I18Z60563-SEM01_SAR_Rev0_part5

FCC ID: 2ACCJH089

RF Exposure Info

Download: PDF
FCCID_3858147

    +      #                                                  uon
Callbratlon Laboratory of                                   i\‘\\w_&/’n}      S   Schweizerischer Kalibrierdienst
Schmid & Partner                                           M                  c   Service suisse d‘étalonnage
  Engineering AG                                           22—R                   Servizio svizzero di taratura
                              £       *                    y n
Zeughausstrasse 43, 8004 Zurich, Switzerland               *g/7/\\\\‘\\*      S   Swiss Calibration Service
                                                              thilals

Accredited by the Swiss Accreditation Service (SAS)                           Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
       (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
   d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
    e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   e Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
   e Feed Point Impedance and Return Loss: These parameters are measured with the dipole
          positioned under the liquid filled phantom. The impedance stated is transformed from the
          measurement at the SMA connector to the feed point. The Return Loss ensures low
          reflected power. No uncertainty required.
      e   Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
          No uncertainty required.
      e   SAR measured: SAR measured at the stated antenna input power.
      e   SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
          connector.
      e   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
          nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D2450V2—853_Jul17                               Page 2 of 8


                                                                                            i




Measurement Conditions
   DASYsystem configuration, as far as not given on page 1.
     DASY Version                                              DASY5                              V52.10.0
     Extrapolation                                   Advanced Extrapolation
     Phantom                                         Modular Flat Phantom
     Distance Dipole Center — TSL                              10 mm                             with Spacer
     Zoom Scan Resolution                              dx, dy, dz =5 mm
     Frequency                                         2450 MHz £ 1 MHz



Head TSL barameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C           39.2               1.80 mho/m
     Measured Head TSL parameters                              (22.0 + 0.2) °C   37.8 16 %          1.87 mho/m £ 6 %
     Head TSL temperature change during test                      <0.5 °C            ~——                     ———

SAR result with Head TSL

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
     SAR measured                                        250 mW input power                     13.4 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W         52.2 Wikg x 17.0 % (k=2)


     SAR averaged over 10 cm* (10 g) of Head TSL                 condition
     SAR measured                                        250 mW input power                     6.26 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W         24.7 Wikg 2 16.5 % (k=2)



Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Body TSL parameters                                  22.0 °C            §2.7              1.95 mho/m
     Measured Body TSL parameters                              (22.0 + 0.2) °C   52.1 26 %          2.04 mho/m x 6 %
      Body TSL temperature change during test                     <0.5 °C            ~——                     ————

SAR result with Body TSL

      SAR averaged over 1 cm* (1 g) of Body TSL                  Condition
      SAR measured                                       250 mW input power                     12.9 W/kg
      SAR for nominal Body TSL parameters                    normalized to 1W         50.4 W/kg x 17.0 % (k=2)

      SAR averaged over 10 cm* (10 g) of Body TSL                condition
      SAR measured                                       250 mW input power                  6.03 W/kg
      SAR for nominal Body TSL parameters                    normalized to 1W         23.8 Wikg + 16.5 % (k=2)



Certificate No: D2450V2—853_Jul17                     Page 3 of 8


(Elll’
         Appendix (Additional assessments outside the scope of SCS 0108)

         Antenna Parameters with Head TSL

               Impedance, transformed to feed point                                           52.0 Q + 5.0 jQ
               Return Loss                                                                       ~25.6 dB



         Antenna Parameters with Body TSL

               Impedance, transformed to feed point                                           49.6 Q + 6.3 jQ
               Return Loss                                                                      —24.0 dB




         General Antenna Parameters and Design

             | Electrical Delay (one direction)                     T                            1.161 ns

         After long term use with 100W radiated power, only a slight warming of the dipole nearthe feedpoint can be measured.

         The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
         second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
         are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
         "Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
         according to the Standard.
         No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
         feedpoint may be damaged.


         Additional EUT Data

               Manufactured by                                                                    SPEAG
               Manufactured on                                                              November 10, 2009




         Certificate No: D2450V2—853_Jul17                         Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                   Date: 20.07.2017

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 853

Communication System: UID 0 — CW; Frequency: 2450 MHz
Medium parameters used: f = 2450 MHz; 0 = 1.87 $/m; & = 37.8; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)
DASY52 Configuration:

   e    Probe: EX3DV4 — SN7349; ConvF(8.12, 8.12, 8.12); Calibrated: 31.05.2017;

   *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e    Electronics: DAE4 S$Sn601; Calibrated: 28.03.2017

   &    Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

   &    DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 112.7 V/m; Power Drift = —0.00 dB
Peak SAR (extrapolated) = 27.0 W/kg
SAR(1 g) = 13.4 W/kg; SAR(10 g) = 6.26 W/kg
Maximum value of SAR (measured) = 21.5 W/kg


       dB
       0

        —4.00

        —8.00

        —12.00

        —16.00

        —20.00
                   0 dB = 21.5 W/kg = 13.32 dBW/kg




Certificate No: D2450V2—853_Jul17                Page 5 of 8


®
Impedance Measurement Plot for Head TSL




                                                                 20 Jul 2017 11:54:15
         CHI sii      1 U FS          4151.965 a   4.9570 a   322.01 pH      2 450.000 000 MHz




          Del

          Ca




          Av
          16 &


          H1d

          CH2




          Ca




          Av
          16 *



          H1d

                    START 2 250.000 000 MHz                            STOP 2 650.000 000 MHz




Certificate No: D2450V2—853_Jul17                   Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                               Date: 21.07.2017

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 853

Communication System: UID 0 — CW; Frequency: 2450 MHz
Medium parameters used: f = 2450 MHz; 0 = 2.04 $/m; & = 52.1; p = 1000 kg/m‘
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C€63.19—2011)

DASY52 Configuration:

   e   Probe: EX3DV4 — SN7349; ConvF(8.1, 8.1, 8.1); Calibrated: 31.05.2017;

   e   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   *   Electronics: DAE4 S$Sn601; Calibrated: 28.03.2017

   *    Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

   *    DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 104.1 V/m; Power Drift = —0.09 dB
Peak SAR (extrapolated) = 25.5 W/kg
SAR(I g) = 12.9 W/kg; SAR(10 g) = 6.03 W/kg
Maximum value of SAR (measured) = 20.0 W/kg


       dB
        0
        —5.00
        —10.00
        —15.00
        —20.00
        —25.00
                   0 dB = 20.0 W/kg = 13.01 dBW/kg




Certificate No: D2450V2—853_Jul17               Page 7 of 8


®
Impedance Measurement Plot for Body TSL




                                                                  24 Jul 2017    14142134
         (CHZ sii     i U s           1149.619 a   6.2656 0 407.02 pH           2 450.000 000 MHz
                                               »   j 3       ~,




          Av
          16 *



          H1d

          CH2




          Ca




          Av
          16s


          H1d

                    START 2 250.000 000 MHz                             STOP 2 650.000 000 MHz




Certificate No: D2450V2—853_Jul17                   Page 8 of 8


Eppiooofiodd whaktvtiietrendteddbtiiceeicielinebeiainiiielialints




              5    P                                                           yol 199y
           Callbfatlon Laboratory of                                         S\\\\\_J//l/’;                            g Schweizerischer Kalibrierdienst
           Schmid & Partner                                                  m                                         c Service suisse d‘étalonnage
              Englneenng AG                                                  T //:\\ >                                        Servizio svizzero di taratura
           Zeughausstrasse 43, 8004 Zurich, Switzerland                       4,”"l/;l\\\‘\‘\\?                        S      gwiss Calibration Service


            Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
            The Swiss Accreditation Service is one of the signatories to the EA
            Multilateral Agreementfor the recognition of calibration certificates

            Client       CTTL—BJ (Auden)                                                                   Certificate No: D2600V2—1012_Jul17

            CALIBRATION CERTIFICATE
             Object                               D2600V2 — SN:1012


             Calibration procedure(s)             QA CAL—05.v9
                                                  Calibration procedure for dipole validation kits above 700 MHz



             Calibration date:                    July 21, 2017


             This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
             The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


             All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


             Calibration Equipment used (M&TE critical for calibration)


             Primary Standards                      ID #                     Cal Date (Certificate No.)                            Scheduled Calibration
             Power meter NRP                       SN: 104778                04—Apr—17 (No. 217—02521/02522)                       Apr—18
             Power sensor NRP—Z91                  SN: 103244                04—Apr—17 (No. 217—02521)                             Apr—18
             Power sensor NRP—Z91                  SN: 103245                04—Apr—17 (No. 217—02522)                             Apr—18
             Reference 20 dB Attenuator             SN: 5058 (20k)           07—Apr—17 (No. 217—02528)                             Apr—18
             Type—N mismatch combination            SN: 5047.2 / 06327       07—Apr—17 (No. 217—02529)                             Apr—18
             Reference Probe EX3DV4                 SN: 7349                 31—May—17 (No. EX3—7349_May17)                        May—18
             DAE4                                   SN: 601                  28—Mar—17 (No. DAE4—601_Mar17)                        Mar—18


             Secondary Standards                    ID #                     Check Date (in house)                                 Scheduled Check
             Power meter EPM—442A                   SN: GB37480704           07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
             Power sensor HP 8481A                  SN: US37202783           07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
             Power sensor HP 8481A                  SN: MY41092317            07—Oct—15 (in house check Oct—16)                    In house check: Oct—18
             RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—16)                    In house check: Oct—18
             Network Analyzer HP 8753E              SN: US37390585            18—Oct—01 (in house check Oct—16)                    In house check: Oct—17


                                                    Name                                     Function                               Signature
             Calibrated by:                        Michael Weber                             Laboratory Technician                 i



             Approved by:                           Katja Pokovic                            Technical Manager             ‘//%%



                                                                                                                                   Issued: July 24, 2017
             This calibration certificate shall not be reproduced except in full without written approvalof the laboratory.



            Certificate No: D2600V2—1012_Jul17                                    Page 1 of 8




                                                                                                                     Nn ey n n en n n n nge e        snn >


uy                                                                                           C




    .      .                                                   al19
Callbfat'on Laboratory of                                    Q\\\\\\_//’z;         g    Schweizerischer Kalibrierdienst
Schmid & Partner                                            m                      c    Service suisse d‘étalonnage
  Englneering AG                                            en itingl                   Servizio svizzero di taratura
                             i      P                       * ty
Zeughausstrasse 43, 8004 Zurich, Switzerland                7///\\\\\\*‘           S    gwiss Calibration Service
                                                                AARS

Accredited by the Swiss Accreditation Service (SAS)                                 Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
      (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
   d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"
Additional Documentation:
    e)   DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
     Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
     Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
     Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
     Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
     SAR measured: SAR measured at the stated antenna input power.
     SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
         SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D2600V2—1012_Jul17                               Page 2 of 8




                                                                               Nn en n n en n n n n e n n n         n n   ey   >


)
n
S
(II
      Measurement Conditions
         DASY system configuration, as far as not given onpage 1.
          DASY Version                                            DASYS5                                 V52.10.0
           Extrapolation                                   Advanced Extrapolation
           Phantom                                          Modular Flat Phantom
           Distance Dipole Center — TSL                              10 mm                              with Spacer
           Zoom Scan Resolution                              dx, dy, dz =5 mm
           Frequency                                         2600 MHz +# 1 MHz


      Head TSL parameters
         The following parameters and calculations were applied.
                                                                     Temperature       Permittivity          Conductivity
           Nominal Head TSL parameters                                  22.0 °C            39.0                 1.96 mho/m
           Measured Head TSL parameters                              (22.0 £0.2) °C     37.2 26 %          2.04 mho/m £ 6 %
           Head TSL temperature change during test                      <0.5 °C             ~—~                      =«»—

      SAR result with Head TSL

           SAR averaged over 1 cm° (1 g) of Head TSL                  Condition
           SAR measured                                        250 mW input power                      14.9 Wikg
           SAR for nominal Head TSL parameters                     normalized to 1W         57.9 Wikg # 17.0 % (k=2)

           SAR averaged over 10 cm* (10 g) of Head TSL                condition
           SAR measured                                        250 mW input power                  6.57 W/kg
           SAR for nominal Head TSL parameters                     normalized to 1W         25.8 Wikg a 16.5 % (k=2)


      Body TSL parameters
         The following parameters and calculations were applied.
                                                                     Temperature        Permittivity            Conductivity
           Nominal Body TSL parameters                                  22.0 °C            52.5                 216 mho/m
           Measured Body TSL parameters                              (22.0 £ 0.2) °C    51.6 46 %          2.22 mho/m * 6 %
            Body TSL temperature change during test                     <0.5 °C             m+>


      SAR result with Body TSL

            SAR averaged over 1 cm* (1 g) of Body TSL                  Condition
            SAR measured                                       250 mW input power                      14.1 Wikg
            SAR for nominal Body TSL parameters                    normalized to 1W          55.5 Wikg + 17.0 % (k=2)

            SAR averaged over 10 cm* (10 g) of Body TSL                condition
            SAR measured                                       250 mW input power                      6.25 Wikg
            SAR for nominal Body TSL parameters                    normalized to 1W          24.8 Wikg 2 16.5 % (k=2)



      Certificate No: D2600V2—1012_Jul17                    Page 3 of 8




                                                                                       Nn opycg m n n       y    >


(llém)
         Appendix (Additional assessments outside the scope of SCS 0108)

         Antenna Parameters with Head TSL

               Impedance, transformed to feed point                                           46.8 Q — 5.0 jQ
               Return Loss                                                                       — 24.2 dB



         Antenna Parameters with Body TSL

               Impedance, transformed to feed point                                           43.5 Q — 5.3 jQ
               Return Loss                                                                      —21.0 dB




         General Antenna Parameters and Design

             fEIectricaI Delay (one direction)                                                   1.1§1 ns

         After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

         The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
         second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
         are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
         "Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
         according to the Standard.
         No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
         feedpoint may be damaged.


         Additional EUT Data

               Manufactured by                                                                    SPEAG
               Manufactured on                                                               October 30, 2007




         Certificate No: D2600V2—1012_Jul17                        Page 4 of 8




                                                                                              ~8 y cans           o                      —


|N
GDASY5 Validation Report for Head TSL



 Test Laboratory: SPEAG, Zurich, Switzerland

 DUT: Dipole 2600 MHz; Type: D2600V2; Serial: D2600V2 — SN: 1012
                                                                                           Date: 20.07.2017




 Communication System: UID 0 — CW; Frequency: 2600 MHz
 Medium parameters used: f = 2600 MHz; 0 = 2.04 S/m; & = 37.2; p = 1000 kg/m*
 Phantom section: Flat Section
 Measurement Standard: DASYS (IEEE/IEC/ANSI C€63.19—2011)

 DASY52 Configuration:

     *   Probe: EX3DV4 — SN7349; ConvF(7.96, 7.96, 7.96); Calibrated: 31.05.2017;

     *   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

     e   Electronics: DAE4 Sn601; Calibrated: 28.03.2017

     *   Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

     e   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


 Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
 Measurement grid: dx=5mm, dy=5mm, dz=5mm
 Reference Value = 113.6 V/m; Power Drift = 0.05 dB
 Peak SAR (extrapolated) = 32.3 W/kg
 SAR(1 g) = 14.9 W/kg; SAR(10 g) = 6.57 W/kg
 Maximum value of SAR (measured) = 25.0 W/kg


         dB
          0

         —5.00

         —10.00

         ~15.00

         —20.00

         —25.00
                    0 dB = 25.0 W/kg = 13.98 dBW/kg




 Certificate No: D2600V2—1012_Jul17              Page 5 of 8




                                                                       Nennma n n n en n m n nc   snn >   —   —


Impedance Measurement Plot for Head TSL




                                                                   20 Jul 2017 12:14:05
         CHI) sii     1 U FS          5: 46.936 a   ~5.0410 a   12.143 pF      2 600.000 000 MHz




         Av
         168


         H1d

          CH2




          Ca




          fivg
          16


          H1d

                      ART 2 400.000 000 MHz                              STOP 2 800.000 000 MHz




Certificate No: D2600V2—1012_Jul17                     Page 6 of 8




                                                                                 8ypyrcgene


DASY5 Validation Report for Body TSL

                                                                                          Date: 21.07.2017

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2600 MHz; Type: D2600V2; Serial: D2600V2 — SN: 1012

Communication System: UID 0 — CW; Frequency: 2600 MHz
Medium parameters used: f = 2600 MHz; 0 = 2.22 $/m; & = 51.6; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   e   Probe: EX3DV4 — SN7349; ConvF(7.94, 7.94, 7.94); Calibrated: 31.05.2017;

   e   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e   Electronics: DAE4 Sn601; Calibrated: 28.03.2017

       Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

   +   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 106.6 V/m; Power Drift = —0.07 dB
Peak SAR (extrapolated) = 30.1 W/kg
SAR(I g) = 14.1 W/kg; SAR(10 g) = 6.25 W/kg
Maximum value of SAR (measured) = 23.4 W/kg


       dB
       0
       —5.20
        —10.40
        ~15.60
        —20.80
        —26.00
                   0 dB = 23.4 W/kg = 13.69 dBW/kg




Certificate No: D2600V2—1012_Jul17             Page 7 of 8




                                                                    Nnpeg n n en n en n en n n n ns n n n   ny   >


)
(I
     Impedance Measurement Plot for Body TSL




                                                                       24 Jul 2017 1i4:50:41
                     §11     i U Fs          1143.541 a   —5.3223 8 114.501 pF     2 600.000 000 MHz
                                                          _1


              Del

              Ca




               Av
               16a


               H1d

               CH2




               Ca




               Av
               16°

               H1d

                           START 2 400.000 000 MHz                           STOP 2 800.000 000 MHz




     Certificate No: D2600V2—1012_Jul17                      Page 8 of 8




                                                                                    Nnepg rrgn en n m n


                                           No. I18Z60563-SEM01
                                               Page 188 of 220


ANNEX I   DAE Calibration Certificate




                                        ©Copyright. All rights reserved by CTTL.


im©
777                                                                                             ty s      n s      n+
=‘-/




     &       &                                                  w‘ ty
 Callbl.'atlon Laboratory of                                  i\‘\\\\ij///'/';         S    Schweizerischer Kalibrierdienst
 Schmid & Partner                                            M                         c    Service suisse d‘étalonnage
    Engineering AG                                           o o mf                         Servizio svizzero di taratura
 Zeughausstrasse 43, 8004 Zurich, Switzerland                 * ///'\\\ y              S    Swiss Calibration Service
                                                                Carhitabe®®

 Accredited by the Swiss Accreditation Service (SAS)                                    Accreditation No.: SCS 0108
 The Swiss Accreditation Service is one of the signatories to the EA
 Multilateral Agreement for the recognition of calibration certificates



 Glossary
 DAE                          data acquisition electronics
 Connector angle              information used in DASY system to align probe sensor X to the robot
                              coordinate system.

 Methods Applied and Interpretation of Parameters
    e  DC Voltage Measurement: Calibration Factor assessed for use in DASY system by
      comparison with a calibrated instrument traceable to national standards. The figure given
      corresponds to the full scale range of the voltmeter in the respective range.
     e    Connector angle: The angle of the connector is assessed measuring the angle
          mechanically by a tool inserted. Uncertainty is not required.
     e    The following parameters as documented in the Appendix contain technical information as a
          result from the performance test and require no uncertainty.
                 DC Voltage Measurement Linearity: Verification of the Linearity at +10%and —10% of
                 the nominal calibration voltage. Influence of offset voltage is included in this
                 measurement.
                 Common mode sensitivity: Influence of a positive or negative common mode voltage on
                 the differential measurement.
                 Channel separation: Influence of a voltage on the neighbor channels not subject to an
                 input voltage.
                 AD Converter Values with inputs shorted: Values on the internal AD converter
                 corresponding to zero input voltage
                 Input Offset Measurement. Output voltage and statistical results over a large number of
                 zero voltage measurements.
                 Input Offset Current: Typical value for information; Maximum channel input offset
                 current, not considering the input resistance.
                 Input resistance: Typical value for information: DAE input resistance at the connector,
                 during internal auto—zeroing and during measurement.
                 Low Battery Alarm Voltage: Typical value for information. Below this voltage, a battery
                 alarm signal is generated.
                 Power consumption: Typical value for information. Supply currents in various operating
                 modes.




 Certificate No : DAE4—1525_Oct17                                 Page 2 of 5




                                                                                 wervopg n ne a io rgn rewr a vewruer e cona wmz


DC Voltage Measurement
   A/D — Converter Resolution nominal
       High Range:          1LSB =          6.1uV ,     full range =    —100...+300 mV
       Low Range:           1LSB =           61nV ,     full range =     —~1....... +3mV
   DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

     Calibration Factors                X                           ¥                           Z
     High Range             405.912 + 0.02% (k=2)     405.954 £0.02% (k=2)        405.400 £ 0.02% (k=2)
     Low Range              8.99166 £ 1.50% (k=2)     4.00980 + 1.50% (k=2)       3.99550 £1.50% (k=2)


Connector Angle

     Connector Angle to be used in DASY system                                             sG56°%1°          J




Certificate No: DAE4—1525_Oct17                       Page 3 of 5




                                                                                Ne n veperg n rag en e mb n ng eene n nene se n en meg n nc



Document Created: 2018-05-16 16:09:05
Document Modified: 2018-05-16 16:09:05

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC