HAC_RF Emission_Appendix A

FCC ID: 2ABZ2-EE133

Test Report

Download: PDF
FCCID_4477749

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab         Date: 2019/7/19


HAC_E_Dipole_835
DUT: HAC-Dipole 835 MHz

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.5 ℃

DASY5 Configuration:
- Probe: EF3DV3 - SN4047; ConvF(1, 1, 1) @ 835 MHz; Calibrated: 2019/1/30
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn914; Calibrated: 2019/6/20
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.10 (2);SEMCAD X Version 14.6.12 (7470)


E Scan - measurement distance from the probe sensor center to CD835 = 10mm & 15mm/Hearing
Aid Compatibility Test at 15mm distance (41x361x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 132.1 V/m; Power Drift = -0.13 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 112.4 V/m
Average value of Total=(108.5+112.4) / 2 = 110.45 V/m

                                               PMF scaled E-field
                                                Grid 1 M4 Grid 2 M4 Grid 3 M4
                                                108.2 V/m         108.5 V/m   102.1 V/m
                                                Grid 4 M4         Grid 5 M4   Grid 6 M4
                                                61.06 V/m         61.11 V/m   59.54 V/m
                                                Grid 7 M4         Grid 8 M4   Grid 9 M4
                                                108.4 V/m         112.4 V/m   110.9 V/m
                                                              Cursor:
                                                         Total = 112.4 V/m
                                                           E Category: M4
                                                     Location: -0.5, 73.5, 9.7 mm




                          0 dB = 112.4 V/m = 41.02 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab         Date: 2019/7/19


HAC_E_Dipole_1880
DUT: HAC Dipole 1880 MHz

Communication System: CW; Frequency: 1880 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.5 ℃

DASY5 Configuration:
- Probe: EF3DV3 - SN4047; ConvF(1, 1, 1) @ 1880 MHz; Calibrated: 2019/1/30
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn914; Calibrated: 2019/6/20
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.10 (2);SEMCAD X Version 14.6.12 (7470)


E Scan - measurement distance from the probe sensor center to CD1880 = 10mm & 15mm/Hearing
Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 166.4 V/m; Power Drift = -0.02 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 90.15 V/m
Average value of Total=(86.62+90.15) / 2 = 88.39 V/m

                                               PMF scaled E-field
                                                Grid 1 M3 Grid 2 M3 Grid 3 M3
                                                86.40 V/m         86.62 V/m   82.90 V/m
                                                Grid 4 M3         Grid 5 M3   Grid 6 M3
                                                64.86 V/m         64.96 V/m   64.11 V/m
                                                Grid 7 M3         Grid 8 M3   Grid 9 M3
                                                88.14 V/m         90.15 V/m   88.05 V/m
                                                              Cursor:
                                                         Total = 90.15 V/m
                                                          E Category: M3
                                                      Location: 0, 33.5, 9.7 mm




                          0 dB = 90.15 V/m = 39.10 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab         Date: 2019.08.03

HAC_E_Dipole_1880

DUT: HAC-Dipole 1880 MHz

Communication System: CW; Frequency: 1880 MHz;Duty Cycle: 1:1

Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.4 ℃

DASY5 Configuration:
- Probe: EF3DV3 - SN4053 (5-6 GHz); ConvF(1, 1, 1); Calibrated: 2019.04.16;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn1437; Calibrated: 2018.10.15
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

E Scan - measurement distance from the probe sensor center to CD1880 = 10mm &
15mm/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid:
dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 169.6 V/m; Power Drift = -0.00 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 96.63 V/m
Average value of Total=(95.27+96.63)/2=95.95 V/m

                                   PMF scaled E-field
                                    Grid 1 M3 Grid 2 M3 Grid 3 M3
                                    93.33 V/m      95.27 V/m      93.23 V/m
                                    Grid 4 M3      Grid 5 M3      Grid 6 M3
                                    68.87 V/m      69.47 V/m      68.78 V/m
                                    Grid 7 M3      Grid 8 M3      Grid 9 M3
                                    94.88 V/m      96.63 V/m      94.75 V/m
                                                   Cursor:
                                              Total = 96.63 V/m
                                               E Category: M3
                                            Location: 0, 34, 9.7 mm


0 dB = 96.63 V/m = 39.85 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab         Date: 2019/7/19


HAC_E_Dipole_2600
DUT: HAC Dipole 2600 MHz

Communication System: CW ; Frequency: 2600 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.5 ℃

DASY5 Configuration:
- Probe: EF3DV3 - SN4047; ConvF(1, 1, 1) @ 2600 MHz; Calibrated: 2019/1/30
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn914; Calibrated: 2019/6/20
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.10 (2);SEMCAD X Version 14.6.12 (7470)


E Scan - measurement distance from the probe sensor center to CD2600 = 10mm & 15mm/Hearing
Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 72.66 V/m; Power Drift = -0.01 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 87.15 V/m
Average value of Total=(87.15+87.12) / 2 = 87.135 V/m

                                               PMF scaled E-field
                                                Grid 1 M3 Grid 2 M3 Grid 3 M3
                                                86.39 V/m         87.15 V/m   83.88 V/m
                                                Grid 4 M3         Grid 5 M3   Grid 6 M3
                                                80.58 V/m         81.00 V/m   78.70 V/m
                                                Grid 7 M3         Grid 8 M3   Grid 9 M3
                                                86.47 V/m         87.12 V/m   84.33 V/m
                                                              Cursor:
                                                         Total = 87.15 V/m
                                                           E Category: M3
                                                      Location: 1, -22.5, 9.7 mm




                          0 dB = 87.15 V/m = 38.81 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab         Date: 2019/7/20


HAC_E_Dipole_3500
DUT: HAC Dipole 3500 MHz

Communication System: CW ; Frequency: 3500 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.5 ℃

DASY5 Configuration:
- Probe: EF3DV3 - SN4047; ConvF(1, 1, 1) @ 3500 MHz; Calibrated: 2019/1/30
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn914; Calibrated: 2019/6/20
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.10 (2);SEMCAD X Version 14.6.12 (7470)


E Scan - measurement distance from the probe sensor center to CD3500 = 10mm & 15mm/Hearing
Aid Compatibility Test at 15mm distance (41x121x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 38.01 V/m; Power Drift = -0.08 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 88.15 V/m
Average value of Total=(88.15+87.0) / 2 = 87.575 V/m

                                               PMF scaled E-field
                                                Grid 1 M3 Grid 2 M3 Grid 3 M3
                                                87.43 V/m         88.15 V/m   85.22 V/m
                                                Grid 4 M3         Grid 5 M3   Grid 6 M3
                                                84.90 V/m         85.40 V/m   83.28 V/m
                                                Grid 7 M3         Grid 8 M3   Grid 9 M3
                                                87.00 V/m         87.00 V/m   83.96 V/m
                                                              Cursor:
                                                         Total = 88.15 V/m
                                                           E Category: M3
                                                     Location: 0.5, -14.5, 9.7 mm




                          0 dB = 88.15 V/m = 38.90 dBV/m



Document Created: 2019-10-24 21:45:28
Document Modified: 2019-10-24 21:45:28