I17Z61902-SEM01_SAR_Rev0_part3

FCC ID: 2ACCJBT10

RF Exposure Info

Download: PDF
FCCID_3666887

mm
 TTL
apge"




  Impedance Measurement Plot for Body TSL




                                                                 19 Jul 2017    @9:16:58
           CHI) sii     1 U FS         1:47.896 a   —3.6926 n 48.966 pF
                                                    *
                                                                               835.000 000 MHz




           fAvq
           16


           H1d

            CH2




           Ca




            Av
            16°

            H1d

                      START 635.000 000 MHz                            STOP 1 035.000 000 MHz




  Certificate No: D835V2—40069_Jul17                   Page 8 of 8




                                                                               Ns epgrge es


—( )
m
~Calibration Laboratory of                                                                                           Schweizerischer Kalibrierdienst




                                                                                                            n 0
 Schmid & Partner                                                                                                    Service suisse d‘étalonnage
   Engineering AG                                                                                                    Servizio svizzero di taratura
 Zeughausstrasse 43, 8004 Zurich, Switzerland                                                                        Swiss Calibration Service



  Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
  The Swiss Accreditation Service is one of the signatories to the EA
  Multilateral Agreementfor the recognition of calibration certificates

  Client      CTTL—BJ (Auden)                                                                    Certificate No: D1900V2—5d101_Jul17

  CALIBRATION CERTIFICATE
  Object                                D1900V2 — SN:5d101


   Calibration procedure(s)             QA CAL—05.v9
                                        Calibration procedure for dipole validation kits above 700 MHz



   Calibration date:                    July 26, 2017


   This calibration certificate documents the traceability to national standards, which realize the physicalunits of measurements (S1).
   The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


   All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


   Calibration Equipment used (M&TE critical for calibration)


   Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
   Power meter NRP                        SN: 104778                04—Apr—17 (No. 217—02521/02522)                       Apr—18
   Power sensor NRP—Z91                   SN: 103244                04—Apr—17 (No. 217—02521)                             Apr—18
   Power sensor NRP—Z91                   SN: 103245                04—Apr—17 (No. 217—02522)                             Apr—18
   Reference 20 dB Attenuator             SN: 5058 (20k)            07—Apr—17 (No. 217—02528)                             Apr—18
   Type—N mismatch combination            SN: 5047.2 / 06327        O7—Apr—17 (No. 217—02529)                             Apr—18
   Reference Probe EX3DV4                 SN: 7349                  31—May—17 (No. EX8—7349_May17)                        May—18
   DAE4                                   SN: 601                   28—Mar—17 (No. DAE4—601_Mar17)                        Mar—18


   Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
   Power meter EPM—442A                   SN: GB37480704            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
   Power sensor HP 8481A                  SN: US37292783            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
   Power sensor HP 8481A                  SN: MY41092317            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
   RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—16)                     In house check: Oct—18
   Network Analyzer HP 8753E              SN: US37390585            18—Oct—01 (in house check Oct—16)                     In house check: Oct—17


                                          Name                                   Function
   Calibrated by:                         Johannes Kurikka                       Laboratory Technician



   Approved by:                           Katia Pokovic                          Technical Manager

                                                                                                                      ePX Issued: July 26, 2017
   This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


  Certificate No: D1900V2—5d101_Jul17                                   Page 1 of 8


m©
777
T.                                                                                                ~o —         200




         s      x                                                   al m
     Callbll'atlon Laboratory of                                  Q\\‘L\\:////’/,;   S   Schweizerischer Kalibrierdienst
     Schmid & Partner                                            ;Ia\&_/fl            c   Service suisse d‘étalonnage
       Engineering AG                                            ;//—/%\3                Servizio svizzero di taratura
     Zeughausstrasse 43, 8004 Zurich, Switzerland                 *g/fi‘\\\\*         S   Swiss Calibration Service
                                                                     "Whilal?



     Accredited by the Swiss Accreditation Service (SAS)                             Accreditation No.: SCS 0108
     The Swiss Accreditation Service is one of the signatories to the EA
     Multilateral Agreement for the recognition of calibration certificates

     Glossary:
     TSL                          tissue simulating liquid
     ConvrF                       sensitivity in TSL / NORM x,y,z
     N/A                          not applicable or not measured

     Calibration is Performed According to the Following Standards:
        a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
           Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
           Communications Devices: Measurement Techniques", June 2013
        b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
            (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
            300 MHz to 6 GHz)", July 2016
        c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
            communication devices used in close proximity to the human body (frequency range of 30
            MHz to 6 GHz)", March 2010
        d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"
     Additional Documentation:
         e) DASY4/5 System Handbook

     Methods Applied and Interpretation of Parameters:
          Measurement Conditions: Further details are available from the Validation Report at the end
          of the certificate. All figures stated in the certificate are valid at the frequency indicated.
          Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
          point exactly below the center marking of the flat phantom section, with the arms oriented
          parallel to the body axis.
          Feed Point Impedance and Return Loss: These parameters are measured with the dipole
          positioned under the liquid filled phantom. The impedance stated is transformed from the
          measurement at the SMA connector to the feed point. The Return Loss ensures low
          reflected power. No uncertainty required.
          Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
          No uncertainty required.
          SAR measured: SAR measured at the stated antenna input power.
          SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
               connector.
               SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
               nominal SAR result.
       The reported uncertainty of measurement is stated as the standard uncertainty of measurement
       multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
       probability of approximately 95%.




     Certificate No: D1900V2—5d101_Jul17                              Page 2 of 8


(EM)
                                                                                                            w




       Measurement Conditions
          DASY system configuration, as far as not given on page 1.
            DASY Version                                              DASY5                              V52.10.0
            Extrapolation                                   Advanced Extrapolation
            Phantom                                         Modular Flat Phantom
            Distance Dipole Center — TSL                              10 mm                             with Spacer
            Zoom Scan Resolution                              dx, dy, dz =5 mm
            Frequency                                         1900 MHz £ 1 MHz



       Head TSL parameters
          The following parameters and calculations were applied.
                                                                      Temperature       Permittivity         Conductivity
            Nominal Head TSL parameters                                  22.0 °C           40.0                 1.40 mho/m
            Measured Head TSL parameters                              (22.0 + 0.2) °C   40.7 #6 %          1.39 mho/m + 6 %
            Head TSL temperature change during test                      <0.5 °C            ~——                     ———

       SAR result with Head TSL

            SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
            SAR measured                                        250 mW input power                     9.93 W/kg
            SAR for nominal Head TSL parameters                     normalized to 1W         40.0 W/kg + 17.0 % (k=2)


            SAR averaged over 10 cm* (10 g) of Head TSL                 condition
            SAR measured                                        250 mW input power                     5.23 W/kg
            SAR for nominal Head TSL parameters                     normalized to 1W         21.0 W/kg + 16.5 % (k=2)


       Body TSL parameters
          The following parameters and calculations were applied.
                                                                      Temperature       Permittivity            Conductivity
            Nominal Body TSL parameters                                   22.0 °C           53.3                1.52 mho/m
            Measured Body TSL parameters                              (22.0 £0.2) °C    54.1 #6 %           1.50 mho/m « 6 %
            Body TSL temperature change during test                      <0.5 °C            ——


       SAR result with Body TSL

            SAR averaged over 1 cm* (1 g) of Body TSL                   Condition
             SAR measured                                       250 mW input power                     10.0 W/kg
             SAR for nominal Body TSL parameters                    normalized to 1W         40.5 Wikg + 17.0 % (k=2)

             SAR averaged over 10 cm‘ (10 g) of Body TSL                condition
             SAR measured                                       250 mW input power                     5.33 W/kg
             SAR for nominal Body TSL parameters                    normalized to 1W         21.5 Wikg * 16.5 % (k=2)



       Certificate No: D1900V2—5d101_Jul17                   Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 0108)

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                          51.7 Q + 5.8 jQ
      Return Loss                                                                      — 24.5 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                          46.2 Q + 6.6 jQ
      Return Loss                                                                      —22.0 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                  1,203 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductorof the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                   SPEAG
      Manufactured on                                                               March 28, 2008




Certificate No: D1900V2—5d101_Jul17                      Page 4 of 8


(IIEIl?
          DASY5 Validation Report for Head TSL

                                                                                            Date: 26.07.2017
          Test Laboratory: SPEAG, Zurich, Switzerland

          DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN:5d101

          Communication System: UID 0 — CW; Frequency: 1900 MHz
          Medium parameters used: £ = 1900 MHz; 0 = 1.39 S/m; & = 40.7; p = 1000 kg/m3
          Phantom section: Flat Section
          Measurement Standard: DASYS5 (IEEE/IEC/ANSI C€63.19—2011)

          DASY52 Configuration:

             e   Probe: EX3DV4 — SN7349; ConvF(8.43, 8.43, 8.43); Calibrated: 31.05.2017;

             e   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

             e   Electronics: DAE4 Sn601; Calibrated: 28.03.2017

             e   Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

             &   DASY32 52.10.0(1446); SEMCAD X 14.6.10(7417)


          Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
          Measurement grid: dx=5mm, dy=5mm, dz=5mm
          Reference Value = 106.3 V/m; Power Drift = —0.04 dB
          Peak SAR (extrapolated) = 18.4 W/kg
          SAR(1 g) = 9.93 W/kg; SAR(10 g) = 5.23 W/kg
          Maximum value of SAR (measured) = 14.9 W/kg




                 —3.60

                 —7.20

                  —10.80

                  ~14.40

                  —18.00
                             0 dB = 14.9 W/kg = 11.73 dBW/kg




          Certificate No: D1900V2—5d101_Jul17            Page 5 of 8


)
~
~
(I
     Impedance Measurement Plot for Head TSL




                                                                     26 Jul 2eir os:s2:00
              CH] sin      i uors          4t 51.727 a   5.8418 0 489.34 pH      1 900.000 oog MHz
                                                         ~~T




              Av
              16 *


              H1d

               CH2




              Ca




              Av
              16°

               H1d

                         START 1 700.000 @00 MHz                           STOP 2 100.000 000 MHz




     Certificate No: D1900V2—5d101_Jul17                   Page 6 of 8


(Ilflll,
           S
          DASY5 Validation Report for Body TSL



      Test Laboratory: SPEAG, Zurich, Switzerland
                                                                                         Date: 26.07.2017



      DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN:5d101

          Communication System: UID 0 — CW; Frequency: 1900 MHz
          Medium parameters used: f = 1900 MHz; 0 = 1.5 S/m; & = 54.1; p = 1000 kg/m*
          Phantom section: Flat Section
          Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)

          DASY52 Configuration:

             e   Probe: EX3DV4 — SN7349; ConvF(8.2, 8.2, 8.2); Calibrated: 31.05.2017;

             e   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

             e   Electronics: DAE4 Sn601; Calibrated: 28.03.2017

             e   Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

             e   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


          Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
          Measurement grid: dx=5mm, dy=5mm, dz=5mm
          Reference Value = 101.8 V/m; Power Drift = —0.09 dB
          Peak SAR (extrapolated) = 17.6 W/kg
          SAR(1 g) = 10 W/kg; SAR(10 g) = 5.33 W/kg
          Maximum value of SAR (measured) = 14.4 W/kg


                 dB
                 0

                 —3.40

                 —6.80

                  —10.20

                  —13.60

                  —17.00
                             0 dB = 14.4 W/kg = 11.58 dBW/kg




          Certificate No: D1900V2—5d101_Jul17            Page 7 of 8


@TTL


)
ue




 Impedance Measurement Plot for Body TSL




                                                               26 Jul 2017    11:15:02
          CHI sii      i uo ors        1: 46.164 o 6.5918 0 55247 pH         1 900.000 000 MHz
                                                 mpe




                     START 1 700.000 000 MHz                         STOP 2 100.000 000 MHz




 Certificate No: D1900V2—5d101_Jul17                 Page 8 of 8


«
Mn esw   en en en nge n n       ee >




           Calibration Laboratory of                                              wl n
                                                                              Sasy"/»                                  S       Schweizerischer Kalibrierdienst




                                                                         §                8
           Schmid & Partner




                                                                                          S
                                                                             4
                                                                                                                       C       Service suisse d‘étalonnage
             Engineering AG                                                                                                    Servizio svizzero di taratura




                                                                                        \>>
                                                                                        4”:,/
                                                                            i
           Zeughausstrasse 43, 8004 Zurich, Switzerland                                                                S       swiss Calibration Service
                                                                                 Aulutuls®

           Accredited by the Swiss Accreditation Service (SAS)                                                          Accreditation No.: SCS 0108
           The Swiss Accreditation Service is one of the signatories to the EA
           Multilateral Agreement for the recognition of calibration certificates

           Client       CTTL—BJ (Auden)                                                                    Certificate No: D2450V2—853_Jul17

           CALIBRATION CERTIFICATE
            Object                                D2450V2 — SN:853


            Calibration procedure(s)              QA CAL—05.v9
                                                  Calibration procedure for dipole validation kits above 700 MHz



            Calibration date:                     July 21, 2017

            This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
            The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


            All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity <70%.


            Calibration Equipment used (M&TE critical for calibration)


            Primary Standards                       ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
            Power meter NRP                         SN: 104778                04—Apr—17 (No. 217—02521/02522)                       Apr—18
            Power sensor NRP—Z91                    SN: 103244                04—Apr—17 (No. 217—02521)                             Apr—18
            Power sensor NRP—Z91                    SN: 103245                04—Apr—17 (No. 217—02522)                             Apr—18
            Reference 20 dB Attenuator              SN: 5058 (20k)            O7—Apr—17 (No. 217—02528)                             Apr—18
            Type—N mismatch combination             SN: 5047.2 / 06327        O7—Apr—17 (No. 217—02529)                             Apr—18
            Reference Probe EX3DV4                  SN: 7349                  31—May—17 (No. EX3—7349_May17)                        May—18
            DAE4                                    SN: 601                   28—Mar—17 (No. DAE4—601_Mar17)                        Mar—18


            Secondary Standards                     ID #                      Check Date (in house)                                 Scheduled Check
            Power meter EPM—442A                    SN: GB37480704            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
            Power sensor HP 8481A                   SN: US37202783            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
             Power sensor HP 8481A                  SN: MY41092317            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
             RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—16)                     In house check: Oct—18
            Network Analyzer HP 8758E               SN: US37390585            18—Oct—01 (in house check Oct—16)                     In house check: Oct—17


                                                    Name                                     Function                                Signature
            Calibrated by:                          Michael Weber                            Laboratory Technician
                                                                                                                                       e


             Approved by:                           KatJa Pokovic                            Technical Manager

                                                                                                                               DERE
                                                                                                                                    Issued: July 24, 2017
             This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


            Certificate No: D2450V2—853_Jul17                                      Page 1 of 8


    +      #                                                  uon
Callbratlon Laboratory of                                   i\‘\\w_&/’n}      S   Schweizerischer Kalibrierdienst
Schmid & Partner                                           M                  c   Service suisse d‘étalonnage
  Engineering AG                                           22—R                   Servizio svizzero di taratura
                              £       *                    y n
Zeughausstrasse 43, 8004 Zurich, Switzerland               *g/7/\\\\‘\\*      S   Swiss Calibration Service
                                                              thilals

Accredited by the Swiss Accreditation Service (SAS)                           Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
       (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
   d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
    e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   e Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
   e Feed Point Impedance and Return Loss: These parameters are measured with the dipole
          positioned under the liquid filled phantom. The impedance stated is transformed from the
          measurement at the SMA connector to the feed point. The Return Loss ensures low
          reflected power. No uncertainty required.
      e   Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
          No uncertainty required.
      e   SAR measured: SAR measured at the stated antenna input power.
      e   SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
          connector.
      e   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
          nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D2450V2—853_Jul17                               Page 2 of 8


                                                                                            i




Measurement Conditions
   DASYsystem configuration, as far as not given on page 1.
     DASY Version                                              DASY5                              V52.10.0
     Extrapolation                                   Advanced Extrapolation
     Phantom                                         Modular Flat Phantom
     Distance Dipole Center — TSL                              10 mm                             with Spacer
     Zoom Scan Resolution                              dx, dy, dz =5 mm
     Frequency                                         2450 MHz £ 1 MHz



Head TSL barameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C           39.2               1.80 mho/m
     Measured Head TSL parameters                              (22.0 + 0.2) °C   37.8 16 %          1.87 mho/m £ 6 %
     Head TSL temperature change during test                      <0.5 °C            ~——                     ———

SAR result with Head TSL

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
     SAR measured                                        250 mW input power                     13.4 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W         52.2 Wikg x 17.0 % (k=2)


     SAR averaged over 10 cm* (10 g) of Head TSL                 condition
     SAR measured                                        250 mW input power                     6.26 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W         24.7 Wikg 2 16.5 % (k=2)



Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Body TSL parameters                                  22.0 °C            §2.7              1.95 mho/m
     Measured Body TSL parameters                              (22.0 + 0.2) °C   52.1 26 %          2.04 mho/m x 6 %
      Body TSL temperature change during test                     <0.5 °C            ~——                     ————

SAR result with Body TSL

      SAR averaged over 1 cm* (1 g) of Body TSL                  Condition
      SAR measured                                       250 mW input power                     12.9 W/kg
      SAR for nominal Body TSL parameters                    normalized to 1W         50.4 W/kg x 17.0 % (k=2)

      SAR averaged over 10 cm* (10 g) of Body TSL                condition
      SAR measured                                       250 mW input power                  6.03 W/kg
      SAR for nominal Body TSL parameters                    normalized to 1W         23.8 Wikg + 16.5 % (k=2)



Certificate No: D2450V2—853_Jul17                     Page 3 of 8


(Elll’
         Appendix (Additional assessments outside the scope of SCS 0108)

         Antenna Parameters with Head TSL

               Impedance, transformed to feed point                                           52.0 Q + 5.0 jQ
               Return Loss                                                                       ~25.6 dB



         Antenna Parameters with Body TSL

               Impedance, transformed to feed point                                           49.6 Q + 6.3 jQ
               Return Loss                                                                      —24.0 dB




         General Antenna Parameters and Design

             | Electrical Delay (one direction)                     T                            1.161 ns

         After long term use with 100W radiated power, only a slight warming of the dipole nearthe feedpoint can be measured.

         The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
         second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
         are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
         "Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
         according to the Standard.
         No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
         feedpoint may be damaged.


         Additional EUT Data

               Manufactured by                                                                    SPEAG
               Manufactured on                                                              November 10, 2009




         Certificate No: D2450V2—853_Jul17                         Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                   Date: 20.07.2017

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 853

Communication System: UID 0 — CW; Frequency: 2450 MHz
Medium parameters used: f = 2450 MHz; 0 = 1.87 $/m; & = 37.8; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)
DASY52 Configuration:

   e    Probe: EX3DV4 — SN7349; ConvF(8.12, 8.12, 8.12); Calibrated: 31.05.2017;

   *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e    Electronics: DAE4 S$Sn601; Calibrated: 28.03.2017

   &    Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

   &    DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 112.7 V/m; Power Drift = —0.00 dB
Peak SAR (extrapolated) = 27.0 W/kg
SAR(1 g) = 13.4 W/kg; SAR(10 g) = 6.26 W/kg
Maximum value of SAR (measured) = 21.5 W/kg


       dB
       0

        —4.00

        —8.00

        —12.00

        —16.00

        —20.00
                   0 dB = 21.5 W/kg = 13.32 dBW/kg




Certificate No: D2450V2—853_Jul17                Page 5 of 8


®
Impedance Measurement Plot for Head TSL




                                                                 20 Jul 2017 11:54:15
         CHI sii      1 U FS          4151.965 a   4.9570 a   322.01 pH      2 450.000 000 MHz




          Del

          Ca




          Av
          16 &


          H1d

          CH2




          Ca




          Av
          16 *



          H1d

                    START 2 250.000 000 MHz                            STOP 2 650.000 000 MHz




Certificate No: D2450V2—853_Jul17                   Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                               Date: 21.07.2017

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 853

Communication System: UID 0 — CW; Frequency: 2450 MHz
Medium parameters used: f = 2450 MHz; 0 = 2.04 $/m; & = 52.1; p = 1000 kg/m‘
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C€63.19—2011)

DASY52 Configuration:

   e   Probe: EX3DV4 — SN7349; ConvF(8.1, 8.1, 8.1); Calibrated: 31.05.2017;

   e   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   *   Electronics: DAE4 S$Sn601; Calibrated: 28.03.2017

   *    Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

   *    DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 104.1 V/m; Power Drift = —0.09 dB
Peak SAR (extrapolated) = 25.5 W/kg
SAR(I g) = 12.9 W/kg; SAR(10 g) = 6.03 W/kg
Maximum value of SAR (measured) = 20.0 W/kg


       dB
        0
        —5.00
        —10.00
        —15.00
        —20.00
        —25.00
                   0 dB = 20.0 W/kg = 13.01 dBW/kg




Certificate No: D2450V2—853_Jul17               Page 7 of 8


®
Impedance Measurement Plot for Body TSL




                                                                  24 Jul 2017    14142134
         (CHZ sii     i U s           1149.619 a   6.2656 0 407.02 pH           2 450.000 000 MHz
                                               »   j 3       ~,




          Av
          16 *



          H1d

          CH2




          Ca




          Av
          16s


          H1d

                    START 2 250.000 000 MHz                             STOP 2 650.000 000 MHz




Certificate No: D2450V2—853_Jul17                   Page 8 of 8


                                                            No. I17Z61902-SEM01
                                                                 Page 111 of 124


ANNEX I      SPOT CHECK for 5026D

I.1 Conducted power of selected case
               Table I.1-1: The conducted power results for GSM850/1900
                                      Conducted Power (dBm)
  GSM
            Channel 251(848.8MHz)      Channel 190(836.6MHz)    Channel 128(824.2MHz)
 850MHz
                     32.68                     32.70                     32.71
                                      Conducted Power (dBm)
   GSM      Channel 810(1909.8MHz)     Channel 661(1880MHz)             Channel
 1900MHz                                                            512(1850.2MHz)
                     29.89                     30.02                     30.06

                  Table I.1-2: The conducted power results for GPRS
            GSM 850                             Measured Power (dBm)
          GPRS (GMSK)                  251                190                     128
            2 Txslots                  31.16            31.16                     31.70
            PCS1900                              Measured Power (dBm)
          GPRS (GMSK)                   810              661                       512
            2 Txslots                  28.33            28.49                     28.45

                      Table I.1-3: The conducted Power for WCDMA
                     band                             FDDV result
     Item
                    ARFCN         4233 (846.6MHz)  4182 (836.4MHz)       4132 (826.4MHz)
   WCDMA               \               22.82             22.93                22.84
                     band                             FDDII result
     Item
                    ARFCN        9538 (1907.6MHz)   9400 (1880MHz)       9262 (1852.4MHz)
   WCDMA               \               22.50             22.85                 22.95

                        Table I.1-5: The conducted Power for WLAN
    Mode / data rate                     Channel               Measured Power (dBm)
    802.11b – 1Mbps                      11                               14.65
    802.11b – 1Mbps                      6                                14.80
    802.11b – 1Mbps                      1                                14.74




                                                         ©Copyright. All rights reserved by CTTL.


                                                                                                    No. I17Z61902-SEM01
                                                                                                        Page 112 of 124

                I.2 Measurement results
                                                    Measured                      Measured Measured      Reported   Reported       Power
  Test Band     Channel    Frequency     Tune-Up                 Test Position                                                               Figure
                                                     Power                        10g SAR   1g SAR       10g SAR     1g SAR         Drift
    GSM850      CH251      848.8 MHz      34.5       32.68      Left Cheek         0.103       0.139     0.16        0.21          -0.05     Fig I.1
    GSM850      CH251      848.8 MHz      32.5       31.16          Rear            0.18       0.329     0.25        0.45          -0.07     Fig I.2
   PCS1900      CH512     1850.2 MHz       31        30.06      Right Cheek        0.049       0.081     0.06        0.10           0.02     Fig I.3
   PCS1900      CH512     1850.2 MHz      29.5       28.45      Bottom edge        0.443       0.847     0.56        1.08          -0.07     Fig I.4
WCDMA1900-BII   CH9262    1852.4 MHz       24        22.95      Right Cheek        0.096       0.159     0.12        0.20           0.09     Fig I.5
WCDMA1900-BII   CH9262    1852.4 MHz       24        22.95      Bottom edge        0.481       0.897     0.61        1.14          -0.02     Fig I.6
 WCDMA850-BV    CH4182     835.4 MHz      24.5       22.93      Left Cheek         0.121       0.164     0.17        0.24           0.03     Fig I.7
 WCDMA850-BV    CH4233     846.6 MHz      24.5       22.82          Rear            0.15       0.273     0.22        0.40          -0.01     Fig I.8
   WLAN2450       CH1       2412 MHz      15.5       14.74      Left Cheek         0.414        0.93     0.49        1.11            0.1     Fig I.9
   WLAN2450       CH6       2437 MHz      15.5       14.74         Front           0.054       0.104     0.06        0.12          -0.03    Fig I.10


                          Table I.2-1: SAR Values (WLAN - Head) – 802.11b (Scaled Reported SAR)
                                   Ambient Temperature: 22.5 oC                     Liquid Temperature: 23.3oC
                  Frequency                                                                        Reported
                                                  Test           Actual          maximum                             Scaled reported
                                       Side                                                          SAR
                 MHz       Ch.                   Position      duty factor       duty factor                         SAR (1g) (W/kg)
                                                                                                  (1g) (W/kg)
                 2412        1         Left        Touch        99.52%             100%                1.11                  1.12


                          Table I.2-2: SAR Values (WLAN - Body) – 802.11b (Scaled Reported SAR)
                                   Ambient Temperature: 22.5 oC                     Liquid Temperature: 23.3oC
                  Frequency              Test         Actual duty        maximum           Reported SAR             Scaled reported
                  MHz        Ch.        Position        factor           duty factor        (1g) (W/kg)             SAR (1g) (W/kg)
                 2437         6          Front         99.53%                100%                 0.12                      0.12




                                          CH1                                                          CH6




                                                                                                 ©Copyright. All rights reserved by CTTL.


                                                                   No. I17Z61902-SEM01
                                                                       Page 113 of 124

I.3 Reported SAR Comparison

                                                        Reported SAR            Reported SAR
  Exposure Configuration      Technology Band
                                                    1g (W/Kg): spot check     1g (W/Kg): original

                                  GSM 850                   0.21                      0.19
          Head                   PCS 1900                   0.10                      0.11
   (Separation Distance         UMTS FDD 2                  0.20                      0.20
          0mm)                  UMTS FDD 5                  0.24                      0.21
                               WLAN 2.4 GHz                 1.11                      1.27
                                  GSM 850                   0.45                      0.31
         Hotspot                 PCS 1900                   1.08                      0.93
   (Separation Distance         UMTS FDD 2                  1.14                      1.16
         10mm)                  UMTS FDD 5                  0.40                      0.28
                               WLAN 2.4 GHz                 0.12                      0.13

Note: All the spot check results marked blue are larger than the original result. So it replace
the original results and others are shared.




                                                              ©Copyright. All rights reserved by CTTL.



Document Created: 2017-11-28 17:53:21
Document Modified: 2017-11-28 17:53:21

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC