SAR Calibration

FCC ID: 2ABV4A2WH

RF Exposure Info

Download: PDF
FCCID_2979678

   COMOSAR E-Field Probe Calibration Report




                                             Ref : ACR.348.1.15.SATU.A




 SHENZHEN STS TEST SERVICES CO., LTD.
1/F, BUILDING 2, ZHUOKE SCIENCE PARK, CHONGQING
                      ROAD
                         RICT, SHENZHEN,CHINA
 MVG COMOSAR DOSIMETRIC E-FIELD PROBE
                    SERIAL NO.: SN 45/15 EPGO281

                           Calibrated at MVG US
                 2105 Barrett Park Dr. - Kennesaw, GA 30144




                        Calibration Date: 12/10/2015

    Summary:
    This document presents the method and results from an accredited COMOSAR Dosimetric E-Field
    Probe calibration performed in MVG USA using the CALISAR / CALIBAIR test bench, for use
    with a COMOSAR system only. All calibration results are traceable to national metrology
    institutions.


                         COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.348.1.15.SATU.A




                                  Name                                Function                        Date                    Signature
Prepared by :                Jérôme LUC                         Product Manager                    12/14/2015

Checked by :                 Jérôme LUC                         Product Manager                    12/14/2015

Approved by :          Kim RUTKOWSKI                             Quality Manager                   12/14/2015




                                                                        Customer Name
                                                                      Shenzhen STS Test
                                       Distribution :
                                                                       Services Co., Ltd.




   Issue                  Date                                                            Modifications
     A                 12/14/2015                  Initial release




                                                               Page: 2/10

                This document shall not be reproduced, except in full or in part, without the written approval of MVG.
                The information contained herein is to be used only for the purpose for which it is submitted and is not to
                be released in whole or part without written approval of MVG.


                             COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.348.1.15.SATU.A




                                                      TABLE OF CONTENTS



1    Device Under Test ..................................................................................................... 4
2    Product Description ................................................................................................... 4
    2.1     General Information _______________________________________________________4
3    Measurement Method ................................................................................................ 4
    3.1     Linearity ________________________________________________________________4
    3.2     Sensitivity _______________________________________________________________5
    3.3     Lower Detection Limit _____________________________________________________5
    3.4     Isotropy _________________________________________________________________5
    3.5     Boundary Effect __________________________________________________________5
4    Measurement Uncertainty.......................................................................................... 5
5    Calibration Measurement Results .............................................................................. 6
    5.1     Sensitivity in air __________________________________________________________6
    5.2     Linearity ________________________________________________________________7
    5.3     Sensitivity in liquid ________________________________________________________7
    5.4     Isotropy _________________________________________________________________8
6    List of Equipment .................................................................................................... 10




                                                                   Page: 3/10

                    This document shall not be reproduced, except in full or in part, without the written approval of MVG.
                    The information contained herein is to be used only for the purpose for which it is submitted and is not to
                    be released in whole or part without written approval of MVG.


                          COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.348.1.15.SATU.A




1              DEVICE UNDER TEST

                                                      Device Under Test
      Device Type                                                     COMOSAR DOSIMETRIC E FIELD PROBE
      Manufacturer                                                    MVG
      Model                                                           SSE2
      Serial Number                                                   SN 45/15 EPGO281
      Product Condition (new / used)                                  New
      Frequency Range of Probe                                        0.45 GHz-6GHz
      Resistance of Three Dipoles at Connector                        Dipole 1: R1=0.186 M
                                                                      Dipole 2: R2=0.194 M
                                                                      Dipole 3: R3=0.191 M

A yearly calibration interval is recommended.

2              PRODUCT DESCRIPTION

2.1          GENERAL INFORMATION
MVG                                                                                                              OET 65 Bulletin C and
CEI/IEC 62209 standards.




                             Figure 1         MVG COMOSAR Dosimetric E field Dipole

      Probe Length                                                    330 mm
      Length of Individual Dipoles                                    2 mm
      Maximum external diameter                                       8 mm
      Probe Tip External Diameter                                     2.5 mm
      Distance between dipoles / probe extremity                      1 mm


         3     MEASUREMENT METHOD
The IEEE 1528, OET 65 Bulletin C, CENELEC EN50361 and CEI/IEC 62209 standards provide
recommended practices for the probe calibrations, including the performance characteristics of
interest and methods by which to assess their affect. All calibrations / measurements performed meet
the fore mentioned standards.

         3.1   LINEARITY
The evaluation of the linearity was done in free space using the waveguide, performing a power
sweep to cover the SAR range 0.01W/kg to 100W/kg.
                                                                Page: 4/10

                 This document shall not be reproduced, except in full or in part, without the written approval of MVG.
                 The information contained herein is to be used only for the purpose for which it is submitted and is not to
                 be released in whole or part without written approval of MVG.


                            COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.348.1.15.SATU.A




        3.2    SENSITIVITY
The sensitivity factors of the three dipoles were determined using a two step calibration method (air
and tissue simulating liquid) using waveguides as outlined in the standards.

        3.3    LOWER DETECTION LIMIT
The lower detection limit was assessed using the same measurement set up as used for the linearity
measurement. The required lower detection limit is 10 mW/kg.

        3.4    ISOTROPY
The axial isotropy was evaluated by exposing the probe to a reference wave from a standard dipole
with the dipole mounted under the flat phantom in the test configuration suggested for system
validations and checks. The probe was rotated along its main axis from 0 - 360 degrees in 15 degree
steps. The hemispherical isotropy is determined by inserting the probe in a thin plastic box filled
with tissue-equivalent liquid, with the plastic box illuminated with the fields from a half wave dipole.
The dipole is rotated about its axis (0° 180°) in 15° increments. At each step the probe is rotated
about its axis (0° 360°).

        3.5    BOUNDARY EFFECT
The boundary effect is defined as the deviation between the SAR measured data and the expected
exponential decay in the liquid when the probe is oriented normal to the interface. To evaluate this
effect, the liquid filled flat phantom is exposed to fields from either a reference dipole or waveguide.
With the probe normal to the phantom surface, the peak spatial average SAR is measured and
compared to the analytical value at the surface.

4             MEASUREMENT UNCERTAINTY

The guidelines outlined in the IEEE 1528, OET 65 Bulletin C, CENELEC EN50361 and CEI/IEC
62209 standards were followed to generate the measurement uncertainty associated with an E-field
probe calibration using the waveguide technique. All uncertainties listed below represent an
expanded uncertainty expressed at approximately the 95% confidence level using a coverage factor
of k=2, traceable to the Internationally Accepted Guides to Measurement Uncertainty.
 Uncertainty analysis of the probe calibration in waveguide
                                                  Uncertainty            Probability                                                Standard
        ERROR SOURCES                                                                             Divisor               ci
                                                   value (%)             Distribution                                            Uncertainty (%)
 Incident or forward power                            3.00%               Rectangular                   3               1             1.732%

 Reflected power                                      3.00%              Rectangular                    3               1             1.732%
                                                                                                   1.732051
 Liquid conductivity                                  5.00%               Rectangular                   3               1             2.887%
                                                                                                   1.732051
 Liquid permittivity                                  4.00%               Rectangular                   3               1             2.309%
                                                                                                   1.732051
 Field homogeneity                                    3.00%               Rectangular                   3               1             1.732%
                                                                                                   1.732051
 Field probe positioning                              5.00%              Rectangular                    3               1             2.887%
                                                                                                   1.732051

                                                                  Page: 5/10

                   This document shall not be reproduced, except in full or in part, without the written approval of MVG.
                   The information contained herein is to be used only for the purpose for which it is submitted and is not to
                   be released in whole or part without written approval of MVG.


                           COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.348.1.15.SATU.A




 Field probe linearity                               3.00%                Rectangular                  3               1             1.732%
                                                                                                  1.732051
  Combined standard uncertainty                                                                                                      5.831%
       Expanded uncertainty
                                                                                                                                     12.0%
     95 % confidence level k = 2



        5     CALIBRATION MEASUREMENT RESULTS


                                                   Calibration Parameters
   Liquid Temperature                                         21 °C
   Lab Temperature                                            21 °C
   Lab Humidity                                               45 %


        5.1     SENSITIVITY IN AIR


                                     Normx dipole Normy dipole Normz dipole
                                     1 ( V/(V/m)2) 2 ( V/(V/m)2) 3 ( V/(V/m)2)
                                          0.77          0.83          0.67


                                     DCP dipole 1 DCP dipole 2 DCP dipole 3
                                       (mV)         (mV)         (mV)
                                         91           90           95


Calibration curves ei=f(V) (i=1,2,3) allow to obtain H-field value using the formula:
                                                       E         E1       E2       E3
                                                                      2        2        2




                                                                 Page: 6/10

                  This document shall not be reproduced, except in full or in part, without the written approval of MVG.
                  The information contained herein is to be used only for the purpose for which it is submitted and is not to
                  be released in whole or part without written approval of MVG.


                      COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.348.1.15.SATU.A




      5.2   LINEARITY




      5.3   SENSITIVITY IN LIQUID

  Liquid        Frequency             Permittivity         Epsilon (S/m)              ConvF
                 (MHz +/-
                100MHz)
  HL450            450                    44.12                  0.88                   1.76
  BL450            450                    58.92                  1.00                   1.81
  HL750            750                    42.24                  0.90                   1.53
  BL750            750                    56.85                  0.99                   1.59
  HL850            835                    43.02                  0.90                   1.78
  BL850            835                    53.72                  0.98                   1.85
  HL900            900                    42.47                  0.99                   1.62
  BL900            900                    56.97                  1.09                   1.67
  HL1800          1800                    42.24                  1.40                   1.83
  BL1800          1800                    53.53                  1.53                   1.87
  HL1900          1900                    40.79                  1.42                   2.10
  BL1900          1900                    54.47                  1.57                   2.16
  HL2000          2000                    40.52                  1.44                   2.01
  BL2000          2000                    54.18                  1.56                   2.09
  HL2450          2450                    38.73                  1.81                   2.21
  BL2450          2450                    53.23                  1.96                   2.28
  HL2600          2600                    38.54                  1.95                   2.32
  BL2600          2600                    52.07                  2.23                   2.38
  HL5200          5200                    36.80                  4.84                   2.46
  BL5200          5200                    51.21                  5.16                   2.52
  HL5400          5400                    36.35                  4.96                   2.70
  BL5400          5400                    50.51                  5.70                   2.79
  HL5600          5600                    35.57                  5.23                   2.74
  BL5600          5600                    49.83                  5.91                   2.83
  HL5800          5800                    35.30                  5.47                   2.53
  BL5800          5800                    49.03                  6.28                   2.60

LOWER DETECTION LIMIT: 9mW/kg


                                                            Page: 7/10

             This document shall not be reproduced, except in full or in part, without the written approval of MVG.
             The information contained herein is to be used only for the purpose for which it is submitted and is not to
             be released in whole or part without written approval of MVG.


                       COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.348.1.15.SATU.A




       5.4   ISOTROPY
HL900 MHz
- Axial isotropy:                              0.04 dB
- Hemispherical isotropy:                      0.06 dB




HL1800 MHz
- Axial isotropy:                              0.04 dB
- Hemispherical isotropy:                      0.08 dB




                                                             Page: 8/10

              This document shall not be reproduced, except in full or in part, without the written approval of MVG.
              The information contained herein is to be used only for the purpose for which it is submitted and is not to
              be released in whole or part without written approval of MVG.


                       COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.348.1.15.SATU.A




HL5600 MHz
- Axial isotropy:                              0.06 dB
- Hemispherical isotropy:                      0.08 dB




                                                             Page: 9/10

              This document shall not be reproduced, except in full or in part, without the written approval of MVG.
              The information contained herein is to be used only for the purpose for which it is submitted and is not to
              be released in whole or part without written approval of MVG.


                            COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.348.1.15.SATU.A




        6    LIST OF EQUIPMENT


                                                  Equipment Summary Sheet

    Equipment                  Manufacturer /                        Current                                               Next Calibration
                                              Identification No.
    Description                   Model                          Calibration Date                                                Date
                                                                                        Validated. No cal               Validated.       No cal
    Flat Phantom                       MVG                   SN-20/09-SAM71
                                                                                        required.                       required.
                                                                                        Validated. No cal               Validated.       No cal
COMOSAR Test Bench                  Version 3                          NA
                                                                                        required.                       required.
                              Rhode & Schwarz
  Network Analyzer                                                SN100132                         02/2013                       02/2016
                                   ZVA

   Reference Probe                     MVG                    EP 94 SN 37/08                       10/2015                       10/2016

      Multimeter                 Keithley 2000                     1188656                         12/2013                       12/2016

   Signal Generator             Agilent E4438C                  MY49070581                         12/2013                       12/2016

                                                                                         Characterized prior to Characterized prior to
       Amplifier                  Aethercomm                        SN 046
                                                                                         test. No cal required. test. No cal required.

     Power Meter                   HP E4418A                    US38261498                         12/2013                       12/2016

    Power Sensor                 HP ECP-E26A                    US37181460                         12/2013                       12/2016

                                                                                         Characterized prior to Characterized prior to
  Directional Coupler           Narda 4216-20                        01386
                                                                                         test. No cal required. test. No cal required.
                                                                                        Validated. No cal               Validated. No cal
     Waveguide                  Mega Industries             069Y7-158-13-712
                                                                                        required.                       required.
                                                                                        Validated. No cal               Validated. No cal
Waveguide Transition            Mega Industries             069Y7-158-13-701
                                                                                        required.                       required.
                                                                                        Validated. No cal               Validated. No cal
Waveguide Termination           Mega Industries             069Y7-158-13-701
                                                                                        required.                       required.
Temperature / Humidity
                       Control Company                           150798832                         10/2015                       10/2017
      Sensor




                                                                 Page: 10/10

                   This document shall not be reproduced, except in full or in part, without the written approval of MVG.
                   The information contained herein is to be used only for the purpose for which it is submitted and is not to
                   be released in whole or part without written approval of MVG.


                                   Ref : ACR.262.5.14.SATU.A




Summary:

This document presents the method and results from an accredited SAR reference dipole calibration
performed in SATIMO USA using the COMOSAR test bench. All calibration results are traceable
to national metrology institutions.


                                                                                                                     Ref: ACR.262.5.14.SATU.A




                                  Name                                Function                        Date                    Signature
Prepared by :                Jérôme LUC                         Product Manager                    9/19/2014

Checked by :                 Jérôme LUC                         Product Manager                    9/19/2014

Approved by :          Kim RUTKOWSKI                            Quality Manager                    9/19/2014




                                                                        Customer Name
                                                                      Shenzhen STS Test
                                       Distribution :
                                                                       Services Co., Ltd.




   Issue                   Date                                                           Modifications
     A                  9/19/2014                  Initial release




                                                               Page: 2/11

                This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                The information contained herein is to be used only for the purpose for which it is submitted and is not to
                be released in whole or part without written approval of SATIMO.


                                                                                                                         Ref: ACR.262.5.14.SATU.A




1    Introduction................................................................................................................4
2    Device Under Test .....................................................................................................4
3    Product Description ...................................................................................................4
    3.1      General Information _______________________________________________________4
4    Measurement Method ................................................................................................5
    4.1      Return Loss Requirements __________________________________________________5
    4.2      Mechanical Requirements___________________________________________________5
5    Measurement Uncertainty..........................................................................................5
    5.1      Return Loss ______________________________________________________________5
    5.2      Dimension Measurement ___________________________________________________5
    5.3      Validation Measurement ____________________________________________________5
6    Calibration Measurement Results ..............................................................................6
    6.1      Return Loss and Impedance In Head Liquid ____________________________________6
    6.2      Return Loss and Impedance In Body Liquid ____________________________________6
    6.3      Mechanical Dimensions ____________________________________________________6
7    Validation measurement ............................................................................................7
    7.1      Head Liquid Measurement __________________________________________________7
    7.2      SAR Measurement Result With Head Liquid____________________________________8
    7.3      Body Liquid Measurement __________________________________________________9
    7.4      SAR Measurement Result With Body Liquid __________________________________10
8    List of Equipment ....................................................................................................11




                                                                   Page: 3/11

                    This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                    The information contained herein is to be used only for the purpose for which it is submitted and is not to
                    be released in whole or part without written approval of SATIMO.


                                                                                                                   Ref: ACR.262.5.14.SATU.A




This document contains a summary of the requirements set forth by the IEEE 1528, OET 65 Bulletin
C and CEI/IEC 62209 standards for reference dipoles used for SAR measurement system validations
and the measurements that were performed to verify that the product complies with the fore
mentioned standards.




                                                   Device Under Test
        Device Type                                       COMOSAR 835 MHz REFERENCE DIPOLE
        Manufacturer                                      Satimo
        Model                                             SID835
        Serial Number                                     SN 30/14 DIP0G835-332
        Product Condition (new / used)                    New

A yearly calibration interval is recommended.



3.1     GENERAL INFORMATION
                                                                           , OET 65 Bulletin
C and CEI/IEC 62209 standards. The product is designed for use with the COMOSAR test bench
only.




                                                Satimo COMOSAR Validation Dipole



                                                             Page: 4/11

              This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
              The information contained herein is to be used only for the purpose for which it is submitted and is not to
              be released in whole or part without written approval of SATIMO.


                                                                                                                   Ref: ACR.262.5.14.SATU.A




The IEEE 1528, OET 65 Bulletin C and CEI/IEC 62209 standards provide requirements for
reference dipoles used for system validation measurements. The following measurements were
performed to verify that the product complies with the fore mentioned standards.

       4.1   RETURN LOSS REQUIREMENTS
The dipole used for SAR system validation measurements and checks must have a return loss of -20
dB or better. The return loss measurement shall be performed against a liquid filled flat phantom,
with the phantom constucted as outlined in the fore mentioned standards.

       4.2   MECHANICAL REQUIREMENTS
The IEEE Std. 1528 and CEI/IEC 62209 standards specify the mechanical components and
dimensions of the validation dipoles, with the dimensions frequency and phantom shell thickness
dependent. The COMOSAR test bench employs a 2 mm phantom shell thickness therefore the
dipoles sold for use with the COMOSAR test bench comply with the requirements set forth for a 2
mm phantom shell thickness.




All uncertainties listed below represent an expanded uncertainty expressed at approximately the 95%
confidence level using a coverage factor of k=2, traceable to the Internationally Accepted Guides to
Measurement Uncertainty.

       5.1   RETURN LOSS
The following uncertainties apply to the return loss measurement:


                      400-6000MHz                                                                0.1 dB

       5.2   DIMENSION MEASUREMENT
The following uncertainties apply to the dimension measurements:


                            3 - 300                                                             0.05 mm


       5.3   VALIDATION MEASUREMENT
The guidelines outlined in the IEEE 1528, OET 65 Bulletin C, CENELEC EN50361 and CEI/IEC
62209 standards were followed to generate the measurement uncertainty for validation
measurements.

                                1g                                                     20.3 %
                               10 g                                                    20.1 %

                                                             Page: 5/11

              This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
              The information contained herein is to be used only for the purpose for which it is submitted and is not to
              be released in whole or part without written approval of SATIMO.


                                                                                                                       Ref: ACR.262.5.14.SATU.A




      6.1     RETURN LOSS AND IMPEDANCE IN HEAD LIQUID




            835                                  -28.11                                     -20                          51.6     + 3.6 j

      6.2     RETURN LOSS AND IMPEDANCE IN BODY LIQUID




            835                                  -23.87                                     -20                          49.0     + 6.3 j

      6.3     MECHANICAL DIMENSIONS
Frequency MHz                            L mm                                         h mm                                      d mm

                            required              measured                required             measured             required           measured

    300                   420.0 ±1 %.                                   250.0 ±1 %.                                6.35 ±1 %.
    450                   290.0 ±1 %.                                   166.7 ±1 %.                                6.35 ±1 %.
    750                   176.0 ±1 %.                                   100.0 ±1 %.                                6.35 ±1 %.
    835                   161.0 ±1 %.                PASS                89.8 ±1 %.               PASS              3.6 ±1 %.            PASS


                                                                 Page: 6/11

                  This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                  The information contained herein is to be used only for the purpose for which it is submitted and is not to
                  be released in whole or part without written approval of SATIMO.


                                                                                                                   Ref: ACR.262.5.14.SATU.A




      900             149.0 ±1 %.                                    83.3 ±1 %.                                  3.6 ±1 %.
     1450              89.1 ±1 %.                                    51.7 ±1 %.                                  3.6 ±1 %.
     1500              80.5 ±1 %.                                    50.0 ±1 %.                                  3.6 ±1 %.
     1640              79.0 ±1 %.                                    45.7 ±1 %.                                  3.6 ±1 %.
     1750              75.2 ±1 %.                                    42.9 ±1 %.                                  3.6 ±1 %.
     1800              72.0 ±1 %.                                    41.7 ±1 %.                                  3.6 ±1 %.
     1900              68.0 ±1 %.                                    39.5 ±1 %.                                  3.6 ±1 %.
     1950              66.3 ±1 %.                                    38.5 ±1 %.                                  3.6 ±1 %.
     2000              64.5 ±1 %.                                    37.5 ±1 %.                                  3.6 ±1 %.
     2100              61.0 ±1 %.                                    35.7 ±1 %.                                  3.6 ±1 %.
     2300              55.5 ±1 %.                                    32.6 ±1 %.                                  3.6 ±1 %.
     2450              51.5 ±1 %.                                    30.4 ±1 %.                                  3.6 ±1 %.
     2600              48.5 ±1 %.                                    28.8 ±1 %.                                  3.6 ±1 %.
     3000              41.5 ±1 %.                                    25.0 ±1 %.                                  3.6 ±1 %.
     3500              37.0±1 %.                                     26.4 ±1 %.                                  3.6 ±1 %.
     3700              34.7±1 %.                                     26.4 ±1 %.                                  3.6 ±1 %.




The IEEE Std. 1528, OET 65 Bulletin C and CEI/IEC 62209 standards state that the system
validation measurements must be performed using a reference dipole meeting the fore mentioned
return loss and mechanical dimension requirements. The validation measurement must be performed
against a liquid filled flat phantom, with the phantom constructed as outlined in the fore mentioned
standards. Per the standards, the dipole shall be positioned below the bottom of the phantom, with
the dipole length centered and parallel to the longest dimension of the flat phantom, with the top
surface of the dipole at the described distance from the bottom surface of the phantom.

       7.1   HEAD LIQUID MEASUREMENT
                       Frequency
                                              Relative permittivity (                  Conductivity ( ) S/m
                          MHz                                            r

                                              required           measured            required          measured
                           300               45.3 ±5 %                               0.87 ±5 %
                           450               43.5 ±5 %                               0.87 ±5 %
                           750               41.9 ±5 %                               0.89 ±5 %
                           835               41.5 ±5 %              PASS             0.90 ±5 %            PASS
                           900               41.5 ±5 %                               0.97 ±5 %
                          1450               40.5 ±5 %                               1.20 ±5 %
                          1500               40.4 ±5 %                               1.23 ±5 %
                          1640               40.2 ±5 %                               1.31 ±5 %
                          1750               40.1 ±5 %                               1.37 ±5 %
                          1800               40.0 ±5 %                               1.40 ±5 %
                          1900               40.0 ±5 %                               1.40 ±5 %
                          1950               40.0 ±5 %                               1.40 ±5 %
                          2000               40.0 ±5 %                               1.40 ±5 %


                                                             Page: 7/11

              This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
              The information contained herein is to be used only for the purpose for which it is submitted and is not to
              be released in whole or part without written approval of SATIMO.


                                                                                                                       Ref: ACR.262.5.14.SATU.A




                             2100               39.8 ±5 %                               1.49 ±5 %
                             2300               39.5 ±5 %                               1.67 ±5 %
                             2450               39.2 ±5 %                               1.80 ±5 %
                             2600               39.0 ±5 %                               1.96 ±5 %
                             3000               38.5 ±5 %                               2.40 ±5 %
                             3500               37.9 ±5 %                               2.91 ±5 %


        7.2    SAR MEASUREMENT RESULT WITH HEAD LIQUID
The IEEE Std. 1528 and CEI/IEC 62209 standards state that the system validation measurements
should produce the SAR values shown below (for phantom thickness of 2 mm), within the
uncertainty for the system validation. All SAR values are normalized to 1 W forward power. In
bracket, the measured SAR is given with the used input power.

Software                                                                OPENSAR V4
Phantom                                                                 SN 20/09 SAM71
Probe                                                                   SN 18/11 EPG122
Liquid                                                                                                     42.3 sigma : 0.92
Distance between dipole center and liquid                               15.0 mm
Area scan resolution                                                    dx=8mm/dy=8mm
Zoon Scan Resolution                                                    dx=8mm/dy=8m/dz=5mm
Frequency                                                               835 MHz
Input power                                                             20 dBm
Liquid Temperature                                                      21 °C
Lab Temperature                                                         21 °C
Lab Humidity                                                            45 %




                          Frequency
                                                     1 g SAR (W/kg/W)                      10 g SAR (W/kg/W)
                             MHz
                                                 required           measured            required          measured
                              300                   2.85                                   1.94
                              450                  4.58                                    3.06
                              750                   8.49                                   5.55
                              835                  9.56             9.63 (0.96)            6.22          6.15 (0.62)
                              900                  10.9                                    6.99
                             1450                    29                                     16
                             1500                   30.5                                   16.8
                             1640                   34.2                                   18.4
                             1750                   36.4                                   19.3
                             1800                   38.4                                   20.1
                             1900                   39.7                                   20.5
                             1950                   40.5                                   20.9
                             2000                   41.1                                   21.1
                             2100                   43.6                                   21.9
                             2300                   48.7                                   23.3


                                                                Page: 8/11

                 This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                 The information contained herein is to be used only for the purpose for which it is submitted and is not to
                 be released in whole or part without written approval of SATIMO.


                                                                                                            Ref: ACR.262.5.14.SATU.A




                   2450                   52.4                                    24
                   2600                   55.3                                   24.6
                   3000                   63.8                                   25.7
                   3500                   67.1                                    25




7.3   BODY LIQUID MEASUREMENT
                Frequency
                                       Relative permittivity (                  Conductivity ( ) S/m
                   MHz                                            r

                                       required           measured            required          measured
                   150                61.9 ±5 %                               0.80 ±5 %
                   300                58.2 ±5 %                               0.92 ±5 %
                   450                56.7 ±5 %                               0.94 ±5 %
                   750                55.5 ±5 %                               0.96 ±5 %
                   835                55.2 ±5 %              PASS             0.97 ±5 %            PASS
                   900                55.0 ±5 %                               1.05 ±5 %
                   915                55.0 ±5 %                               1.06 ±5 %
                   1450               54.0 ±5 %                               1.30 ±5 %
                   1610               53.8 ±5 %                               1.40 ±5 %
                   1800               53.3 ±5 %                               1.52 ±5 %
                   1900               53.3 ±5 %                               1.52 ±5 %
                   2000               53.3 ±5 %                               1.52 ±5 %
                   2100               53.2 ±5 %                               1.62 ±5 %
                   2450               52.7 ±5 %                               1.95 ±5 %
                   2600               52.5 ±5 %                               2.16 ±5 %
                   3000               52.0 ±5 %                               2.73 ±5 %
                   3500               51.3 ±5 %                               3.31 ±5 %
                   5200               49.0 ±10 %                             5.30 ±10 %
                   5300               48.9 ±10 %                             5.42 ±10 %
                   5400               48.7 ±10 %                             5.53 ±10 %


                                                      Page: 9/11

       This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
       The information contained herein is to be used only for the purpose for which it is submitted and is not to
       be released in whole or part without written approval of SATIMO.


                                                                                                                      Ref: ACR.262.5.14.SATU.A




                             5500               48.6 ±10 %                             5.65 ±10 %
                             5600               48.5 ±10 %                             5.77 ±10 %
                             5800               48.2 ±10 %                             6.00 ±10 %




        7.4    SAR MEASUREMENT RESULT WITH BODY LIQUID
Software                                                                OPENSAR V4
Phantom                                                                 SN 20/09 SAM71
Probe                                                                   SN 18/11 EPG122
Liquid                                                                  Body                54.1 sigma : 0.97
Distance between dipole center and liquid                               15.0 mm
Area scan resolution                                                    dx=8mm/dy=8mm
Zoon Scan Resolution                                                    dx=8mm/dy=8m/dz=5mm
Frequency                                                               835 MHz
Input power                                                             20 dBm
Liquid Temperature                                                      21 °C
Lab Temperature                                                         21 °C
Lab Humidity                                                            45 %



                          Frequency
                                                     1 g SAR (W/kg/W)                      10 g SAR (W/kg/W)
                             MHz
                                                          measured                              measured
                              835                         9.93 (0.99)                           6.35 (0.63)




                                                               Page: 10/11

                 This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                 The information contained herein is to be used only for the purpose for which it is submitted and is not to
                 be released in whole or part without written approval of SATIMO.


                                                                                                                       Ref: ACR.262.5.14.SATU.A




                                                 Equipment Summary Sheet




                                                                                       Validated. No cal               Validated.      No cal
   SAM Phantom                       Satimo                SN-20/09-SAM71
                                                                                       required.                       required.
                                                                                       Validated. No cal               Validated.      No cal
COMOSAR Test Bench                 Version 3                          NA
                                                                                       required.                       required.
                             Rhode & Schwarz
  Network Analyzer                                              SN100132                         02/2013                        02/2016
                                  ZVA

      Calipers                       Carrera                   CALIPER-01                        12/2013                        12/2016

  Reference Probe                    Satimo                EPG122 SN 18/11                       10/2013                        10/2014

     Multimeter                 Keithley 2000                     1188656                        12/2013                        12/2016

  Signal Generator             Agilent E4438C                 MY49070581                         12/2013                        12/2016

                                                                                        Characterized prior to Characterized prior to
      Amplifier                  Aethercomm                        SN 046
                                                                                        test. No cal required. test. No cal required.

    Power Meter                   HP E4418A                    US38261498                        12/2013                        12/2016

   Power Sensor                 HP ECP-E26A                    US37181460                        12/2013                        12/2016

                                                                                        Characterized prior to Characterized prior to
 Directional Coupler           Narda 4216-20                       01386
                                                                                        test. No cal required. test. No cal required.
  Temperature and
                              Control Company                     11-661-9                        8/2012                        8/2015
  Humidity Sensor




                                                                Page: 11/11

                  This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                  The information contained herein is to be used only for the purpose for which it is submitted and is not to
                  be released in whole or part without written approval of SATIMO.


                                   Ref : ACR.262.8.14.SATU.A




Summary:

This document presents the method and results from an accredited SAR reference dipole calibration
performed in SATIMO USA using the COMOSAR test bench. All calibration results are traceable
to national metrology institutions.


                                                                                                                     Ref: ACR.262.8.14.SATU.A




                                  Name                                Function                        Date                    Signature
Prepared by :                Jérôme LUC                         Product Manager                    9/19/2014

Checked by :                 Jérôme LUC                         Product Manager                    9/19/2014

Approved by :          Kim RUTKOWSKI                            Quality Manager                    9/19/2014




                                                                        Customer Name
                                                                      Shenzhen STS Test
                                       Distribution :
                                                                       Services Co., Ltd.




   Issue                   Date                                                           Modifications
     A                  9/19/2014                  Initial release




                                                               Page: 2/11

                This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                The information contained herein is to be used only for the purpose for which it is submitted and is not to
                be released in whole or part without written approval of SATIMO.


                                                                                                                         Ref: ACR.262.8.14.SATU.A




1    Introduction................................................................................................................4
2    Device Under Test .....................................................................................................4
3    Product Description ...................................................................................................4
    3.1      General Information _______________________________________________________4
4    Measurement Method ................................................................................................5
    4.1      Return Loss Requirements __________________________________________________5
    4.2      Mechanical Requirements___________________________________________________5
5    Measurement Uncertainty..........................................................................................5
    5.1      Return Loss ______________________________________________________________5
    5.2      Dimension Measurement ___________________________________________________5
    5.3      Validation Measurement ____________________________________________________5
6    Calibration Measurement Results ..............................................................................6
    6.1      Return Loss and Impedance In Head Liquid ____________________________________6
    6.2      Return Loss and Impedance In Body Liquid ____________________________________6
    6.3      Mechanical Dimensions ____________________________________________________6
7    Validation measurement ............................................................................................7
    7.1      Head Liquid Measurement __________________________________________________7
    7.2      SAR Measurement Result With Head Liquid____________________________________8
    7.3      Body Liquid Measurement __________________________________________________9
    7.4      SAR Measurement Result With Body Liquid __________________________________10
8    List of Equipment ....................................................................................................11




                                                                   Page: 3/11

                    This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                    The information contained herein is to be used only for the purpose for which it is submitted and is not to
                    be released in whole or part without written approval of SATIMO.


                                                                                                                   Ref: ACR.262.8.14.SATU.A




This document contains a summary of the requirements set forth by the IEEE 1528, OET 65 Bulletin
C and CEI/IEC 62209 standards for reference dipoles used for SAR measurement system validations
and the measurements that were performed to verify that the product complies with the fore
mentioned standards.




                                                   Device Under Test
        Device Type                                       COMOSAR 1900 MHz REFERENCE DIPOLE
        Manufacturer                                      Satimo
        Model                                             SID1900
        Serial Number                                     SN 30/14 DIP1G900-333
        Product Condition (new / used)                    New

A yearly calibration interval is recommended.



3.1     GENERAL INFORMATION
                                                                           , OET 65 Bulletin
C and CEI/IEC 62209 standards. The product is designed for use with the COMOSAR test bench
only.




                                                Satimo COMOSAR Validation Dipole



                                                             Page: 4/11

              This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
              The information contained herein is to be used only for the purpose for which it is submitted and is not to
              be released in whole or part without written approval of SATIMO.


                                                                                                                   Ref: ACR.262.8.14.SATU.A




The IEEE 1528, OET 65 Bulletin C and CEI/IEC 62209 standards provide requirements for
reference dipoles used for system validation measurements. The following measurements were
performed to verify that the product complies with the fore mentioned standards.

       4.1   RETURN LOSS REQUIREMENTS
The dipole used for SAR system validation measurements and checks must have a return loss of -20
dB or better. The return loss measurement shall be performed against a liquid filled flat phantom,
with the phantom constucted as outlined in the fore mentioned standards.

       4.2   MECHANICAL REQUIREMENTS
The IEEE Std. 1528 and CEI/IEC 62209 standards specify the mechanical components and
dimensions of the validation dipoles, with the dimensions frequency and phantom shell thickness
dependent. The COMOSAR test bench employs a 2 mm phantom shell thickness therefore the
dipoles sold for use with the COMOSAR test bench comply with the requirements set forth for a 2
mm phantom shell thickness.




All uncertainties listed below represent an expanded uncertainty expressed at approximately the 95%
confidence level using a coverage factor of k=2, traceable to the Internationally Accepted Guides to
Measurement Uncertainty.

       5.1   RETURN LOSS
The following uncertainties apply to the return loss measurement:


                      400-6000MHz                                                                0.1 dB

       5.2   DIMENSION MEASUREMENT
The following uncertainties apply to the dimension measurements:


                            3 - 300                                                             0.05 mm


       5.3   VALIDATION MEASUREMENT
The guidelines outlined in the IEEE 1528, OET 65 Bulletin C, CENELEC EN50361 and CEI/IEC
62209 standards were followed to generate the measurement uncertainty for validation
measurements.

                                1g                                                     20.3 %
                               10 g                                                    20.1 %

                                                             Page: 5/11

              This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
              The information contained herein is to be used only for the purpose for which it is submitted and is not to
              be released in whole or part without written approval of SATIMO.


                                                                                                                      Ref: ACR.262.8.14.SATU.A




      6.1   RETURN LOSS AND IMPEDANCE IN HEAD LIQUID




          1900                                  -23.68                                     -20                          51.2     + 6.4 j

      6.2   RETURN LOSS AND IMPEDANCE IN BODY LIQUID




          1900                                  -20.22                                     -20                          48.8     + 9.6 j

      6.3   MECHANICAL DIMENSIONS
Frequency MHz                           L mm                                         h mm                                      d mm

                           required              measured                required             measured             required           measured

    300                  420.0 ±1 %.                                   250.0 ±1 %.                                6.35 ±1 %.
    450                  290.0 ±1 %.                                   166.7 ±1 %.                                6.35 ±1 %.
    750                  176.0 ±1 %.                                   100.0 ±1 %.                                6.35 ±1 %.
    835                  161.0 ±1 %.                                    89.8 ±1 %.                                 3.6 ±1 %.


                                                                Page: 6/11

                 This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                 The information contained herein is to be used only for the purpose for which it is submitted and is not to
                 be released in whole or part without written approval of SATIMO.


                                                                                                                   Ref: ACR.262.8.14.SATU.A




      900             149.0 ±1 %.                                    83.3 ±1 %.                                  3.6 ±1 %.
     1450              89.1 ±1 %.                                    51.7 ±1 %.                                  3.6 ±1 %.
     1500              80.5 ±1 %.                                    50.0 ±1 %.                                  3.6 ±1 %.
     1640              79.0 ±1 %.                                    45.7 ±1 %.                                  3.6 ±1 %.
     1750              75.2 ±1 %.                                    42.9 ±1 %.                                  3.6 ±1 %.
     1800              72.0 ±1 %.                                    41.7 ±1 %.                                  3.6 ±1 %.
     1900              68.0 ±1 %.                PASS                39.5 ±1 %.               PASS               3.6 ±1 %.         PASS

     1950              66.3 ±1 %.                                    38.5 ±1 %.                                  3.6 ±1 %.
     2000              64.5 ±1 %.                                    37.5 ±1 %.                                  3.6 ±1 %.
     2100              61.0 ±1 %.                                    35.7 ±1 %.                                  3.6 ±1 %.
     2300              55.5 ±1 %.                                    32.6 ±1 %.                                  3.6 ±1 %.
     2450              51.5 ±1 %.                                    30.4 ±1 %.                                  3.6 ±1 %.
     2600              48.5 ±1 %.                                    28.8 ±1 %.                                  3.6 ±1 %.
     3000              41.5 ±1 %.                                    25.0 ±1 %.                                  3.6 ±1 %.
     3500              37.0±1 %.                                     26.4 ±1 %.                                  3.6 ±1 %.
     3700              34.7±1 %.                                     26.4 ±1 %.                                  3.6 ±1 %.




The IEEE Std. 1528, OET 65 Bulletin C and CEI/IEC 62209 standards state that the system
validation measurements must be performed using a reference dipole meeting the fore mentioned
return loss and mechanical dimension requirements. The validation measurement must be performed
against a liquid filled flat phantom, with the phantom constructed as outlined in the fore mentioned
standards. Per the standards, the dipole shall be positioned below the bottom of the phantom, with
the dipole length centered and parallel to the longest dimension of the flat phantom, with the top
surface of the dipole at the described distance from the bottom surface of the phantom.

       7.1   HEAD LIQUID MEASUREMENT
                       Frequency
                                              Relative permittivity (                  Conductivity ( ) S/m
                          MHz                                            r

                                              required           measured            required          measured
                           300               45.3 ±5 %                               0.87 ±5 %
                           450               43.5 ±5 %                               0.87 ±5 %
                           750               41.9 ±5 %                               0.89 ±5 %
                           835               41.5 ±5 %                               0.90 ±5 %
                           900               41.5 ±5 %                               0.97 ±5 %
                          1450               40.5 ±5 %                               1.20 ±5 %
                          1500               40.4 ±5 %                               1.23 ±5 %
                          1640               40.2 ±5 %                               1.31 ±5 %
                          1750               40.1 ±5 %                               1.37 ±5 %
                          1800               40.0 ±5 %                               1.40 ±5 %
                          1900               40.0 ±5 %              PASS             1.40 ±5 %            PASS
                          1950               40.0 ±5 %                               1.40 ±5 %
                          2000               40.0 ±5 %                               1.40 ±5 %


                                                             Page: 7/11

              This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
              The information contained herein is to be used only for the purpose for which it is submitted and is not to
              be released in whole or part without written approval of SATIMO.


                                                                                                                      Ref: ACR.262.8.14.SATU.A




                             2100               39.8 ±5 %                               1.49 ±5 %
                             2300               39.5 ±5 %                               1.67 ±5 %
                             2450               39.2 ±5 %                               1.80 ±5 %
                             2600               39.0 ±5 %                               1.96 ±5 %
                             3000               38.5 ±5 %                               2.40 ±5 %
                             3500               37.9 ±5 %                               2.91 ±5 %


        7.2    SAR MEASUREMENT RESULT WITH HEAD LIQUID
The IEEE Std. 1528 and CEI/IEC 62209 standards state that the system validation measurements
should produce the SAR values shown below (for phantom thickness of 2 mm), within the
uncertainty for the system validation. All SAR values are normalized to 1 W forward power. In
bracket, the measured SAR is given with the used input power.

Software                                                                OPENSAR V4
Phantom                                                                 SN 20/09 SAM71
Probe                                                                   SN 18/11 EPG122
Liquid                                                                                                     41.1 sigma : 1.42
Distance between dipole center and liquid                               10.0 mm
Area scan resolution                                                    dx=8mm/dy=8mm
Zoon Scan Resolution                                                    dx=8mm/dy=8m/dz=5mm
Frequency                                                               1900 MHz
Input power                                                             20 dBm
Liquid Temperature                                                      21 °C
Lab Temperature                                                         21 °C
Lab Humidity                                                            45 %




                          Frequency
                                                     1 g SAR (W/kg/W)                      10 g SAR (W/kg/W)
                             MHz
                                                 required           measured            required          measured
                              300                   2.85                                   1.94
                              450                  4.58                                    3.06
                              750                   8.49                                   5.55
                              835                  9.56                                    6.22
                              900                  10.9                                    6.99
                             1450                    29                                     16
                             1500                   30.5                                   16.8
                             1640                   34.2                                   18.4
                             1750                   36.4                                   19.3
                             1800                   38.4                                   20.1
                             1900                   39.7           39.84 (3.98)            20.5          20.20 (2.02)
                             1950                   40.5                                   20.9
                             2000                   41.1                                   21.1
                             2100                   43.6                                   21.9
                             2300                   48.7                                   23.3


                                                                Page: 8/11

                 This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                 The information contained herein is to be used only for the purpose for which it is submitted and is not to
                 be released in whole or part without written approval of SATIMO.


                                                                                                            Ref: ACR.262.8.14.SATU.A




                   2450                   52.4                                    24
                   2600                   55.3                                   24.6
                   3000                   63.8                                   25.7
                   3500                   67.1                                    25




7.3   BODY LIQUID MEASUREMENT
                Frequency
                                       Relative permittivity (                  Conductivity ( ) S/m
                   MHz                                            r

                                       required           measured            required          measured
                   150                61.9 ±5 %                               0.80 ±5 %
                   300                58.2 ±5 %                               0.92 ±5 %
                   450                56.7 ±5 %                               0.94 ±5 %
                   750                55.5 ±5 %                               0.96 ±5 %
                   835                55.2 ±5 %                               0.97 ±5 %
                   900                55.0 ±5 %                               1.05 ±5 %
                   915                55.0 ±5 %                               1.06 ±5 %
                   1450               54.0 ±5 %                               1.30 ±5 %
                   1610               53.8 ±5 %                               1.40 ±5 %
                   1800               53.3 ±5 %                               1.52 ±5 %
                   1900               53.3 ±5 %              PASS             1.52 ±5 %            PASS
                   2000               53.3 ±5 %                               1.52 ±5 %
                   2100               53.2 ±5 %                               1.62 ±5 %
                   2450               52.7 ±5 %                               1.95 ±5 %
                   2600               52.5 ±5 %                               2.16 ±5 %
                   3000               52.0 ±5 %                               2.73 ±5 %
                   3500               51.3 ±5 %                               3.31 ±5 %
                   5200               49.0 ±10 %                             5.30 ±10 %
                   5300               48.9 ±10 %                             5.42 ±10 %
                   5400               48.7 ±10 %                             5.53 ±10 %


                                                      Page: 9/11

       This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
       The information contained herein is to be used only for the purpose for which it is submitted and is not to
       be released in whole or part without written approval of SATIMO.


                                                                                                                      Ref: ACR.262.8.14.SATU.A




                             5500               48.6 ±10 %                             5.65 ±10 %
                             5600               48.5 ±10 %                             5.77 ±10 %
                             5800               48.2 ±10 %                             6.00 ±10 %




        7.4    SAR MEASUREMENT RESULT WITH BODY LIQUID
Software                                                                OPENSAR V4
Phantom                                                                 SN 20/09 SAM71
Probe                                                                   SN 18/11 EPG122
Liquid                                                                  Body                54.2 sigma : 1.54
Distance between dipole center and liquid                               10.0 mm
Area scan resolution                                                    dx=8mm/dy=8mm
Zoon Scan Resolution                                                    dx=8mm/dy=8m/dz=5mm
Frequency                                                               1900 MHz
Input power                                                             20 dBm
Liquid Temperature                                                      21 °C
Lab Temperature                                                         21 °C
Lab Humidity                                                            45 %



                          Frequency
                                                     1 g SAR (W/kg/W)                      10 g SAR (W/kg/W)
                             MHz
                                                          measured                              measured
                             1900                        43.33 (4.33)                          21.59 (2.16)




                                                               Page: 10/11

                 This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                 The information contained herein is to be used only for the purpose for which it is submitted and is not to
                 be released in whole or part without written approval of SATIMO.


                                                                                                                       Ref: ACR.262.8.14.SATU.A




                                                 Equipment Summary Sheet




                                                                                       Validated. No cal               Validated.      No cal
   SAM Phantom                       Satimo                SN-20/09-SAM71
                                                                                       required.                       required.
                                                                                       Validated. No cal               Validated.      No cal
COMOSAR Test Bench                 Version 3                          NA
                                                                                       required.                       required.
                             Rhode & Schwarz
  Network Analyzer                                              SN100132                         02/2013                        02/2016
                                  ZVA

      Calipers                       Carrera                   CALIPER-01                        12/2013                        12/2016

  Reference Probe                    Satimo                EPG122 SN 18/11                       10/2013                        10/2014

     Multimeter                 Keithley 2000                     1188656                        12/2013                        12/2016

  Signal Generator             Agilent E4438C                 MY49070581                         12/2013                        12/2016

                                                                                        Characterized prior to Characterized prior to
      Amplifier                  Aethercomm                        SN 046
                                                                                        test. No cal required. test. No cal required.

    Power Meter                   HP E4418A                    US38261498                        12/2013                        12/2016

   Power Sensor                 HP ECP-E26A                    US37181460                        12/2013                        12/2016

                                                                                        Characterized prior to Characterized prior to
 Directional Coupler           Narda 4216-20                       01386
                                                                                        test. No cal required. test. No cal required.
  Temperature and
                              Control Company                     11-661-9                        8/2012                        8/2015
  Humidity Sensor




                                                                Page: 11/11

                  This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                  The information contained herein is to be used only for the purpose for which it is submitted and is not to
                  be released in whole or part without written approval of SATIMO.



Document Created: 2016-04-20 15:41:30
Document Modified: 2016-04-20 15:41:30

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC