Test Report HAC RF Emission Appendix A

FCC ID: 2ABZ2-GM1915

Test Report

Download: PDF
FCCID_4226313

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab        Date: 2019/2/21


HAC_E_Dipole_835_190221
DUT: HAC-Dipole 835 MHz

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.5 ℃

DASY5 Configuration:
- Probe: EF3DV3 - SN4047; ConvF(1, 1, 1) @ 835 MHz; Calibrated: 2019/1/30
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn854; Calibrated: 2018/6/14
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


E Scan - measurement distance from the probe sensor center to CD835 = 10mm &
15mm/Hearing Aid Compatibility Test at 15mm distance (41x361x1): Interpolated grid: dx=0.5000
mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 124.3 V/m; Power Drift = 0.02 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 115.1 V/m
Average value of Total=(114.3+115.1) / 2 = 114.7 V/m

                                           PMF scaled E-field
                                            Grid 1 M4 Grid 2 M4 Grid 3 M4
                                            113.4 V/m      114.3 V/m   109.0 V/m
                                            Grid 4 M4      Grid 5 M4   Grid 6 M4
                                            64.34 V/m      64.51 V/m   61.54 V/m
                                            Grid 7 M4      Grid 8 M4   Grid 9 M4
                                            113.2 V/m      115.1 V/m   111.5 V/m
                                                          Cursor:
                                                     Total = 115.1 V/m
                                                      E Category: M4
                                                  Location: 0.5, 74, 8.7 mm




                      0 dB = 115.1 V/m = 41.22 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab        Date: 2019/2/21


HAC_E_Dipole_1880_190221
DUT: HAC Dipole 1880 MHz

Communication System: CW; Frequency: 1880 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.5 ℃

DASY5 Configuration:
- Probe: EF3DV3 - SN4047; ConvF(1, 1, 1) @ 1880 MHz; Calibrated: 2019/1/30
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn854; Calibrated: 2018/6/14
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


E Scan - measurement distance from the probe sensor center to CD1880 = 10mm &
15mm/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid: dx=0.5000
mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 154.1 V/m; Power Drift = 0.03 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 94.34 V/m
Average value of Total=(94.34+93.71) / 2 = 94.025 V/m

                                           PMF scaled E-field
                                            Grid 1 M3 Grid 2 M3 Grid 3 M3
                                            93.35 V/m      94.34 V/m   91.18 V/m
                                            Grid 4 M3      Grid 5 M3   Grid 6 M3
                                            67.39 V/m      67.67 V/m   66.15 V/m
                                            Grid 7 M3      Grid 8 M3   Grid 9 M3
                                            92.84 V/m      93.71 V/m   90.30 V/m
                                                         Cursor:
                                                    Total = 94.34 V/m
                                                      E Category: M3
                                                Location: 0.5, -33.5, 8.7 mm




                      0 dB = 94.34 V/m = 39.49 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab       Date: 2019/2/25


HAC_E._Dipole_2450_190225
DUT: HAC Dipole 2450 MHz

Communication System: CW; Frequency : 2450 MHz;Duty Cycle: 1: 1
Medium: Air Medium parameters used: 0 = 0 S/m, s, = 1; p= 0 kg/m3
Ambient Temperature : 23.5 °C

DASYS5 Configuration:
— Probe: EF3DV3 — SN4047; ConvF(1, 1, 1) @ 2450 MHz; Calibrated: 2019/1/30
— Sensor—Surface: (Fix Surface)
— Electronics: DAF4 Sn8§54; Calibrated: 2018/6/14
— Phantom: HAC Test Arch with AMCC; Type: SD HAC POI BA;
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


E Scan — measurement distance from the probe sensor center to CD2450 =10mm & 15mm
2/Mearing Aid Compatibility Test at 1 5mm distance (41x181x1): Interpolated grid: dx=0.5000 mm,
dy=0.5000 mm
Device Reference Point: 0, 0, —6.3 mm
Reference Value = 82.49 V/m; Power Drift =0.01 dB
PMR not calibrated. PMF = 1.000 is applied.
E—field emissions = 92.85 V/m
Average value of Total=(91.81+92.85) / 2 = 92.33 V/m

                                           PMF scaled E—field

                                            Grid 1 M3 Grid 2 M3 Grid 3 M3
                                            90.96 V/m 91.81 V/m 88.05 V/m
                                            Grid 4 M3 Grid 5 M3 Grid 6 M3
                                            80.43 V/m 80.82 V/m 78.64 V/m
                                            Grid 7 M3 Grid 8 M3 Grid 9 M3
                                            91.65 V/m 92.85 V/m 90.38 V/m
                                                            Cursor:
                                                      Total = 92.85 V/m
                                                       E Category: M3
                                                 Location: 0.5, 24.5, 8.7 mm




          —1.42


          —2.85


          —4.27


          —5.70


          —7.12
                      —
                      0 dB = 92.85 V/m = 39.36 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab        Date: 2019/2/21


HAC_E_Dipole_2600_190221
DUT: HAC Dipole 2600 MHz

Communication System: CW ; Frequency: 2600 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.5 ℃

DASY5 Configuration:
- Probe: EF3DV3 - SN4047; ConvF(1, 1, 1) @ 2600 MHz; Calibrated: 2019/1/30
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn854; Calibrated: 2018/6/14
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


E Scan - measurement distance from the probe sensor center to CD2600 = 10mm &
15mm/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid: dx=0.5000
mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 73.21 V/m; Power Drift = -0.03 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 94.50 V/m
Average value of Total=(94.5+94.18) / 2 = 94.34 V/m

                                           PMF scaled E-field
                                            Grid 1 M3 Grid 2 M3 Grid 3 M3
                                            93.63 V/m      94.50 V/m   91.04 V/m
                                            Grid 4 M3      Grid 5 M3   Grid 6 M3
                                            86.26 V/m      86.66 V/m   84.23 V/m
                                            Grid 7 M3      Grid 8 M3   Grid 9 M3
                                            92.66 V/m      94.18 V/m   92.01 V/m
                                                          Cursor:
                                                     Total = 94.50 V/m
                                                       E Category: M3
                                                  Location: 1, -22.5, 8.7 mm




                      0 dB = 94.50 V/m = 39.51 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab        Date: 2019/2/25


HAC_E_Dipole_5500_190225
DUT: HAC Dipole 5500 MHz

Communication System: CW ; Frequency: 5500 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.5 ℃

DASY5 Configuration:
- Probe: EF3DV3 - SN4047; ConvF(1, 1, 1) @ 5500 MHz; Calibrated: 2019/1/30
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn854; Calibrated: 2018/6/14
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


E Scan - measurement distance from the probe sensor center to CD5500 = 10mm & 15mm
2/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid: dx=0.5000 mm,
dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 28.57 V/m; Power Drift = 0.04 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 129.8 V/m
Average value of Total=(105.6+109.2) / 2 = 107.4 V/m

                                               PMF scaled E-field
                                                Grid 1 M3 Grid 2 M3 Grid 3 M3
                                                103.3 V/m 105.6 V/m 103.5 V/m
                                                Grid 4 M2 Grid 5 M2 Grid 6 M2
                                                125.9 V/m 129.8 V/m 128.2 V/m
                                                Grid 7 M3 Grid 8 M3 Grid 9 M3
                                                106.7 V/m 109.2 V/m 107.4 V/m
                                                              Cursor:
                                                          Total = 129.8 V/m
                                                           E Category: M2
                                                      Location: -0.5, 3.5, 8.7 mm




                          0 dB = 129.8 V/m = 42.27 dBV/m