I17Z60860-SEM01_SAR_Rev0_part4

FCC ID: 2ACCJB095

RF Exposure Info

Download: PDF
FCCID_3470530

                                                                                                   fy—       <—0 —




Appendix (Additional assessments outside the scope of SCS 0108)

Antenna Parameters with Head TSL

      Impedance,transformed to feed point                                           sosn+55jn
      Retum Loss                                                                      ~25.1 dB


Antenna Parameters with Body TSL

      Impedane,transformed to feed point                                            4700 +670
      Retum Loss                                                                       ~224 0B



General Antenna Parameters and Design

     [ Electrical Detay (one direction}                                                 1.202 ns
After long term use with 100W radiated power, only a sight warming ofthe dipole near the feedpoint can be measured.
The dipole is made of standard semirigid coaxial cable. The center conductor ofthe feeding line is direcily connected to the
second arm of the dipole. The antenna is therefore shot—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
*Measurement Condiions® paragraph. The SAR data are not affected by this change. The overall dipole length is sti
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the saldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                   Speaq
      Manufactured on                                                               March 28, 2008




Cortficate No: D1900V2—5d101_.Jult6                      Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                   Date: 22.07.2016

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1900 MHz D1900V2; Type: D1900V2; Serial: D1900V2 — SN:5d101

Communication System: UID 0 — CW; Frequency: 1900 MHz
Mediumparameters used: f= 1900 MHz; 0 = 1.38 S/m; & = 39.8; p = 1000 ke/m}
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEEAEC/ANST C63.19—2011)

DASY52 Configuration:

   *    Probe: EX3DV4 — SN7349; ConvB(7.99, 7.99, 7.99); Calibrated: 15.06.2016;

   *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   +    Flectronics: DAF4 Sn601; Calibrated: 30.12.2015

   *    Phantom: Flat Phantom 5.0 (front); Type: QDOOOP5OAA; Serial: 1001

   «_   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7372)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x8x7)/Cube 0:
Measurementgrid: dx=3mm, dy=5mm, dz=5mm
Reference Value = 107.9 V/m; Power Drift = —0.02 dB
Peak SAR (extrapolated) = 19.0 W/kg
SAR(L g) = 10.1 W/kg; SAR(IO g) = 5.28 W/ke
Maximum value of SAR (measured) = 15.8 W/kg




        3.80

        —7.60

        1.40

        15.20

        —19.00
                  0 dB = 15.8 Wikg = 11.99 dBW/ke




Certficate No: D1900V2—5d101_Jult6             Page 5 of 8


                                                                                   fy—   <—0 —   —




Impedance Measurement Plot for Head TSL




                                                        22 aat soss ssroorse
          EBD sa      i ucss          isaese _ ssssae
                                               2okges seaarpi      1 900.000 000 mtz




          hy
          152


          ma
          one



          ca




                    Srhet 1 re0.000 a00 mz                      srom 2 100.000 ooo mz




Cortficate No: D1900V2.5d101_.Jult8               Page 8 of 8


DASY5 Validation Report for Body TSL
                                                                                  Date: 28.07.2016

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1900 MHz D1900V2; Type: D1900V2; Serial: D1900V2 — SN: 54101

Communication System: UID 0 — CW; Frequency: 1900 MHz
Medium parameters used: (= 1900 MHz; 0 = 1.31 S/m; & = 52.7; p = 1000 kg/m?
Phantom section: Flat Section
Measurement Standard: DASY3 (IEEE/AEC/ANST C63.19—2011)

DASY52 Configuration:
   *   Probe: EX3DV4 — SN7349; ConvB(8.03, 8.03, 8.03); Calibrated: 15.06.2016;

   + Sensor—Surface: 1.4mm (Mechanical Surface Detection)
   *   Electronics: DAE4 Sn601: Calibrated: 30.12.2015

   *   Phantom: Flat Phantom 5.0 (back); Type: QDOOOPSOAA; Serial: 1002

   *   DASY32 52.8.8(1258); SEMCAD X 14.6.10(7372)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurementgrid: dx=5mm, dy=Smm, dz=Smm
Reference Value = 104.2 V/m; Power Drift=—0.03 dB
Peak SAR (extrapolated) = 17.6 W/kg
SAR(L g) = 10 W/kg; SAR(1O g) = 5.32 We
Maximum value ofSAR (measured) = 15.0 W/ke




       3.33

       5.67

       +10.00

       413.34

       +16.67
                  0 dB = 15.0 W/kg = 11.76 dBW/kg




Certficate No: D1900V2—5d101_Jul16            Page 7 of 8


Impedance Measurement Plot for Body TSL



                                                          2e aur zes coirore«
          gBD s       a urs          assrse   erszee seasspi         1 se0.000 c00 nz
                                              .




          ca




                   Stae1 700.000 o0 mz                         srom 2 Los.coo onomiz




Certficate No: D1900V2—6d101_Julté            Fage 8 o8


Calibration
     a      Laboratory of                                  «us
                                                          al\/n                                & Sctweizeriacher Kallbrlerdtanst
Schmid & Partner                                          ;fi/&                                 _ Service suisse d‘étaionnage
   Engineering AG                                         ie t                                    Servirio      svizzero ditaratura
Zeughausstrasse 43, 8004 zurich, Suitzerland              x,’fl@?                               5. swiss Caliboration Service

Aceredted by e Sviss Accrodition Sorvice (GAS)                                                 Accreditation No: SCS 0108
"The Siiss Accreditaion Serviceis one of the signatories to the EA
MulllateralAgreementfor therecognition ofcabration corticates
Ciem       CTTL—BJ (Auden)                                                          Conificate No: D2450V2—853_Jul16

|CALIBRATION CERTIFICATE
 Object                           D2450V2 — SN:853


 Calbrationprocedure(e)           QA CAL—05.v9
                                  Galibration procedure for dipole validation kits above 700 MHz



 Calbraton date:                  July 25, 2016


 Thiscalbraion corifcate documents the traceabity t national standards, which realze the physicalunis of moasuromants (S)
 The measurements and the uncerainties wih confidence probabity are given on thefalowing pages and are pat ofthe cortcate

 Allcalbrations have beenconducted in the clased laboratoryfaclty: anvronment tomporature (22 2 3)°C and hunily <70%

 Calbration Equipment used (MATE cricalfr caltration}
 Primary Standards                 n#                     Gal Date (Contfcate No)                        scheduied Cattration
 Power mater NrP                   sic roure              d6—0r18 (No. 217—czzsa02200)                   Aoctr
 Power sensor NrP—291              SN: rosae              o6—A0r16 (No. 217.0z000)                       Aperr
 Power sensor NRP291               s ro0ss                d6—Apets (No. 217—0z000)                       Aperr
 Reference 20 B Attenuator         sn: sose rog           0—Apt8 (No. 217.02202)                         Aperr
 Type—N mismalch combination       SN: 8047.2/06027..     O6—Apre16 (No.217—02205)                       Aperr
 Relerence Probe EOVA              Sht row                15—Jun—16 (No. BB—7340.Junt)                   Juretr
 vace                              sN: co1                30—Dee—15 (No. DAEA—601.Dect)                  Dec1s
 Secondary Standards               o#                     Check Date in house)                           Scheduled Chack
 Power mater EPM—44zA              sN: aearigoros         0rOct1s No 217.0rz22)                          in house checic Oct—16
 Power sensor HP sagtA             sh: ussraserea         ar—0et1s (No.2t7—orze2)                        in house checkc Oct16
 Power sensor HP 8481A             Sh: invatosear?        or0et1s (No.217.02220)                         in house checic Oct16
 AF gonerator RAS SMT—46           stt roogre             16—Jun—15 (n house cheok dun—15)               in house check: Oct16
 Network Analyzer h sr5se          snt ussrasoses         160001 (inhousecheckOc+15)                     In house check: Oct16
                                   Name                              Funcion                              Signature
 Caltrated by:                     Michaal Wober                     Laboratoy Technician

 Approved by:                      Katia Potovic                      ‘Techrical Managor         %7g

                                                                                                        Issued: Juy 26, 20106
 "This caibraton certicate shanot be reproduced except in ful wihout writen approvalof t aboritoy,

Certficate No: D2450V2—853_Jut6                               Page 1 of 8


®
Calibration Laboratory of
Schmid & Partner
  Engineering AG
Zeughausstrasse 43, 8004 Zurich, Switzerland



Accredted by ho Siss Accredtation Sorvice (GAS)
                                                                                 Sehwelzerischer Kalibrardionst
                                                                                 Service suisse «‘étalonnage
                                                                                 Serviclo vizzero dtaratura
                                                                                 Swiss Callbration Service

                                                                              Accreditation No.: SCS 0108
The Siiss Accreditation Service is one ofthe signatories to the EA
MultLateral Agreoment for the recognition of allbration certficates
Glossary:
TSL                        tissue simulating liquid
ConvF                      sensitivity in TSL / NORM xy,z
NA                         not applicable or not measured

Calibration is Performed According to the Following Standards:
    a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
       Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
       Communications Devices: Measurement Techniques®, June 2013
    b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
        devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz),
       February 2005
    c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
       communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)*, March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
    e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   * Measurement Conditions: Further details are available from the Validation Report at the end
        of the certificate. All figures stated in the certificate are valid at the frequency indicated.
    *   Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
        point exacily below the center marking of the flat phantom section, with the arms oriented
        parallel to the body axis.
    *   Feed Point Impedance and Return Loss: These parameters are measured with the dipole
        positioned under the liquid filled phantom. The impedancestated is transformed from the
        measurement at the SMA connector to the feed point. The Return Loss ensures low
        reflected power. No uncertainty required.
    *   Electrical Delay: One—way delay between the SMA connector and the antennafeed point.
        No uncertainty required.
    *   SAR measured: SAR measured at the stated antenna input power.
        SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
        connector.
    +   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
        nominal SAR result.
 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.




Cortficate No: D2450V2.853.Jult6                             Page 2 of 8


Measurement Con
    DASY system contiquration, as far as not given on page 1.
     DASY Version                                             Dasvs                              vsess
     Extrapolation                                    Advanced Extrapolation
     Phantom                                           Modular Fiat Phantom
     Distance Dipole Center — TSL                              10 mm                            with Spacer
     Zoom Scan Resolution                                k. dy, dz =5 mm
     Frequency                                          2450 MHz + 1 Mz

Head TSL parameters
   The following parameters and calculations were applied.
                                                               Temporature      Permittivity         Conductivity
     Nominal Head TSL parametors                                 220°C              soe               1.80 mhoim
     Measured Head TSL parameters                              22002)°C          se026%            1.86 mhoim£6%
     Head TSL temperature change during tost                      «05°C

SAR result with Head TSL

     SAR averaged over 1 em(1 g) of Head TL                     Condition
     SAR measured                                        250 mW input power                13.5 Wig
     SAR for nominal Head TSL parameters                  normalized to 1W          52.0 Wikg 2 17.0 % (k=2)

     SAR averaged over 10 om* (10 g) of Head TSL                condition
     SAR measured                                        250 mW input power                623 Wikg
     SAR for nominal Head TSL paramoters                     normalized to 1W       24.6 Wikg 16.5 % (k=2)

Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temporature      Permittivity         Conductivity
     Nominal Body TSL parameters                                 220°0              se7               1.95 mhoim
     Measured Body TSL parameters                              @2002)°C          51.826%           2.03 mhoim=6%
     Body TSL temperature change during test                     <as°c              ——                    —
SAR result with Body TSL

     SAR averaged over 1 om" (1 g) of Body TSL                Condition
     SAR measured                                        250 mW input power                    18.1 Whg
     SAR for nominal Body TSL parameters                     normalized to 1W       51.2 Whka £ 17.0 % (kx2)
     SAR averaged over 10 em" (10 g) of Body TSL                 condition
     SAR measured                                        250 mW input power                8.10 Whg
     SAR for nominal Body TSL paramaters                     normalized to 1W       24.1 Wikg 16.5 % (ke2)



Cortficate No: D240V2—953_Jult6                       Page 3 of 8


G
                                                                                                  vgs> snn sc e e +




Appendix (Additional assessments outside the scope of SCS 0108)
Antenna Parameters with Head TSL

      Impedance, transformad to faed point                                           sern+s1jn
      Return Loss                                                                      +243 08

Antenna Parameters with Body TSL

      Impedance, transformedto feed point                                            soon+«sj0
      Return Loss                                                                      —27.0 08



General Antenna Parameters and Design

     [ Electrical Delay (one direction)                      |                          1.182ns
After long term use with 100W radiated power, only a sight warming of the dipole near the feedpoint can be measured.
The dipole is made of standard sermirigid coaxial cable. The center conductor ofthe feeding line is directly connocted to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipolas, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions* paragraph. The SAR data are not affected by this change. The overall dipole length is stil
according to the Standard.
No excessive force must be applied to the cipole arms, because they might bend or the soldered connections near the
feedpoint maybdamaged.

Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                              November 10, 2009




Certficate No: D240V2—853_Ju6                            Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                Date: 25.07.2016

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz D2450V2; Type: D2450V2; Serial: D2450V2 — SN:853

Communication System: UID 0 — CW; Frequency: 2450 MHz
Mediumparameters used: f= 2450 MHz; a = 1.86 $/m; e, = 38; p = 1000 kg/l“‘
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/EC/ANST C63.19—2011)

DASY52 Configuration:
   + Probe: EX3DV4 — SN7349; ConvB(7.72. 7.72, 7.72); Calibrated: 15.06.2016;
   + Sensor—Surface: 1.4mm (Mechanical Surface Detection)
   + Electronics: DAE4 Sn601; Calibrated: 30.12.2015
   + Phantom: Flat Phantom 5.0 (front); Type: QDOOOPSOAA; Serial: 1001
   «_   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7372)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurementgrid: dx=3mm, dy=5mm, d
Reference Value      15.0 V/m; Power Dri
Peak SAR (extrapolated) = 27.5 W/kg
SAR( g)       5 Wi/kg; SAR(I g) = 6,23 Wike
Maximum value of SAR (measured) = 22.2 W/ke




        ~4.60

        —9.20

        —13.80

        +18.40

        —23.00
                  0 dB =22.2 Wke = 13.46 dBW/ke




Cortficate No: D2450V2.953_Jut6               Page 5 of 8


Impedance Measurement Plot for Head TSL




                                                             as w zes oerssias
         EBD san      a ucre         1: s3.e00 a          szo            2 430.000 ooo nz




         he
         155




         onz



         to


         y
         1.53


         ie
                   "Staer 2 250.000 a00 mz                        stoe 2 ss0.000 a00 hiz




Cortficate No: D240V2—853_Juf16                    Page 6 of 6


DASY5 Validation Report for Body TSL

                                                                                   Date: 25.07.2016

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz D2450V2; Type: D2450V2; Serial: D2450V2 — SN:853

Communication System: UID 0 — CW; Frequency: 2450 MHz
Medium parameters used: f= 2450 MHz; a=2.03 $/m; ¢, = 51.8; p = 1000 kg/m‘
Phantomsection: Flat Section
Measurement Standard: DASY5 (IEEEAEC/ANSI C63.19—2011)
DASY52 Configuration:

    *   Probe: EX3DV4 — SN7349; ConvB(7.79, 7.79, 7.79); Calibrated: 15.06.2016;

    *   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    *   Electronics: DAE Sn601; Calibrated: 30.12.2015

    *   Phantom: Flat Phantom 5.0(back); Type: QDOOOPSOAA; Serial: 1002

    *   DASY52 52.8.8(1258); SEMCAD X 146.10(7372)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurementgrid: dx=3mm, dy=Smm, dz=Smm
Reference Value = 107.4 V/m; Power Drift = —0.02 dB
Peak SAR (extrapolated) = 26.3 W/kg
SAR(L g) = 13.1 W/ke; SAR(IO g)
Maximum value of SAR (measured




        ~4.39

        —8.78

        413.16

        17.55

        —21.94
                    0 dB =21.6 Wikg = 13.34 dBW/ke




Cortficate No: D2450V2.—853_.Jult6             Page 7 of 8


_
7TL                                                                             s—    —
~@m~




 Impedance Measurement Plot for Body TSL




                                                         2s au zo seroors
            in sa        i us           asaos o casssn 2sns pi    2 450.000 000 hkz




            con


            hy
            it°   t


            wha       [
                       srher 2 250.000 000mmz




  Cortficate No: D2450V2.—853_.Jult6               Page 8 of8


                                         No.I17Z60860-SEM01
                                             Page 119 of 119


ANNEX I   Accreditation Certificate




                                      ©Copyright. All rights reserved by CTTL.



Document Created: 2017-07-06 10:57:35
Document Modified: 2017-07-06 10:57:35

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC