SAR report part 3

FCC ID: 2ABOSSKYELITE4

RF Exposure Info

Download: PDF
FCCID_4031465

                                                                              Report No: CCISE180814401

                Dipole Impedance and Return Loss calibration Report
 Object:                           D835V2 - SN: 4d154

 Calibration Date:                 June 10, 2018

 Calibration reference:            IEEE Std 1528:2013, IEC 62209-1:2016, FCC KDB 865664 D01

 Calibrated By:
                                                              (Janet Wei, SAR project engineer)

 Reviewed By:
                                                        (Bruce Zhang, Technical manager)


Environment of Test Site
 Temperature:                      21 ~ 23C
 Humidity:                         50~60% RH
 Atmospheric Pressure:             1011 mbar

Test Data
     Measurement Plot for Head TSL In 2017                          Measurement Plot for Body TSL In 2017




Shenzhen Zhongjian Nanfang Testing Co., Ltd.                                           Project No.: CCISE1808144
No. B-C, 1/F., Building 2, Laodong No.2 Industrial Park, Xixiang Road,
Bao’an District, Shenzhen, Guangdong, China
Telephone: +86 (0) 755 23118282 Fax: +86 (0) 755 23116366, E-mail: info@ccis-cb.com               Page 121 of 156


                                                                                Report No: CCISE180814401

       Measurement Plot for Head TSL In 2018                          Measurement Plot for Body TSL In 2018




  Comparison with Original report
                               Calibrated By       Calibrated By CCIS
         Items                                                                Deviation                 Limit
                               CCIS In 2017              In 2018
Impendence for Head TSL       51.46Ω –4.26jΩ         51.57Ω –3.48jΩ        0.11Ω+0.78jΩ                  ±5Ω
Return Loss for Head TSL         -27.05dB                -28.5dB               -9.2%          ±20%(No less than 20 dB)
Impendence for Body TSL       46.24Ω-4.46 jΩ         45.62Ω-5.28 jΩ        -0.62Ω-0.82 jΩ                ±5Ω
Return Loss for Body TSL          -26.8dB                -28.5dB                6.3%          ±20%(No less than 20 dB)
  Result
  Compliance




  Shenzhen Zhongjian Nanfang Testing Co., Ltd.                                              Project No.: CCISE1808144
  No. B-C, 1/F., Building 2, Laodong No.2 Industrial Park, Xixiang Road,
  Bao’an District, Shenzhen, Guangdong, China
  Telephone: +86 (0) 755 23118282 Fax: +86 (0) 755 23116366, E-mail: info@ccis-cb.com                 Page 122 of 156


CCIS


            SAR Reference Dipole Calibration Report



                                                Ref : ACR.93.5.17.SATU.A




       WALTEK SERVICES(SHENZHEN) CO.,LTD
        1/F., FUKANGTAI BUILDING,WEST BAIMA ROAD,
                   SONGGANG STREET
          BAOAN DISTRICT,SHENZHEN GUANGDONG
                     518105,CHINA
           MVG COMOSAR REFERENCE DIPOLE
                            FREQUENCY: 1800 MHZ
                            SERIAL NO.: SN 09/15 DIP 1G800—360
                                   Calibrated at MVG US
                        2105 Barrett Park Dr. — Kennesaw, GA 30144
                                   \\\\“\HIH,"/’,




                                 *z(accrepitep)
                                   s
                                     i us             Caribrationcext snzeeq2

                               Calibration Date: 02/28/2018


          Summary:

          This document presents the method and results from an accredited SAR reference dipole calibration
          performed in MVG USA using the COMOSAR test bench. All calibration results are traceable to
          national metrology institutions.




       oo e                                                                               hn egeses n nwe e n ne n e en n a s


CCIS                                                                                                             |
  enuc                      SAR REFERENCE DIPOLE CALIBRATION REPORT                                                     Ref: ACR.93.5.17.SATU.A




                                    Name                                Function                          Date                  Signature
  Prepared by :                Jérome LUC                         Product Manager                    3/13/2018                    )/5

  Checked by :                 Jérome LUC                         Product Manager                    3/13/2018                    fi’

  Approved by :           Kim RUTKOWSKTI                          Quality Manager                    3/13/2018                 hum Patthowth!




                                                                           Customer Name

                                                                          Waltek Services
                                         Distribution :                    (Shenzhen)Co.,
                                                                                Ltd




     Issue                   Date                                                          Mc         cations
       A                   3/13/2018                 Initial release




                                                                Page: 2/11

                  This document shall not be reproduced, except in full or in part, without the written approval ofMVG.
                  The information contained herein is to be used onlyfor the purpose for which it is submitted and is not to
                  be released in whole orpart without written approval ofMVG.


CCIS                                                                                                                |
  mvG                          SAR REFERENCE DIPOLE CALIBRATION REPORT                                                     Ref: ACR.93.5.17.SATU.A




                                                      TABLE OF CONTENTS



  1     IMEROGUCHOM..... .. .e mm renereneee en erne en reee en en en en reerne en en eree en en en en ennn en en 4
  2     D@eViG@ UNG@P TS .......... 0e renerneereeee en en en neee en en en eeereee en en nenn en en 4

  3     PFOGUCt DSCTIDHON ........ ce e e mm erneene en ernnernneereeerenern en en en en en rnenrnv en nenenen en 4
      34       General Information

  4     Megsurement MethOd ———mmmmommummmommummmomummmomumummenmmmemmame 5
      4.1      Return Loss Requirements

      4.2      Mechanical Requirements
  5     MEASUP@MENE UNC@THMNEY ........... en en ereneneeeereen en e reer en en neneneerev en ene en 5

      §.1      Return Loss
      §2       Dimension Measurement

      5.3      Validation Measurement
  6     CalibFatiON MEASUP@M@Nt ReSUIS ..............00eeee e mm en erreerneeereen en en en en es 6

      6.1      Return Loss and Impedance In Head Liquid
      6.2      Return Loss and Impedance In Body Liquid
      6.3      Mechanical Dimensions

  7     VAIiQAHON M@ASUT@M@NE........... .cesn en en en en ree en en en reen en en nenenen en 7
      7.1      Head Liquid Measurement

      72       SAR Measurement Result With Head Liquid
      7.3      Body Liquid Measurement

      7.4      SAR Measurement Result With Body Liquid                                                                                         10
  $     Listrdl® HqUIPMENE m=mmmmommmmummomommmmmmeommmmmnmmomenn 11




                                                                   Page: 3/11

                     This document shall not be reproduced, except in full or in part, without the written approval ofMVG.
                     The information contained herein is to be used onlyfor the purpose for which it is submitted and is not to
                     be released in whole orpart without written approval ofMVG.




                                                                                                                                  regscs ce


CCIS                                                                                                                |
   mvG                          SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.93.5.17.SATU.A




   1              INTRODUCTION


   This document contains a summary of the requirements set forth by the IEEE 1528, FCC KDBs and
   CEIIEC 62209 standards for reference dipoles used for SAR measurement system validations and

  the measurements that were performed to verify that the product complies with the fore mentioned
   standards.




  2               DEVICE UNDER TEST




                                                           Device Under Test

              Device Type                                         COMOSAR 1800 MHz REFERENCE DIPOLE
              Manufacturer                                        MVG
              Model                                               $ID1800
              Serial Number                                       SN 09/15 DIP 1G800—360
              Product Condition (new / used)                      New

   A yearly calibration interval is recommended.

  3               PRODUCT DESCRIPTION


   3.1         GENERAL INFORMATION

   MVG‘s COMOSAR Validation Dipoles are built in accordance to the IEEE 1528, FCC KDBs and
   CEIIEC 62209 standards. The product is designed for use with the COMOSAR test bench only.




                                       Figure 1 —MVG COMOSAR Validation Dipole




                                                                    Page: 4/11

                      This document shall not be reproduced, except in full or in part, without the written approval ofMVG.
                      The information contained herein is to be used onlyfor the purposefor which it is submitted and is not to
                      be released in whole orpart without written approval ofMVG.




  «on n n maanwnn n n nwnn n wen ty n Nee nnnnag n eng ccv es                                                                     hn egeses n nwe e n ne n e en n a s


CCIS
   enuc                         SAR REFERENCE DIPOLE CALIBRATION REPORT                                                     Ref: ACR.93.5.17.SATU.A



            4      MEASUREMENT METHOD
   The IEEE 1528, FCC KDBs and CEIIEC 62209 standards provide requirements for reference
   dipoles used for system validation measurements. The following measurements were performed to
   verify that the product complies with the fore mentioned standards.

            4.1       RETURN LOSS REQUIREMENTS
   The dipole used for SAR system validation measurements and checks must have a return loss of —20
   dB or better. The return loss measurement shall be performed against a liquid filled flat phantom,
   with the phantom constucted as outlined in the fore mentioned standards.

            4.2       MECHANICAL REQUIREMENTS
  The IEEE Std. 1528 and CEIIEC 62209 standards specify the mechanical components and
  dimensions of the validation dipoles, with the dimensions frequency and phantom shell thickness
  dependent. The COMOSAR test bench employs a 2 mm phantom shell thickness therefore the
  dipoles sold for use with the COMOSAR test bench comply with the requirements set forth for a 2
  mm phantom shell thickness.


   5               MEASUREMENT UNCERTAINTY

   All uncertainties listed below represent an expanded uncertainty expressed at approximately the 95%
   confidence level using a coverage factor of k=2, traceable to the Internationally Accepted Guides to
   Measurement Uncertainty.

             5.1    RETURN LOSS
   The following uncertainties apply to the return loss measurement:

                            Frequency band                                     Expanded Uncertainty on Return Loss

                              400—6000MHz                                                                0.1 dB

             5.2    DIMENSION MEASUREMENT
   The following uncertainties apply to the dimension measurements:
                               Length (mm)                                         Expanded Uncertainty on Length

                                    3 — 300                                                             0.05 mm



             5.3    VALIDATION MEASUREMENT
   The guidelines outlined in the IEEE 1528, FCC KDBs, CENELEC EN50361 and CEVIEC 62209
   standards were followed to generate the measurement uncertainty for validation measurements.
                               Scan Volume                                        Expanded Uncertainty
                                        1g                                                     20.3 %

                                                                    Page: 5/11

                      This document shall not be reproduced, except in full or in part, without the written approval ofMVG.
                      The information contained herein is to be used onlyfor the purpose for which it is submitted and is not to
                      be released in whole orpart without written approval ofMVG.




  «on n n maanwnn n n nwnn n wen ty n Nee nnnnag n eng ccv es                                                                      hn egeses n nwe e n ne n e en n a s


CCIS
                             SAR REFERENCE DIPOLE CALIBRATION REPORT                                                     Ref: ACR.93.5.17.SATU.A




                                     10 g                                                   20.1%


         6     CALIBRATION MEASUREMENT RESULTS

         6.1       RETURN LOSS AND IMPEDANCE IN HEAD LIQUID
                                                                    Frequency, MHz
                              1700     1720    1740     1760    1780      800     1820     1840      i860    1880    1900




                            50—
    Frequency (MHz)                      Return Loss (dB)                        Requirement (dB)                             Impedance
          1800                                —25.49                                   —20                                  45.4 02 + 2.6 jC2

         6.2       RETURN LOSS AND IMPEDANCE IN BODY LIQUID
                                                                    Frequency, MHz
                              1700     1720    1740     1760    1780      800     1820     1840     i1860    1880    1900
                             o—




                            40—
    Frequency (MHz)                      Return Loss (dB)                        Requirement (dB)                             Impedance
          1800                                —26.35                                   —20                                  45.4 0 — 1.5 j02

         6.3       MECHANICAL DIMENSIONS

   Frequency MHz                            L mm                                         h mm                                    d mm

                             required              measured                required               measured           required           measured

       300                 420.0 +1 %.                                   250.0 £1 %.                                6.35 +1 %.

                                                                 Page: 6/11

                   This document shall not be reproduced, except in full or in part, without the written approval ofMVG.
                   The information contained herein is to be used onlyfor the purpose for which it is submitted and is not to
                   be released in whole orpart without written approval ofMVG.




                                                                                                                                  regscs ce


CCIS                                                                                                           |
  mvG                     SAR REFERENCE DIPOLE CALIBRATION REPORT                                                      Ref: ACR.93.5.17.SATU.A




       450              290.0 £1 %.                                   166.7 +1 %.                                  6.35 +1 %.
       750              176.0 +1 %.                                   100.0 +1 %.                                  6.35 +1 %.
       835              161.0 +1 %.                                   89.8 +1 %.                                    3.6 +1 %.
       900              149.0 +1 %.                                   83.3 +1 %.                                    3.6 +1 %.
       1450              89.1 +1 %.                                   51.7 +1 %.                                    3.6 41 %.
       1500              80.5 £1 %.                                   50.0 £1 %.                                    3.6 41 %.
       1640              79.0 +1 %.                                   45.7 £1 %.                                    3.6 +1 %.
       1750              75.2 £1%.                                    42.9 £1 %.                                    3.6 +1 %.
       1800              72.0 +1 %.                PASS                41.7 +1 %.               PASS               3.6 +1 %.           PASS
       1900              68.0 +1 %.                                    39.5 £1 %.                                  3.6 +1 %.
       1950              66.3 41 %.                                    38.5 £1 %.                                  3.6 +1 %.
       2000              64.5 +1 %.                                    37.5 +1 %.                                  3.6 +1 %.
       2100              61.0£1%.                                      35.7 £1 %.                                  3.6 +1 %.
       2300              55.5 +1 %.                                    32.6 +1 %.                                  3.6 +1 %.
       2450              51.5 +1 %.                                    30.4 £1 %.                                  3.6 +1 %.
       2600              48.5 £1 %.                                    28.8 +1 %.                                  3.6 +1 %.
       3000              41.5 £1%.                                     25.0 +1 %.                                  3.6 +1 %.
       3500              37.0£1 %.                                     26.4 +1 %.                                  3.6 +1 %.
       3700              34.741 %.                                     26.4 +1 %.                                  3.6 +1 %.

         7     VALIDATION MEASUREMENT
  The IEEE Std. 1528, FCC KDBs and CEVIEC 62209 standards state that the system validation
  measurements must be performed using a reference dipole meeting the fore mentioned return loss
  and mechanical dimension requirements. The validation measurement must be performed against a
  liquid filled flat phantom, with the phantom constructed as outlined in the fore mentioned standards.
  Per the standards, the dipole shall be positioned below the bottom of the phantom, with the dipole
  length centered and parallel to the longest dimension of the flat phantom, with the top surface of the
  dipole at the described distance from the bottom surface of the phantom.

         7.1   HEAD LIQUID MEASUREMENT

                         Fre'?,lliiezncy        Relative permittivity (,)                Conductivity {0) S/m

                                                required           measured            required          measured

                             300               45.3 45 %                               0.87 +5 %

                             450               43.5 45 %                               0.87 25 %

                             750               41.9 £5 %                               0.89 45 %

                             835               41.5 £5 %                               0.90 +5 %

                             900               41.5 45 %                               0.97 25 %

                             1450              40.5 45 %                               1,20 45 %
                             1500              40.4 £5 %                               1.3545%
                             1640              40.2 £5 %                               1.31 +5 %

                             1750              40.1 £5 %                               1.3745%

                                                              Page: 7/11

                This document shall not be reproduced, except in full or in part, without the written approval ofMVG.
                The information contained herein is to be used onlyfor the purpose for which it is submitted and is not to
                be released in whole orpart without written approval ofMVG.




                                                                                                                                regeses n nwe e n nen e en n a s


CCIS
                                                                                                                          Ref: ACR.93.5.17.SATU.A
  mvG                        SAR REFERENCE DIPOLE CALIBRATION REPORT




                                1800              40.0 £5 %              PASS             1.40 £5 %            PASS

                                1900              40.0 £5 %                               1.40 £5 %

                                1950              40.0 £5 %                               1.40 £5 %

                                2000              40.0 £5 %                               1.40 45 %

                                2100              39.8 +5 %                               1.49 £5 %

                                2300              39.5 45 %                               1.67 45 %

                                2450              39.245%                                 1.80 45 %

                                2600              39.0 £5 %                               1.96 45 %

                                3000              38.5 45 %                               2.40 £5 %

                                3500              37.9 £5 %                               2.91 25 %


             7.2   SAR MEASUREMENT RESULT WITH HEAD LIQUID
  The IEEE Std. 1528 and CELIIEC 62209 standards state that the system validation measurements
  should produce the SAR values shown below (for phantom thickness of 2 mm), within the
  uncertainty for the system validation. All SAR values are normalized to 1 W forward power. In
  bracket, the measured SAR is given with the used input power.

  Software                                                                OPENSAR V4
  Phantom                                                                 SN 20/09 SAM71
  Probe                                                                   SN18/11 EPG122
  L    d                                                                  Head   L   Values:              ‘:41.7s         a : 1.46
  Distance                                                                 10.0 mm
  Area scan resolution                                                    dx=8mm        mm
  Zoon Scan Reso       n                                                       8      8
  Fi                                                                       1800 MHz
                                                                          20 dBm
  T
      id T                                                                2 "C
  Lab T.                                                                  21 °C
  Lab Humidi                                                              45 %




                            F'eme:‘"                   1 g SAR (W/ke/w)                      10 g SAR (W/ke/w)
                                                   required           measured            required          measured

                                300                  2.85                                    1.94

                                450                  4.58                                    3.06

                                750                  8.49                                    5.55

                                835                  9.56                                    6.22

                                900                   10.9                                   6.99

                                1450                   29                                     16

                                1500                 30.5                                    16.8
                                1640                 34.2                                    184
                                1750                 36.4                                    19.3

                                1800                 38.4            38.76 (3.88)            20.1          20.29 (2.03)

                                                                 Page: 8/11

                   This document shall not be reproduced, except in full or in part, without the written approval ofMVG.
                   The information contained herein is to be used onlyfor the purpose for which it is submitted and is not to
                   be released in whole orpart without written approval ofMVG.




                                                                                                                                 regsse e >


CCIS
  nuC                  SAR REFERENCE DIPOLE CALIBRATION REPORT                                                         Ref: ACR.93.5.17.SATU.A




                          1900

                          1950

                          2000

                          2100

                          2300

                          2450

                          2600

                          3000

                          3500

                          3700


               n rrarmtarm
                                             Zoevieta



       z4
    sheo
    Shie
    se
    ns
                                                                                           ao
                                                                                           o
                                                                                      SaR (Wikgl
                                                                                           s
                                                                                           w




                                                                                                   2

                                                                                                   1
                                                                                                   048—
                                                                                                       0       10 12 14 1 19 20 22 2b 2s ce t
              TT KbT          Ybed                                                                                    2 (nn}




       7.3   BODY LIQUID MEASUREMENT

                      Fre'?/lliiezncy        Relative permittivity (€;)                Conductivity (0) S/m

                                             required           measured            required               measured

                           150              61.9 +5 %                               0.80 +5 %

                          300               58.2 45 %                               0.92 +5 %

                          450               56.7 £5 %                               0.94 +5 %

                          750               55.5 45 %                               0.96 +5 %

                          835               55.2 #5 %                               0.97 25 %

                          900               55.0 £5 %                               1.05 5 %

                          915               55.0 £5 %                               1.06 5 %

                          1450              54.0 £5 %                               1.30 5 %

                          1610              53.8 45 %                               1.40 £5%

                          1800               53.3 45 %             PASS             1.52 45 %                PASS
                          1900               53.3 45 %                              1.52 45 %
                          2000               53.3 15 %                              1.52%5%

                          2100              53.2 45 %                               1.62 45 %

                                                           Page: 9/11

             This document shall not be reproduced, except in full or in part, without the written approval ofMVG.
             The information contained herein is to be used onlyfor the purpose for which it is submitted and is not to
             be released in whole orpart without written approval ofMVG.


CCIS
                                                                                                                               Ref: ACR.93.5.17.SATU.A
  mvG                         SAR REFERENCE DIPOLE CALIBRATION REPORT




                                     2300         52.9 £5 %                              1.81 +5 %

                                     2450         52.7.45%                               1.95 45 %

                                     2600         52.5 45 %                              2.16 £5 %

                                     3000         52.0 45 %                              2.73 25 %

                                     3500         51.3 45 %                              3.31 45 %

                                     3700         51.0 +5 %                              3.55 45 %

                                     5200         49.0 £10 %                            5.30 £10 %

                                     5300         48.9 £10 %                            5.42 10 %

                                     5400         48.7 £10 %                            5.53 £10 %

                                     5500         48.6 £10 %                            5.65 £10 %

                                     5600         48.5 £10 %                            5.77 £10 %

                                     5800         48.2 £10 %                            6.00 £10 %


          7.4   SAR MEASUREMENT RESULT WITH BODY LIQUID
  Software                                                                OPENSAR V4
  Phantom                                                                 SN 20/09 SAMT71
  Probe                                                                   SN18/11 EPG122
  Liguid                                                                        Li    Values:
  Distance between           center and li   id                           10.0 mm
  Area scan resolution                                                    dx=8mm      $mm
  Zoon Scan Resolution                                                    dx=8m       $mm/dz=Smm
  F                                                                       1800 MHz
                                                                          20    dBm
  L                                                                       2t °C
                                                                          21 °C
                                                                          45 %




                              F'emez““’               1 g SAR (W/ke/w)                           10 g SAR (W/ke/w)
                                                           measured                                       measured

                                     1800                  38.90 (3.89)                                  20.84 (2.08)

                     S reabninicerce m                      Z                             £

                  Sutace Rulad mss                 Fows:
                                                                                               on
                                                                                           SeR (Wigl
                                                                                               s
                                                                                               w
                                                                                               n




                                                                                                       ost—
                                                                                                          o             to 12 14 1e 1e 20 2e 2 2s 20 so
                                                                                                                             Z (na}



                                                                Page: 10/11

                  This document shall not be reproduced, except in full or in part, without the written approval ofMVG.
                  The information contained herein is to be used onlyfor the purpose for which it is submitted and is not to
                  be released in whole orpart without written approval ofMVG.




                                                                                                                                        Ataae


CCIS
                                                                                                                         Ref: ACR.93.5.17.SATU.A
   mvG                        SAR REFERENCE DIPOLE CALIBRATION REPORT




         8    LIST OF EQUIPMENT



                                                  Equipment Summary Sheet

      Equipment                Manufacturer /              Identification No                    Current                   Next Calibration
      Description                  Model                                              ‘|    Calibration Date                    Date

     SAM Phantom                        mvoe                 SNGntggayy,; |flCaisd
                                                                          required.
                                                                                    Nowal                               NVelidated.
                                                                                                                        required.
                                                                                                                                        No wal
  COMOSAR Test Bench|                Version 3                         NA                  feidsisd.
                                                                                           required.
                                                                                                     Nosal              NWalldeied
                                                                                                                        required.
                                                                                                                                        ho cal
    Network Analyzer           RhOde8Schwarz                      sN100132                        o2p016                        0222019
        Calipers                       Carrera                  CALIPER—01                        01/2017                       01/2020

    Reference Probe                     MVG                 EPG122 SN 1811                        10/2017                       10/2018

       Multimeter                 Keithley 2000                     1188656                       01/2017                       01/2020

    Signal Generator             Agilent E4438C                 MY49070581                        01/2017                       01/2020

             e                                                                             Characterized prior to |\Characterized prior to
        Amplifier                  Asthersomn                       SN 948                 test. No cal required. |test. No cal required.

      Power Meter                   HP E4418A                   US38261498                        01/2017                       01/2020

     Power Sensor                 HP ECP—E26A                    US37181460                       01/2017                       01/2020

   Directional Coupler           Narda 4216—20                       01386                 Characterized pripr to |Characterized pri_or to
                                                                                           test. No cal required. |test. No cal required.
    Temperature and
    Humidiy Sensor              Control Company                   150798832                       10/2017                       10/2019




                                                                 Page: 11/11

                    This document shall not be reproduced, except in full or in part, without the written approval ofMVG.
                    The information contained herein is to be used onlyfor the purposefor which it is submitted and is not to
                    be released in whole orpart without written approval ofMVG.




                                                                                                                                regscs ce


CCGIS                                                                                             |
                   im5
                   4#                      in Collsboration with                   t /y"/‘s
                                                                                  SJm                 A +EA
                   a=7T"]]J,
                   v
                             e 3lgc=MR/
                             CALIBRATION LABORATORY
                                                    . CNAs AREW
                                                                                                                     Bot
              Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China   'f,,/,\\\?‘         v CALIBRATION
              Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504               "Ardili(sS®         CNAS LO570
              E—mail: cttl@chinattl.com        Hrtp://wwrw.chinattl.on
        Client             CCIs                                               Certificate No:         Z16—97090

  CALIBRATION CERTIFICATE

  Object                                    D1900V2 — SN: 5d175

  Calibration Procedure(s)                  FD—211—2—003—01

                                            Calibration Procedures for dipole validation kits

  Calibration date:                         Jun 15, 2016

  This calibration Certificate documents the traceability to national standards, which realize the physical units of
  measurements(S1). The measurements and the uncertainties with confidence probability are given on the following
  pages and are part of the certificate.

  All calibrations have been conducted in the closed laboratory facility: environment temperature(22+3)‘c and
  humidity<70%.

  Calibration Equipment used (M&TE critical for calibration)


  Primary Standards                 ID #                Cal Date(Calibrated by, Certificate No.)         Scheduled Calibration
   Power Meter   NRP2               101919             01—Jul—15 (CTTL, No.J15X04256)                           Jun—16
   Power sensor NRP—Z91             101547             01—Jul—15 (CTTL, No.J15X04256)                           Jun—16
   Reference Probe EX3DV4           SN 7307            19—Feb—16(SPEAG,No.EX3—7307_Feb16)                        Feb—17
   DAE4                             SN 771             02—Feb—16(CTTL—SPEAG,No.216—97011)                        Feb—17

   Secondary Standards              ID #               Cal Date(Calibrated by, Certificate No.)          Scheduled Calibration
   Signal Generator E4438C          MY49071430          O1—Feb—16 (CTTL, No.J16X00893)                         Jan—17
   Network Analyzer E5071C          MY46110673          26—Jan—16 (CTTL, No.J16X00894)                         Jan—17



                                    Name                           Function                                   Signature
  Calibrated by:                   Zhao Jing                  SAR Test Engineer                               %

  Reviewed by:                     Qi Dianyuan                 SAR Project Leader                     M;/

  Approved by:                     Lu Bingsong                 Deputy Director of the laboratory         [‘ %?Z

                                                                                  Issued: Jun 17, 2016
  This calibration certificate shall not be reproduced exceptin full without written approval of the laboratory.



      Certificate No: Z16—97090                               Page 1 of 8


CCIS                                                                          |
            .t\"                      in Collaboration with


                                      CALIBRATION LABORATORY
            y
          Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
          Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
          E—mail: cttl@chinattl.com       Hitp://www.chinattl.en


  Glossary:
  TSL                       tissue simulating liquid
  ConvF                      sensitivity in TSL / NORMx,y,z
  N/A                       not applicable or not measured


  Calibration is Performed According to the Following Standards:
  a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak
     Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
     Communications Devices: Measurement Techniques", June 2013
  b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) For hand—held
     devices used in close proximity to the ear (frequency range of 300MHz to 3GHz)", February
     2005
  c) IEC 62209—2, "Procedure to measure the Specific Absorption Rate (SAR) For wireless
     communication devices used in close proximity to the human body (frequency range of
     30MHz to 6GHz)", March 2010
  d) KDB865664, SAR Measurement Requirements for 100 MHz to 6 GHz


  Additional Documentation:
  e) DASY4/5 System Handbook


  Methods Applied and Interpretation of Parameters:
  * Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
  * Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     paralle! to the body axis.
  * Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
  * Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
  * SAR measured: SAR measured at the stated antenna input power.
     SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
       connector.
  e    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
       nominal SAR result.


      The reported uncertainty of measurement is stated as the standard uncertainty of
      Measurement multiplied by the coverage factor k=2, which for a normal distribution
      Corresponds to a coverage probability of approximately 95%.



  Certificate No: Z16—97090                             Page 2 of 8


CCGIS
|
                   m ® in Collaboration with
                   TTL y.B__e_a
                   V                        CALIBRATION LABORATORY
                Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
                Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
                E—mail: cttl@chinattl.com       Hitp://www.chinattl.cn


         Measurement Conditions
            DASY system configuration, as far as not given on page 1.
              DASY Version                                       DASY52                                      52.8.8.1258

              Extrapolation                             Advanced Extrapolation

              Phantom                                    Triple Flat Phantom 5.1C

              Distance Dipole Center— TSL                        10 mm                                     with Spacer

              Zoom Scan Resolution                          dx, dy, dz = 5 mm

              Frequency                                     1900 MHz + 1 MHz


         Head TSL parameters
            The following parameters and calculations were applied.
                                                                  Temperature               Permittivity           Conductivity
              Nominal Head TSL parameters                            22.0 °C                   40.0                1.40 mho/m

              Measured Head TSL parameters                       (22.0 £ 0.2) °C             40.3 + 6 %          1.38 mho/m + 6 %

              Head TSL temperature change during test                <1.0 °C                    ——                       ———
         SAR result with Head TSL
              SAR averaged over 1_cm" _(1 g) of Head TSL                        Condition
              SAR measured                                                250 mW input power                   9.99 mW / g

              SAR for nominal Head TSL parameters                            normalized to 1W          40.4 mW /g + 20.8 % (k=2)

‘l            SAR averaged over 10 cm° (10 g) of Head TSL                       Condition
              SAR measured                                                250 mW input power                   5.28 mW / g

              SAR for nominal Head TSL parameters                            normalized to 1W          21.3 mW ig + 20.4 % (k=2)


         Body TSL parameters
            The following parameters and calculations were applied.
                                                                  Temperature               Permittivity           Conductivity

 |            Nominal Body TSL parameters                            22.0 °C                   53.3                1.52 mho/m

 ;            Measured Body TSL parameters                       (22.0 £ 0.2) °C             53.3 + 6 %          1.54 mho/m + 6 %

 1            Body TSL temperature change during test                <1.0 °C                    ————                     See
         SAR result with Body TSL
 \            SAR averaged over 1 em‘ (1 g) of Body TSL                         Condition
 !            SAR measured                                                250 mW input power                   10.1 mW /g

 v‘           SAR for nominal Body TSL parameters                            normalized to 1W          40.1 mW /g £ 20.8 % (k=2)

     ?        SAR averaged over 10 cm° (10 g) of Body TSL                       Condition
              SAR measured                                                250 mW input power                   5.39 mW /g

              SAR for nominal Body TSL parameters                            normalized to 1W          21.5 mW /g + 20.4 % (k=2)

    |    Certificate No: Z16—97090                             Page 3 of 8


       CCIS
                    zi Ees
                       liRmamyye~             CALIBRATION LABORATORY
                  Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
                  Tel: +86—10—62304633—2079        Fax: +86—10—62304633—2504
                  E—mail: cttl@chinattl.com        Hitp://www.chinattl.en


            Appendix

           Antenna Parameters with Head TSL

                Impedance, transformed to feed point                                  53.20+ 5.44j0

                Return Loss                                                             — 24.30B


            Antenna Parameters with Body TSL

                Impedance, transformed to feed point                                  48.90+ 5.75j0

                Return Loss                                                             —24.60B


            General Antenna Parameters and Design

                Electrical Delay (one direction)                                        1.304 ns




            After long term use with 100W radiated power, only a slight warming of the dipole nearthe feedpoint can
            be measured.

            The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly
            connected to the second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some
            of the dipoles, small end caps are added to the dipole arms in order to improve matching when loaded
            according to the position as explained in the "Measurement Conditions" paragraph. The SAR data are not
            affected by this change. The overall dipole length is still according to the Standard.
            No excessive force must be applied to the dipole arms, because they might bend or the soldered
            connections near the feedpoint may be damaged.



            Additional EUT Data

                Manufactured by                                                            SPEAG




           Certificate No: Z16—97090                             Page 4 of 8
_i m > c


CCGIS
                                        in Collaboration with

              aa=‘/"/°J, a
               v                        CALIBRATION LABORATORY

            Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
            Tel: +86—10—62304633—2079      Fax: +86—10—62304633—2504
            E—mail: cttl@chinattl.com       Hitp://www.chinattl.en


    DASYS5 Validation Report for Head TSL                                        Date: 06.15.2016
    Test Laboratory: CTTL, Beijing, China
    DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN: 5d175
       Communication System: UID 0, CW; Frequency: 1900 MHz; Duty Cycle: 1:1
       Medium parameters used: f= 1900 MHz; 0 = 1.381 $/m; sr = 40.33; p = 1000 kg/m3
       Phantom section: Right Section
       Measurement Standard: DASY5 (IEEE/IEC/ANSI €63.19—2007)
    DASYS Configuration:

                *    Probe: EX3DV4 — SN7307; ConvF(8.1, 8.1, 8.1); Calibrated: 2/19/2016;
                *    Sensor—Surface: 2mm (Mechanical Surface Detection)
                *    Electronics: DAE4 Sn771; Calibrated: 2/2/2016
|               *    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
|               +    Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7372)

            System Performance Check/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid:
            dx=5mm, dy=5mm, dz=5mm
|           Reference Value = 103.5V/m; Power Drift = 0.00 dB
            Peak SAR (extrapolated) = 18.0W/kg
            SAR(I g) =9.99 W/kg; SAR(10 g) = 5.28 W/kg
            Maximum value of SAR (measured) =14.1 W/kg

               dB
               0


                —3.41


                —6.82


                —10.24


               ~13.65

                                                   a
               —17.06                            fl—_—

                          0 dB =14.1 W/kg =11.49 dBW/kg




    Certificate No: Z16—97090                             Page 5 of 8


CCGIS
                 fi                         I;Coflabomtionewith 3

                 V                         CALIBRATION LABORATORY

               Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
               Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
               E—mail: cttl@chinattl.com       Hitp://www.chinattl.en




               Impedance Measurement Plot for Head TSL


     [ Tri 11 Log Mag 10.00d8/ Ref 0.000d8 [F1]
      |     so.c0      L   1.9000000 onz —24.303 do_
            40. 00
            30. 00

            20.00

            10. 00
            0. 000p)
           —10. 00

           ~20. 00


           ~30. 00


           ~40. 00
           «8B 86 fe..__
          MIR 511 smith (R+jX) scale 1.000U [F1 bel]
            »1 1.9000000 Ghz §3.173 n 5.4373 n 455.46—pH




      [1 Strt 17 GHe                             i              IFSW 100 He        Stop 2.1 GHtc BB




    Certificate No: Z16—97090                                 Page 6 of 8


CCGIS
            i!”?\l"/                   I;Co!laborationewiflu 3


             ‘/                        CALIBRATION LABORATORY
           Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
           Tel: +86—10—62304633—2079      Fax: +86—10—62304633—2504
           E—mail: cttl@chinattl.com       Hitp://www.chinattl.en




    DASYS Validation Report for Body TSL                                       Date: 06.15.2016
    Test Laboratory: CTTL, Beijing, China
    DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN: 50175
       Communication System: UID 0, CW; Frequency: 1900 MHz; Duty Cycle: 1:1
      Medium parameters used: f= 1900 MHz; 0 = 1.537 S/m; s, = 53.34; p = 1000 kg/m3
      Phantom section: Center Section
      Measurement Standard: DASYS (IEEE/IEC/ANSI €63.19—2007)
    DASYS Configuration:

               +    Probe: EX3DV4 — SN7307; ConvF(7.67, 7.67, 7.67); Calibrated: 2/19/2016;
               +    Sensor—Surface: 2mm (Mechanical Surface Detection)
|              *    Electronics: DAE4 Sn771; Calibrated: 2/2/2016
|              +    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1 161/1
               +    Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7372)

           System Performance Check/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid:
           dx=5mm, dy=5mm, dz=5mm
           Reference Value = 96.11 V/m; Power Drift =—0.03 dB
           Peak SAR (extrapolated) = 17.8 W/kg
           SAR(I g) = 10.1 W/kg; SAR(10 g) = 5.39 W/kg
           Maximum value of SAR (measured) = 14.3 W/kg


               dB
               0


                ~3.23


                —6.47


                —9.70


                ~12.94


                1617       |2
                                £.            kassmmmmmmmmmmmgaceeeee
                                                         Tess
                           0 dB = 14.3 W/kg = 11.55 dBW/kg




    Certificate No: Z16—97090                              Page 7 of 8



Document Created: 2018-10-10 15:15:16
Document Modified: 2018-10-10 15:15:16

© 2025 FCC.report
This site is not affiliated with or endorsed by the FCC