Test Report_SAR-2

FCC ID: 2ACCJH087

RF Exposure Info

Download: PDF
FCCID_3880672

                                     In Collaboration with

          @TTL a                     CALIBRATION LABORATORY
          y
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Http://www.chinattl.cn


Measurement Conditions
   DASY system configuration, as far as notgiven on page 1.
     DASY Version                                         DASY52                                     52.8.8.1258

     Extrapolation                               Advanced Extrapolation

     Phantom                                     Triple Flat Phantom 5.1C

     Distance Dipole Center — TSL                         15 mm                                    with Spacer

     Zoom Scan Resolution                           dx, dy, dz = 5 mm

     Frequency                                       835 MHz + 1 MHz


Head TSL parameters
   The following parameters and calculations were applied.
                                                             Temperature            Permittivity           Conductivity

     Nominal Head TSL parameters                              22.0 °C                  41.5                0.90 mho/m

     Measured Head TSL parameters                        (22.0 + 0.2) °C             40.8 +6 %           0.91 mho/m +6 %

     Head TSL temperature change during test                  <1.0 °C                   ————                     _———
SAR result with Head TSL
     SAR averaged over 1 Cmm‘ (1 g) of Head TSL                         Condition
     SAR measured                                                 250 mW input power                   2.43 mW / g
     SAR for nominal Head TSL parameters                             normalized to 1W          9.59 mW /g + 20.8 % (k=2)
     SAR averaged over 10 cm (10 g) of Head TSL                         Condition

     SAR measured                                                 250 mW input power                   1.59 mW / g

     SAR for nominal Head TSL parameters                             normalized to 1W          6.29 mW /g + 20.4 % (k=2)

Body TSL parameters
   The following parameters and calculations were applied.
                                                             Temperature            Permittivity           Conductivity

     Nominal Body TSL parameters                              22.0 °C                  55.2                0.97 mho/m

     Measured Body TSL parameters                        (22.0 + 0.2) °C             54.7 £6 %           0.98 mho/m +6 %

     Body TSL temperature change during test                  <1.0 °C                                            ————
SAR result with Body TSL
     SAR averaged over 1 Cm" (1 g) of Body TSL                          Condition
     SAR measured                                                 250 mW input power                   2.44 mW / g

     SAR for nominal Body TSL parameters                             normalized to 1W          9.65 mW /g + 20.8 % (k=2)
     SAR averaged over 10 cm> (10 g) of Body TSL                        Condition

     SAR measured                                                 250 mW input power                   1.63 mW / g

     SAR for nominal Body TSL parameters                             normalized to 1W          6.46 mW /g + 20.4 % (k=2)
Certificate No: Z16—97239                              Page 3 of 8


          E7TTL $.P.g—2
          V
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Http://www.chinattl.cn


 Appendix

 Antenna Parameters with Head TSL


     Impedance, transformed to feed point                                   49.20— 3.41j0Q

     Return Loss                                                              — 29.10B


 Antenna Parameters with Body TSL

     Impedance, transformed to feed point                                   45.80— 3.25j0

     Return Loss                                                              — 25.10B


 General Antenna Parameters and Design

     Electrical Delay (one direction)                                         1.500 ns




 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can
 be measured.

 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly
 connected to the second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some
 of the dipoles, small end caps are added to the dipole arms in order to improve matching when loaded
 according to the position as explained in the "Measurement Conditions" paragraph. The SAR data are not
 affected by this change. The overall dipole length is still according to the Standard.
 No excessive force must be applied to the dipole arms, because they might bend or the soldered
 connections near the feedpoint may be damaged.



 Additional EUT Data

     Manufactured by                                                             SPEAG




Certificate No: Z16—97239                             Page 4 of 8


         ETTL $.2.g.2
          ‘/
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Http://www.chinattl.cn


DASYS5 Validation Report for Head TSL                                   Date: 12.08.2016
Test Laboratory: CTTL, Beijing, China
DUT : Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN: 40105
   Communication System: UID 0, CW; Frequency: 835 MHz; Duty Cycle: 1: 1
   Medium parameters used: f= 835 MHz; 0 = 0.912 S/m; s, = 40.78; p = 1000 kg/m*
   Phantom section: Center Section
   Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2007)
DASY5 Configuration:

            e    Probe: EX3DV4 — SN7433; ConvF(9.82, 9.82, 9.82); Calibrated: 9/26/2016;
            e    Sensor—Surface: 2mm (Mechanical Surface Detection)
            e    Electronics: DAF4 Sn771; Calibrated: 2/2/2016
            e    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
            e    Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7372)

        Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=5mm,
        dy=5mm, dz=5mm
        Reference Value = 49.08V/m; Power Drift = 0.05 dB
        Peak SAR (extrapolated) = 3.62 W/kg
        SAR(I g) =2.43 W/kg; SAR(10 g) = 1.59 W/kg
        Maximum value of SAR (measured) = 3.08 W/kg

           dB
           0


           —2.09


           —4.18


           —6.26


           —8.35




                      0 dB = 3.08 W/kg = 4.89 dBW/kg




Certificate No: Z16—97239                              Page 5 of 8


            fi                         l;ColIabonmonewith 3


            ‘                         CALIBRATION LABORATORY

        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079         Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com         Http://www.chinattl.cn




        Impedance Measurement Plot for Head TSL


      Tri s11 Log Mag 10.00dB/ Ref 0.000dsB [F1]
        900       1 ~g35. 000006 mhz —29. 063 ds
       40. 00


       30. 00
       20. 00
       10. 00
       0. 000 p
      —10. 00
      —20. 00                                                         4
      —30. 00
      —40. 00

      —50. 00                                                         s
    |PM@H s11 smith (R+jx) scale 1.000Uu [F1 del]
       »1   835.00000 mhz    49.239 o —3.4139 o    SSW’               \




                                                        IFBW 100Hz




Certificate No: Z16—97239                              Page 6 of 8


                              8      In Collaboration with

         "__—j"TTI, _
                                     CALIBRATION LABORATORY
          V
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079      Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com      Http://www.chinattl.cn




DASYS Validation Report for Body TSL                                  Date: 12.08.2016
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN: 40105
   Communication System: UID 0, CW; Frequency: 835 MHz; Duty Cycle: 1:1
   Medium parameters used: f= 835 MHz; 0 = 0.983 S/m; g, = 54.74; p = 1000 kg/m3
  Phantom section: Right Section
  Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2007)
DASY5 Configuration:

            e    Probe: EX3DV4 — SN7433; ConvF(9.5,9.5, 9.5); Calibrated: 9/26/2016;
            e    Sensor—Surface: 2mm (Mechanical Surface Detection)
            e    Electronics: DAE4 Sn771; Calibrated: 2/2/2016
            e    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
            e    Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7372)

        Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=5mm,
        dy=5mm, dz=5mm
        Reference Value = 57.10 V/m; Power Drift = —0.02 dB
        Peak SAR (extrapolated) = 3.54 W/kg
        SAR(1 g) = 2.44 W/kg; SAR(10 g) = 1.63 W/kg
        Maximum value of SAR (measured) = 3.06 W/kg


           dB
            0


            —1.97


            —3.94


            —5.91


            —1.88

                          L* _———1
            —9.85

                       0 dB = 3.06 W/kg = 4.86 dBW/kg




Certificate No: Z16—97239                              Page 7 of 8


            iiEEEF!EF::;;E;EiE;/      ::Lllflllflfl'flflJfiE;UB()iz;NUI'Y
            m
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079        Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com        Http://www.chinattl.cn


        Impedance Measurement Plot for Body TSL


      Tri s11 Log Mag 10.00dB/ Rref 0.000ds [F1]

       990         1 ~g35. 00000 mhz —25. 066 d
       40. 00
       30. 00
       20. 00
       10. 00
       0. 000 p,
      —10. 00
      —20. 00
      —30. 00
      —40. 00

      —50. 00




                                              7
                                                                     Ah.
    P0 sii smith (r+jx) scale 1.000u [F1 del]
       »1   835.00000 mHz     45.756 o —3.2528 o   58.59
                                                                            _




Certificate No: Z16—97239                              Page 8 of 8


Calibration
      f     Laboratory of                                        um
                                                                 s3                                       g Schweizerischer Kalibrierdienst
Schmid & Partner                                                SBE/W/BE                                  @ Service suisse d‘étalonnage
  Engineering AG                                                 Tof                                        Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                     */,fi. w                                  S swiss Calibration Service

Accredited by the Swiss Accreditation Service (SAS)                                                        Accreditation No.: SCS 0108
The Swiss Accreditation Service is oneof the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client       SGS—SZ (Auden)                                                                   Certificate No: D1750V2—1149_Jun16

CALIBRATION CERTIFICATE                                                                                    $                                  I
 Object                               D1750V2 — SN:1149


 Calibration procedure(s)             QA CAL—05.v9
                                      Calibration procedure for dipole validation kits above 700 MHz



 Calibration date:                   June 23, 2016


This calibration certificate documents the traceabilty to national standards, which realize the physical units of measurements (S1).
The measurements and the uncertainties with confidence probabilty are given on the following pages and are part of the cortficate.

All callbrations have been conducted in the closed laboratory facifty: environment temperature (22 + 3)°C and humidity < 70%.


Calibration Equipment used (M&TE crtical for calibration)

 Primary Standards                     1D #                      Cal Date (Certificate No.)                       Scheduled Calibration
 Power meter NRP                       SN: 104778                06—Apr—16 (No. 217—02288/02289)                  Apr—17
 Power sensor NRP—Z91                  SN: 108244                06—Apr—16 (No. 217—02288)                        Apr—17
 Power sensor NRP—Z91                  SN: 103245                06—Apr—16 (No. 217—02289)                        Apr—17
 Reference 20 dB Attenuator            SN: 5058 (20k)            05—Apr—16 (No. 217—02202)                        Apr—17
Type—N mismatch combination            SN: 5047.2 / 06827        05—Apr—16 (No. 217—02205)                        Apr—17
 Reference Probe EX3DV4                SN: 7849                  15—Jun—16 (No. EX3—7349_Junt6)                   Jun—17
 DAE4                                  SN: 601                   30—Dec—15 (No. DAE4—601_Dect5)                    Dec—16


Secondary Standards                    1D #                      Check Date (in house)                            Scheduled Check
 Power meter EPM—442A                  SN: GB37480704            07—0ct—15 (No. 217—02222)                         In house check: Oct—16
 Power sensor HP 8481A                 SN: US37202783            07—0ct—15 (No. 217—02222)                        in house check: Oct—16
 Power sensor HP 8481A                 SN: MY41092317            07—0ct—15 (No. 217—02223)                        in house chack: Oct—16
 RF generator R&S SMT—06               SN: 100972                15—Jun—15 (in house check Jun—15)                in house check: Oct—16
 Network Analyzer HP 8753E             S Ussrsaoses              18—0ct—01 (in house check Oct—15)                in house check: Oct—16

                                       Name                                   Function                             Signature
Calibrated by:                        Michael Waber                           Laboratory Technician               %



                                      Katja Pokovic                          Technical Manager                   ////%/:
Approved by:



                                                                                                                  Issued: June 28, 2016
This callbration certificate shall not be reproduced exceptin full without written approval of the laboratory.

Certificate No: D1750V2—1149_Jun16                                   Page 1 of 8


Calibration Laboratory of                                                     g Schweizerischer Kalibrierdienst
Schmid & Partner                                                                Service suisse d‘étalonnage
  Engineering AG                                                              C   Servizio svizzero di taratura
Zeughausstrasse 48, 8004 Zurich, Switzerland
                                                                 mlutaied
                                                                              S swiss Calibration Service

Accredited by the Swiss Accreditation Service (GAS)                            Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM xy,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques®, June 2013
   b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
      devices used in close proximity to the ear(frequency range of 300 MHz to 3 GHz)",
         February 2005
    c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
       communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
    e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
    *    Measurement Conditions: Further details are available from the Validation Report at the end
         of the certificate. All figures stated in the certificate are valid at the frequency indicated.
    *    Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
         point exactly below the center marking of the flat phantom section, with the arms oriented
         parallel to the body axis.
    *    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
         positioned under the liquid filled phantom. The impedance stated is transformed from the
         measurement at the SMA connector to the feed point. The Return Loss ensures low
         reflected power. No uncertainty required.
    *    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
         No uncertainty required.
    *    SAR measured: SAR measured at the stated antenna input power.
    e    SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    «_   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
 The reported uncertainty of measurementis stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D1750V2—1149_Jun16                              Page 2 of 8


Measurement Conditions
    DASY system contfiguration, as far as not iven on page 1.
      DASY Version                                              DASY5                               V52.8.8
      Extrapolation                                   Advanced Extrapolation
      Phantom                                          Modular Flat Phantom
      Distance Dipole Center — TSL                               10 mm                            with Spacer
      Zoom Scan Resolution                               dx. dy, dz =5 mm
      Frequency                                         1750 MHz a 1 MHz



Head TSL parameters
    The following parameters and calculations were applied.
                                                                Temperature       Permittivity          Conductivity
      Nominal Head TSL parameters                                  22.0 °C           40.1               1.37 mho/m
      Measured Head TSL parameters                              (22.0 £0.2) °C    39.8 2 6 %          1.36 mho/m + 6 %
      Head TSL temperature change during test                      <0.5 °C                                    ——

SAR result with Head TSL

      SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
      SAR measured                                        250 mW input power                     9.16 Wikg
      SAR for nominal Head TSL parameters                     normalized to 1W        36.7 Wikg # 17.0 % (k=2)

      SAR averaged over 10 cm* (10 g) of Head TSL                 condition
      SAR measured                                        250 mW input power                     4.87 Wikg
      SAR for nominal Head TSL parameters                     normalized to 1W        19.5 Wikg 2 16.5 % (k=2)


Body TSL parameters
   The following parameters and calculations were applied.
                                                                Temperature                            Conductivity
      Nominal Body TSL parameters                                  22.0 °C           53.4               1.49 mho/m
      Measured Body TSL parameters                              (22.0 £ 0.2) °C   53.8 £ 6 %         1.50 mho/m + 6 %
      Body TSL temperature change during test                      <0.5 °C


SAR result with Body TSL

      SAR averaged over 1 cm° (1 g) of Body TSL                  Condition
      SAR measured                                        250 mW input power                     9.27 Wikg
      SAR for nominal Body TSL parameters                     normalized to 1W        37.0 Wikg » 17.0 % (k=2)

      SAR averaged over 10 cm* (10 g) of Body TSL                 condition

      SAR measured                                        250 mW input power                     4.93 Wikg
      SAR for nominal Body TSL parameters                     normalized to 1W        19.7 Wikg 2 16.5 % (k=2)



Certificate No: D1750V2—1149_Jun16                     Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 0108)
Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           49.0 0 — 0.5 j

      Return Loss                                                                       —38.7 dB


Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           43.6 Q + 0.0 jQ
      Return Loss                                                                       —23.3 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                   1.213 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions® paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.



Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                       +         April 10, 2015




Certificate No: D1750V2—1149_Jun16                       Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                    Date: 23.06.2016

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1750 MHz; Type: D1750V2; Serial: D1750V2 — SN: 1149

Communication System: UID 0 — CW; Frequency: 1750 MHz
Medium parameters used: f = 1750 MHz; 0 = 1.36 S/m; s:= 39.8; p = 1000 kg/m‘
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

    *    Probe: EX3DV4 — SN7349; ConvF(8.46, 8.46, 8.46); Calibrated: 15.06.2016;

    *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    +    Electronics: DAE4 Sn601; Calibrated: 30.12.2015

    *    Phantom: Flat Phantom 5.0 (front); Type: QDOOOP50AA; Serial: 1001

    «_   DASYS52 52.8.8(1222); SEMCAD X 14.6.10(7372)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=Smm, dz=5mm
Reference Value = 105.0 V/m; Power Drift = 0.04 dB
Peak SAR (extrapolated) = 16.7 W/kg
SAR(I g) =9.16 W/kg; SAR(10 g) = 4.87 W/kg
Maximum value of SAR (measured) = 13.9 W/kg
         dB
         0

         —3.60

         —7.20

         —10.80

         14.40

         ~18.00
                   0 dB = 13.9 W/kg = 11.43 dBW/kg




Certificate No: D1750V2—1149_Jun16              Page 5 of 8


Impedance Measurement Plot for Head TSL




                                                                   23 Jun 2016    10:05:27
          CHI sai     i uors          4: 48,973 a   ~S14.72 me 477.73 pF         1 750.000 008 MHz




          fvg
          16                                                       =

          Hia

          cn2 si1__Los          _5 dB/REF —20_a8        _ 41=38.724 dB           1 750.000 doe IHz




          by
          169

          Hd

                    start 1 550.000 ooo Mz




Certificate No: D1750V2—1149_Jun16                    Page6 of 8


DASY5 Validation Report for Body TSL

                                                                                    Date: 23.06.2016

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1750 MHz; Type: D1750V2; Serial: D1750V2 — SN: 1149

Communication System: UID 0 — CW; Frequency: 1750 MHz
Medium parameters used: f = 1750 MHz; 0 = 1.5 S/m; s = 53.8; p = 1000 kg/m‘
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

    «_   Probe: EX3DV4 — SN7349; ConvF(8.25, 8.25, 8.25); Calibrated: 15.06.2016;

    *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    *    Electronics: DAE4 Sn601; Calibrated: 30.12.2015

    *    Phantom: Flat Phantom 5.0 (back); Type: QDOOOP50AA; Serial: 1002

    «_   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7372)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 100.6 V/m; Power Drift =0.01 dB
Peak SAR (extrapolated) = 16.4 W/kg
SAR(I g) = 9.27 W/kg:; SAR(10 g) = 4.93 W/kg
Maximum value of SAR (measured) = 13.9 W/kg

         dB
         0

         —3.60

         —7.20

         —10.80

         14.40

         —18.00
                   0 dB = 13.9 W/kg = 11.43 dBW/kg




Certificate No: D1750V2—1149_Jun16              Page 7 of 8


Impedance Measurement Plot for Body TSL




                                                                23 Jun 2016 1g:e2:14
          CHI siai     i uors           4143594 a   0.0449 0. 40054 pH      1 75e.000 000 MHz




          Avg
          16

          HLd

          cn2 si1      Los      _5 de/Rer —20 ce       _       1:—23.296 ds    1 750.000 go0 Mz




                                                                     {—LL_L



          fva
                                               J
        m
        @




                     staRrt 1 s50.000 oo8 nz                            SToF 1 950.000 000 Niz




Certificate No: D1750V2—1149_Jun16                   Page 8 of 8


                 t“                      In Collaboration with                  \\‘\y"/
                                                                                      7          A EUn

                 a=‘7"]"J, e ::—~~*                                                                              aREW
             Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China "4,,//'_\\3: v CAUBRAHON
             Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504           '/,,,,,h‘\\\\                   CNAS LO570
             E—mail: cttl@chinattl.com       Http://www.chinattl.cn
      Client              SGS(Boce)                                         Certificate No:      Z16—97240

CALIBRATION CERTIFICATE

Object                                     D1900V2 — SN: 5028

Calibration Procedure(s)                   FD—211—003—01
                                          Calibration Procedures for dipole validation kits

Calibration date:                          December 7, 2016


This calibration Certificate documents the traceability to national standards, which realize the physical units of
measurements(Sl). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.


All calibrations have been conducted in the closed laboratory facility: environment temperature(22+3)‘C and
humidity<70%.


Calibration Equipment used (M&TE critical for calibration)


Primary Standards                 ID #               Cal Date(Calibrated by, Certificate No.)       Scheduled Calibration
 Power Meter     NRP2             101919            27—Jun—16 (CTTL, No.J16X04777)                         Jun—17
 Power sensor       NRP—Z91       101547            27—Jun—16 (CTTL, No.J16X04777)                         Jun—17
 Reference Probe EX3DV4           SN 7433           26—Sep—16(SPEAG,No.EX3—7433_Sep16)                     Sep—17
 DAE4                             SN 771            02—Feb—16(CTTL—SPEAG,No.Z16—97011)                     Feb—17


 Secondary Standards              ID #              Cal Date(Calibrated by, Certificate No.)        Scheduled Calibration
 Signal Generator E4438C          MY49071430        O01—Feb—16 (CTTL, No.J16X00893)                       Jan—17
 Network Analyzer E5071C          MY46110673         26—Jan—16 (CTTL, No.J16X00894)                       Jan—17



                                   Name                          Function                                Signature
Calibrated by:                   Zhao Jing                 SAR Test Engineer                             z Z/

Reviewed by:                     Qi Dianyuan                SAR Project Leader                    ML/

Approved by:                     Lu Bingsong                Deputy Director of the laboratory       ’%_W;{C 6

                                                                                      Issued: Decemer 11, 2016
This calibration certificate shall not be reproduced except in full without written approval of the laboratory.



    Certificate No: Z16—97240                              Page 1 of 8


         i’Tfl:                     ;Collabomtionewith a

                                    CALIBRATION LABORATORY
          e
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Http://www.chinattl.cn


Glossary:
TSL                       tissue simulating liquid
Convr                      sensitivity in TSL / NORMx,y,z
N/A                       not applicable or not measured


Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak
   Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
  Communications Devices: Measurement Techniques", June 2013
b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) For hand—held
    devices used in close proximity to the ear (frequency range of 300MHz to 3GHz)", February
    2005
c) IEC 62209—2, "Procedure to measure the Specific Absorption Rate (SAR) For wireless
   communication devices used in close proximity to the human body (frequency range of
    30MHz to 6GHz)", March 2010
d) KDB865664, SAR Measurement Requirements for 100 MHz to 6 GHz


Additional Documentation:
e) DASY4/5 System Handbook


Methods Applied and Interpretation of Parameters:
e  Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
e    Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
e    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
e    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
     SAR measured: SAR measured at the stated antenna input power.
e    SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
     connector.
e    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
     nominal SAR result.


    The reported uncertainty of measurement is stated as the standard uncertainty of
    Measurement multiplied by the coverage factor k=2, which for a normal distribution
    Corresponds to a coverage probability of approximately 95%.


Certificate No: Z16—97240                              Page 2 of 8


                                    In Collaboration with

          a_‘/"/‘J
           imaggye
                   a
                   CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Http://www.chinattl.cn


Measurement Conditions
   DASY system configuration, as far as not given on page 1.
     DASY Version                                           DASY52                                   52.8.8.1258

     Extrapolation                               Advanced Extrapolation

     Phantom                                     Triple Flat Phantom 5.1C

     Distance Dipole Center — TSL                           10 mm                                  with Spacer

     Zoom Scan Resolution                           dx, dy, dz = 5 mm

     Frequency                                       1900 MHz + 1 MHz


Head TSL parameters
   The following parameters and calculations were applied.
                                                            Temperature             Permittivity           Conductivity

     Nominal Head TSL parameters                              22.0 °C                  40.0                1.40 mho/m

     Measured Head TSL parameters                        (22.0 + 0.2) °C             40.2 + 6 %          1.38 mho/m +6 %

     Head TSL temperature change during test                  <1.0 °C                   ————                     ————
SAR result with Head TSL
   SAR averaged over 1 C# (1 g) of Head TSL                             Condition
     SAR measured                                                   250 mW input power                 10.1 mW / g

     SAR for nominal Head TSL parameters                             normalized to 1W          40.7 mW ig + 20.8 % (k=2)
     SAR averaged over 10 cm (10 g) of Head TSL                         Condition

     SAR measured                                                   250 mW input power                 5.24 mW / g

     SAR for nominal Head TSL parameters                             normalized to 1W          21.1 mW /g + 20.4 % (k=2)

Body TSL parameters
   The following parameters and calculations were applied.
                                                            Temperature             Permittivity           Conductivity

     Nominal Body TSL parameters                              22.0 °C                  53.3                1.52 mho/m

     Measured Body TSL parameters                        (22.0 + 0.2) °C             54.3 + 6 %          1.51 mho/m +6 %

     Body TSL temperature change during test                  <1.0 °C
SAR result with Body TSL
     SAR averaged over 1 cm> (1 g) of Body TSL                          Condition

     SAR measured                                                   250 mW input power                 10.3 mW / g

     SAR for nominal Body TSL parameters                             normalized to 1W          41.6 mW ig + 20.8 % (k=2)

     SAR averaged over 10 cm> (10 g) of Body TSL                        Condition

     SAR measured                                                   250 mW input power                 5.32 mW / g
     SAR for nominal Body TSL parameters                             normalized to 1W          21.4 mW ig + 20.4 % (k=2)
Certificate No: Z16—97240                              Page 3 of 8


          E!T\fl: ;Collabonm'onewith a

          iiaagyye~                 CALBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Http://www.chinattl.cn


 Appendix

 Antenna Parameters with Head TSL


     Impedance, transformed to feed point                                   51.80+ 5.90j0

     Return Loss                                                              — 24.40B



 Antenna Parameters with Body TSL

     Impedance, transformed to feed point                                   48.10+ 5.82j0

     Return Loss                                                              — 24.10B


 General Antenna Parameters and Design

     Electrical Delay (one direction)                                         1.306 ns




 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can
 be measured.

 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly
 connected to the second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some
 of the dipoles, small end caps are added to the dipole arms in order to improve matching when loaded
 according to the position as explained in the "Measurement Conditions" paragraph. The SAR data are not
 affected by this change. The overall dipole length is still according to the Standard.
 No excessive force must be applied to the dipole arms, because they might bend or the soldered
 connections near the feedpoint may be damaged.



 Additional EUT Data

     Manufactured by                                                             SPEAG




Certificate No: Z16—97240                              Page 4 of 8


          mm&                       In Collaboration with

         w‘/"/"J, @9
                                    CALIBRATION LABORATORY
          apg»
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Http://www.chinattl.cn


DASY5 Validation Report for Head TSL                                     Date: 12.07.2016
Test Laboratory: CTTL, Beijing, China
DUT : Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN: 50028
   Communication System: UID 0, CW; Frequency: 1900 MHz; Duty Cycle: 1:1
   Medium parameters used: f= 1900 MHz; 0 = 1.383 S/m; er = 40.16; p = 1000 kg/m3
   Phantom section: Center Section
   Measurement Standard: DASYS5 (IEEE/IEC/ANSI C63.19—2007)
DASY5 Configuration:

            e    Probe: EX3DV4 — SN7433; ConvF(7.98, 7.98, 7.98); Calibrated: 9/26/2016;
            e    Sensor—Surface: 2mm (Mechanical Surface Detection)
            e    Electronics: DAF4 Sn771; Calibrated: 2/2/2016
            e    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
            e    Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7372)

        System Performance Check/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid:
        dx=5mm, dy=5mm, dz=5mm
        Reference Value = 102.5 V/m; Power Drift = 0.06 dB
        Peak SAR (extrapolated) = 18.7 W/kg
        SAR(1 g) =10.1 W/kg; SAR(10 g) = 5.24 W/kg
        Maximum value of SAR (measured) = 14.4 W/kg


           dB
            0


            —3.61


            —1.22


            —10.83


            —14.44

                          k
            —18.05

                       0 dB = 14.4 W/kg = 11.58 dBW/kg




Certificate No: Z16—97240                              Page 5 of 8


           b@                            in Collaboration with
          a_‘/"/‘]J, a
           V                             CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com      Http:/www.chinattl.cn




        Impedance Measurement Plot for Head TSL


     Tri s11 Log Mag 10.00dB/ Rref 0.000dB [F1]
       50. 00
                  »1   1.9000000 GHz —24.374 dB
       40. 00

       30. 00

       20. 00

       10. 00

       0. 000 p




                                                                            P &A
      —10. 00

      —20. 00

      —30. 00

      —40. 00

      —50. 00                                                           a
    >WGi sil1 smith (R+jx) scale 1.000Uu [F1 Ddel]
      »1 1.9000000 GHz        51.759 o     5.9039 n;}f\




    Torempopmprremurceerorenronoommiraciirrarrrtaisoreonies




Certificate No: Z16—97240                                 Page 6 of 8


          mm &                      In Collaboration with

          aa_‘/"/"J, a
                                    CALIBRATION LABORATORY
            \pppemr~
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Http://www.chinattl.cn




DASY5S Validation Report for Body TSL                                    Date: 12.07.2016
Test Laboratory: CTTL, Beijing, China
DUT : Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN: 50028
   Communication System: UID 0, CW; Frequency: 1900 MHz; Duty Cycle: 1:1
   Medium parameters used: f= 1900 MHz; 0 = 1.506 S/m; &, = 54.26; p = 1000 kg/m*
   Phantom section: Right Section
   Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2007)
DASY5 Configuration:

            e    Probe: EX3DV4 — SN7433; ConvF(7.7, 7.7, 7.7); Calibrated: 9/26/2016;
            e    Sensor—Surface: 2mm (Mechanical Surface Detection)
            e    Electronics: DAF4 Sn771; Calibrated: 2/2/2016
            e    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
            e    Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7372)

        System Performance Check/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid:
        dx=5mm, dy=5mm, dz=5mm
        Reference Value = 99.69 V/m; Power Drift = —0.03 dB
        Peak SAR (extrapolated) = 18.8 W/kg
        SAR(I g) = 10.3 W/kg; SAR(10 g) = 5.32 W/kg
        Maximum value of SAR (measured) = 14.8 W/kg


           dB
            0


            —3.50


            —£.00


            —10.50


            —14.00

                         L.                       y            es           y
            —17.50

                      0 dB = 14.8 W/kg =11.70 dBW/kg




Certificate No: Z16—97240                             Page 7 of 8


            _ifi©                     in Collaboration with
                      a
            w ‘/"/"J, CALIBRATION
            -‘,                   LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com      Http://www.chinattl.cn


        Impedance Measurement Plot for Body TSL


      Trl Ss11 Log Mag 10.00ds/ ref 0.000ds [F1]
        ‘9— 90     (<1~1,5000000 enz —24. 127 d
        40. 00
        30. 00
        20. 00
        10. 00
        0. 000 p                                                                       4
       —10. 00
      —20. 00
      —30. 00
       —40. 00

      —50. 00                                                         +
     PB si1 smith (R+jxX) scale 1.000U [F1 Ddel]
       »1   1.9000000 GoHz    48.141 o   5.8232 o   487.7



                                               /




                                                                              inould
                                                                            a e men




Certificate No: Z16—97240                               Page 8 of 8



Document Created: 2018-06-04 11:13:51
Document Modified: 2018-06-04 11:13:51

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC