2AAUFSPS001 TRpt Qi

FCC ID: 2AAUFSPS001

Test Report

Download: PDF
FCCID_3407297

                                                       FCC TEST REPORT
                                                        FCC 47 CFR Part 15C
                                                           ISED RSS-210
                                               License exempt radio equipment

Report Reference No. .................... : G0M-1702-6295-TFC209LP-MU-V01

Testing Laboratory ........................ : Eurofins Product Service GmbH

Address ............................................ : Storkower Str. 38c
                                                       15526 Reichenwalde
                                                       Germany

Accreditation .................................... :




                                                       FCC Test Firm Designation Number: DE0008
                                                       IC Testing Laboratory site: 3470A-2

Applicant’s name ........................... : eResearchTechnology GmbH

Address ............................................ : Sieboldstrasse 3
                                                       97230 Estenfeld
                                                       GERMANY

Test specification:

Standard........................................... : 47 CFR Part 15C
                                                      RSS-210, Issue 9, 2016-08
Test scope………….. ....................... : complete Radio compliance test
Equipment under test (EUT):

Product description                                    Spirometer

Model No.                                              SpiroSphere - Main Unit

Additional Model(s)                                    None

Brand Name(s)                                          SpiroSphere

Hardware version                                       04.04.03

Firmware / Software version                            Jet_Lib + Test_APP 0.14.0 ERT
                                                       App: sd_SpiroSpherePackage-v1.1.19tgz

                                                       FCC-ID: 2AAUFSPS001            IC: 11335A-SPS001

Test result                                            Passed



                                  Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
                                                   Eurofins Product Service GmbH
                                         Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                                                                      Page 1 of 26


 <= eurofins
                                      Product Service



Possible test case verdicts:

=R@IRGF AES@BSBA NOF I@BIBO ... mcmmesmmmemmcecmmmnece : N/N
— required by standard but not appl. to test object ....... : N/A

— required by standard but not teSted .......................... : NT

— not required by standard for the test object............... : NR

— test object does meet the requirement ..................... : P (Pass)

— test object does not meet the requirement................ : F (Fail)

Testing:

TESt L@BD T@MP@FANUN® ...........s.scses en eneeemmememmmemmeen. : 20 —23 °C

TESt LQD HUMHCY .....2.2000000emem en rerernrereereenrereeemernes : 32— 38 %

Dat@ Of F@C@IPt Of t@St It@M..............clccscemmmeemmcss. : 2017—03—24

Date (s) of performance of tests....                              : 2017—05—02

Compiled ByY ............c.c.clc.ccc.l.l.l......: Wilfried Treffke                         /


Zgj;%‘é:éé;ig”?;fi) =                          Wilfried Treffke                   ...Z/% .....................
Approved by (+ signature)................:
(Head of Lab)                                  Christian
                                                   ristian Weber Weber           ......2l.. Iemsntant frommmnnces

Date of issue .....                         : 2017—05—12

Total number of pages .....................: 26

General remarks:

The test results presented in this report relate only to the objecttested.
The results contained in this report reflect the results for this particular model and serial
number. It is the responsibility of the manufacturer to ensure thatall production models meet
the intent of the requirements detailed within this report.


This report shall not be reproduced, except in full, without the written approval of the Issuing testing
laboratory.

Additional comments:




                             Test Report No.: GOM—1702—6295—TFC209LP—MU—VO01
                                              Eurofins Product Service GmbH
                                   Storkower Str. 38¢, D—15526 Reichenwalde, Germany
                                                                                                     Page 2 of 26


Version History
Version   Issue Date   Remarks                                             Revised by
01        2017-05-12   Initial Release




                   Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
                                 Eurofins Product Service GmbH
                       Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                                            Page 3 of 26


REPORT INDEX


1     EQUIPMENT (TEST ITEM) DESCRIPTION                                                   5

1.1   Photos – Equipment External                                                         6
1.2   Photos – Equipment internal                                                         8
1.3   Photos – Test setup                                                                10
1.4   Supporting Equipment Used During Testing                                           11
1.5   Test Modes                                                                         12
1.6   Test Equipment Used During Testing                                                 13
1.7   Sample emission level calculation                                                  14

2     RESULT SUMMARY                                                                     15

3     TEST CONDITIONS AND RESULTS                                                        16

3.1   Test Conditions and Results – Occupied Bandwidth                                   16
3.2   Test Conditions and Results – Fundamental field strength emissions                 18
3.3   Test Conditions and Results – Receiver radiated emissions                          20

ANNEX A          Transmitter Field Strength Emissions                                    22
ANNEX B          Receiver Radiated Spurious Emissions                                    25




                       Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
                                      Eurofins Product Service GmbH
                            Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                                                Page 4 of 26


1      Equipment (Test item) Description

Description                      Spirometer
Model                            SpiroSphere - Main Unit
Additional Model(s)              None
Brand Name(s)                    SpiroSphere
Serial number                    12567
Hardware version                 04.04.03
                                 Jet_Lib + Test_APP 0.14.0 ERT
Software / Firmware version
                                 App: sd_SpiroSpherePackage-v1.1.19tgz
PMN                              SpiroSphere
HVIN                             SpiroSphere
FVIN                             N/A
HMN                              N/A
FCC-ID                           2AAUFSPS001
IC                               11335A-SPS001
Equipment type                   End product
Radio type                       Transceiver
Radio technology                 custom
Operating frequency range        112 - 205 kHz
Frequency range                  FMID                    128 kHz
Number of channels               1
Channel spacing                  None
Number of antennas               1
                                 Type               integrated
                                 Model              unspecified
Antenna
                                 Manufacturer       eResearch Technology GmbH
                                 Gain               unspecified
                                 eResearchTechnology GmbH
                                 Sieboldstrasse 3
Manufacturer
                                 97230 Estenfeld
                                 GERMANY
                                 VNOM                    120 VAC
Power supply                     VMIN                    N/A
                                 VMIN                    N/A
                                 Model                   RR9KA6000YL4CRVB3046
                                 Vendor                  Globtek
AC/DC-Adaptor
                                 Input                   100 to 240V /50/60Hz
                                 Output                  5V




                      Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
                                      Eurofins Product Service GmbH
                            Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                                                Page 5 of 26


1.1   Photos – Equipment External




                                          EUT FRONT




                                         EUT BOTTOM




                     Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
                                   Eurofins Product Service GmbH
                         Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                                             Page 6 of 26


                  EUT INTERFACE




                      EUT Label




Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
              Eurofins Product Service GmbH
    Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                        Page 7 of 26


1.2   Photos – Equipment internal



                                            PCB TOP




                                              Module




                     Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
                                    Eurofins Product Service GmbH
                          Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                                              Page 8 of 26


                        Battery




                  AE: Sensor Unit




Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
              Eurofins Product Service GmbH
    Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                        Page 9 of 26


1.3   Photos – Test setup




                                           Radiated Setup




                     Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
                                      Eurofins Product Service GmbH
                            Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                                                Page 10 of 26


1.4      Supporting Equipment Used During Testing


Product
                     Device                      Manufacturer           Model No.        Comments
 Type*
                                          eResearchTechnology
   AE              Spirometer                                           SpiroSphere      SensorUnit
                                                GmbH
*Note: Use the following abbreviations:
        AE : Auxiliary/Associated Equipment, or

       SIM : Simulator (Not Subjected to Test)

      CABL : Connecting cables




                              Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
                                               Eurofins Product Service GmbH
                                     Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                                                            Page 11 of 26


1.5     Test Modes


      Mode #                                           Description

                General conditions: EUT powered by DC/DC adaptor
  Charging      Radio conditions:    Mode = wireless charging
                                     Power level = Maximum
                General conditions: EUT powered by DC/DC adaptor
      Standby
                Radio conditions:    Mode = standalone standby




                       Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
                                      Eurofins Product Service GmbH
                            Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                                                Page 12 of 26


1.6   Test Equipment Used During Testing


                                     Measurement Software
    Description               Manufacturer                         Name                    Version
  EMC Test Software          Dare Instruments                    Radimation                2015.2.4

                                      Occupied Bandwidth
   Description        Manufacturer        Model             Identifier         Cal. Date      Cal. Due
Spectrum Analyzer        R&S             FSP 30             EF00312            2017-03        2018-03

                                     Field strength emissions
   Description        Manufacturer        Model             Identifier         Cal. Date      Cal. Due
 Semi-Anechoic
                       Frankonia          AC 1              EF00062            2017-02        2020-02
    chamber
  Loop antenna           R&S            HFH2-Z2             EF00184            2016-12        2018-12
Spectrum Analyzer        R&S             FSP30              EF00312            2017-03        2018-03




                      Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
                                     Eurofins Product Service GmbH
                           Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                                                           Page 13 of 26


1.7      Sample emission level calculation

The following is a description of terms and a sample calculation, as appears in the radiated emissions
data table. The numbers used in the calculation are for example only. There is no direct correlation to
the specific data taken for the product described in this document:

Reading:

This is the reading obtained on the spectrum analyzer in dBµV. Any external preamplifiers used are
taken into account through internal analyzer settings.

A.F.:

This is the antenna factor for the receiving antenna. It is a conversion factor, which converts electric
fields strengths to voltages, which can be measured directly on the spectrum analyzer. It is treated as
a loss in dB. Cable losses have been included with the A.F. to simplify the calculations. The antenna
factor is used in calculations as follows:

                 Reading on Analyzer (dBµV) + A.F. (dB) = Net field strength (dBµV/m)

Net:

This is the net field strength measurement (as shown above).

Limit:

This is the FCC Class B radiated emission limit (in units of dBµV/m). The FCC limits are given in units
of µV/m. The following formula is used to convert the units of µV/m to dBµV/m:

                                    Limit (dBµV/m) = 20*log (µV/m)

Margin:

This is the margin of compliance below the FCC limit. The units are given in dB. A negative margin
indicates the emission was below the limit. A positive margin indicates that the emission exceeds the
limit.

Example only:

          Reading + AF      =       Net Reading        :   Net reading - FCC limit = Margin
          21.5 dBµV + 26 dB =       47.5 dBµV/m        :   47.5 dBµV/m - 57.0 dBµV/m = -9.5 dB




                        Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
                                        Eurofins Product Service GmbH
                              Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                                                          Page 14 of 26


2     Result Summary


                                    FCC 47 CFR Part 15C, ISED RSS-210

    Product Specific                                               Reference
                                  Requirement – Test                                 Result      Remarks
    Standard Section                                                Method

      RSS-Gen 6.6        Occupied Bandwidth                       RSS-Gen 6.6         N/R     Informational only

     FCC 15.209
                         Field strength emissions                 ANSI C63.10        PASS
 ISED RSS-210 4.3, 4.4

    ISED RSS-210 3.1
                         Receiver radiated spurious emissions     ANSI C63.10        PASS
    ISED RSS-Gen 7.1
Remarks:




                          Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
                                           Eurofins Product Service GmbH
                                 Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                                                                  Page 15 of 26


3      Test Conditions and Results

3.1        Test Conditions and Results – Occupied Bandwidth


Occupied Bandwidth acc. to ISED RSS-Gen                                                      Verdict: PASS

             Test according to                                        Reference Method
           measurement reference                                         RSS-Gen 6.6
                                                                      Tested frequencies
            Test frequency range
                                                                              FMID
                  EUT test mode                                            Charging
                                                      Limits
                                            None (Informational only)
                                                    Test setup


                                          Spectrum
                                          Analyzer                          EUT



                                                 Test procedure
      1.   EUT set to test mode (Communication tester is used if needed)
      2.   Span set to at least twice the emission spectrum
      3.   Resolution bandwidth set to 1% to 5% of Ocupied Bandwidth
      4.   Occupied Bandwidth (99 %) measurement with spectrum analyzer built in measurement
           function
                                                   Test results
       Channel            Frequency [kHz]                         Occupied Bandwidth [kHz]
           FMID                   64                                           0.08
Comments:




                             Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
                                            Eurofins Product Service GmbH
                                  Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                                                              Page 16 of 26


                                Occupied Bandwidth - FMID
Occupied Bandwidth acc. to RSS-Gen

Project Number:                  G0M-1702-6295
Applicant                        eResearchTechnology GmbH
Model Description                Spirometer
Model:                           SpiroSphere - Main Unit
Test Sample ID:                  12567
Operator:                        W. Treffke
Test Site:                       Eurofins Product Service GmbH
Test Date:                       2017-05-02
Note 1:                          A spectrum analyzer with an integrated 99% power bandwidth
                                 function is used
Note 2:                          NONE (INFORMATION ONLY)




                    Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
                                  Eurofins Product Service GmbH
                        Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                                                Page 17 of 26


3.2   Test Conditions and Results – Fundamental field strength emissions


Field strength emissions acc. to FCC 47 CFR 15.209 / ISED RSS-210                       Verdict: PASS

       Test according referenced                                  Reference Method
               standards                                 FCC 15.209 / ISED RSS-210 4.3, 4.4

          Test according to                                       Reference Method
        measurement reference                                        ANSI C63.10
                                                                 Tested frequencies
         Test frequency range                                               th
                                                                9 kHz – 10 Harmonic
            EUT test mode                                                Single
                                                Limits
 Frequency range [MHz]       Detector          Limit [µV/m]        Limit [dBµV/m]     Limit Distance [m]
      0.009 – 0.490         Quasi-Peak         2400/F[kHz]           48.5 – 13.8             300
      0.490 – 1.705         Quasi-Peak         2400/F[kHz]            13.8 – 1.4              30
       1.705 – 30           Quasi-Peak              30                   29.5                 30
        30 – 88             Quasi-Peak              100                   40                  3
        88 – 216            Quasi-Peak              150                  43.5                 3
       216 – 960            Quasi-Peak              200                   46                  3
       960 – 1000           Quasi-Peak              500                   54                  3
         > 1000              Average                500                   54                  3
The emission limits shown in the above table are based on measurements employing a CISPR quasi-
peak detector except for the frequency bands 9–90 kHz, 110–490 kHz and above 1000 MHz. Radiated
emission limits in these three bands are based on measurements employing an average detector.




                       Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
                                      Eurofins Product Service GmbH
                            Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                                                          Page 18 of 26


                                               Test setup



                                            Semi-anechoic Chamber



                                                                                  EUT



                                                                             Turn table


                                                 Ground Plane



                       Amplifier           Measurement
                        Matrix              Receiver

                                            Test procedure

   1. EUT set to test mode
   2. Span it set according to measurement range
   3. Resolution bandwidth below 1 GHz is set according to CISPR 16 with peak/quasi-peak detector
      and RBW of 1 MHz with peak/average detector is used above 1 GHz
   4. Markers are set to maximum emission levels

                                               Test results
           Frequency    Emission       Level                      Limit          Measurement
Channel                                            Detector                                        Margin [dB]
              [kHz]      [kHz]       [dBµV/m]                   [dBµV/m]         distance [m]*
  FMID         128      127.746         2.9           av          25.5                 3              -22.60
  FMID        128       392.517        -23.1          av           15.7                 3             -38.84
Comments: * Physical distance between EUT and measurement antenna.




                        Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
                                       Eurofins Product Service GmbH
                             Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                                                                 Page 19 of 26


3.3   Test Conditions and Results – Receiver radiated emissions


Receiver radiated emissions acc. to ISED RSS-210                                                Verdict: PASS

      Test according referenced                                    Reference Method
              standards                                            ISED RSS-210 3.1

         Test according to                                         Reference Method
       measurement reference                                          ANSI C63.10
                                                                  Tested frequencies
        Test frequency range                                                 th
                                                                 30 MHz – 5 Harmonic
           EUT test mode                                                 Standby
                                                   Limits
 Frequency range [MHz]          Detector          Limit [µV/m]         Limit [dBµV/m]          Limit Distance [m]
      0.009 – 0.490         Quasi-Peak            2400/F[kHz]            48.5 – 13.8                  300
      0.490 – 1.705         Quasi-Peak            2400/F[kHz]             13.8 – 1.4                  30
       1.705 – 30           Quasi-Peak                 30                    29.5                     30
         30 – 88            Quasi-Peak                 100                    40                       3
        88 – 216            Quasi-Peak                 150                   43.5                      3
        216 – 960           Quasi-Peak                 200                    46                       3
       960 – 1000           Quasi-Peak                 500                    54                       3
         > 1000                 Average                500                    54                       3
                                                 Test setup



                                              Semi-anechoic Chamber



                                                                                    EUT



                                                                                  Turn table


                                                   Ground Plane



                       Amplifier             Measurement
                        Matrix                Receiver




                       Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
                                         Eurofins Product Service GmbH
                               Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                                                                  Page 20 of 26


                                            Test procedure

   1. EUT set to receive mode (Communication tester is used if needed)
   2. Span it set according to measurement range
   3. Resolution bandwidth below 1 GHz is set according to CISPR 16 with peak/quasi-peak detector
      and RBW of 1 MHz with peak/average detector is used above 1 GHz
   4. Markers are set to peak emission levels

                                             Test results
            Frequency   Emission    Emission Level                                 Limit         Margin
Channel                                                   Det.
              [kHz]      [kHz]        [dbµV/m]                                   [dBdµV/m]     [dBµV/m]
  FMID         N/A      178.531          -4.9              av                       22.6         -27.48
Comments:




                        Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
                                       Eurofins Product Service GmbH
                             Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                                                             Page 21 of 26


  ANNEX A                                     Transmitter Field Strength Emissions
  Carrier according to FCC 15.209

  Project number: G0M-1702-6295

  Applicant:                                                        eResearchTechnology GmbH
  EUT Name:                                                         Spirometer
  Model:                                                            SpiroSphere - Main Unit
  Test Site:                                                        Eurofins Product Service GmbH
  Operator:                                                         W. Treffke
  Test Conditions:                                                  Tnom: 20°C, Vnom: 120 VAC
  Antenna:                                                          Rohde & Schwarz HFH 2-Z2
  Measurement distance:                                             3 m converted to 300 m
  Mode:                                                             TX; Charging
  Test Date:                                                        2017-05-02
  Note:                                                             carrier
                                                                                                                                                 Index 15


                                     FCC 15.209 AV2         RBW: 200 Hz, Vertical Max Average



                            40




                            20


                                                                                         1
                             0
Electrical Field (dBµV/m)




                            -20




                            -40




                            -60




                            -80



                             126 k         126.5 k          127 k            127.5 k            128 k     128.5 k       129 k     129.5 k        130 k

                                                                                         Frequency (Hz)




  Frequency                                           Average                     Average Limit            Average Difference   Average Status
  127.746 kHz                                         2.9 dBµV/m                  25.5 dBµV/m              -22.61 dB            Pass




                                                         Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
                                                                          Eurofins Product Service GmbH
                                                                Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                                                                                                       Page 22 of 26


  Spurious emissions according to FCC 15.209

  Project number: G0M-1702-6295

  Applicant:                                                       eResearchTechnology GmbH
  EUT Name:                                                        Spirometer
  Model:                                                           SpiroSphere - Main Unit
  Test Site:                                                       Eurofins Product Service GmbH
  Operator:                                                        W. Treffke
  Test Conditions:                                                 Tnom: 20°C, Vnom: 120 VAC
  Antenna:                                                         Rohde & Schwarz HFH 2-Z2
  Measurement distance:                                            3 m converted to 300 m
  Mode:                                                            TX; Charging
  Test Date:                                                       2017-05-02
  Note:
                                                                                                                                                                     Index 8


                                   FCC 15.209 AV1            FCC 15.209 AV2         FCC 15.209 QP1              RBW: 9 kHz, Vertical Max Average
                                   RBW: 200 Hz, Vertical Max Average          RBW: 200 Hz, Vertical Max Peak



                            60



                            40



                            20
Electrical Field (dBµV/m)




                                                                                                                          1
                             0



                            -20                                                                                                                               2



                            -40



                            -60



                            -80



                              9k                     20 k          30 k              50 k                      100 k                   200 k         300 k          490 k

                                                                                         Frequency (Hz)




  Frequency                                        Average                        Average Limit                   Average Difference               Average Status
  130.979 kHz                                      1.7 dBµV/m                     25.3 dBµV/m                     -23.62 dB                        Pass
  392.517 kHz                                      -23.1 dBµV/m                   15.7 dBµV/m                     -38.84 dB                        Pass




                                                        Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
                                                                          Eurofins Product Service GmbH
                                                                Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                                                                                                                        Page 23 of 26


  Spurious emissions according to FCC 15.209

  Project number: G0M-1702-6295

  Applicant:                                                      eResearchTechnology GmbH
  EUT Name:                                                       Spirometer
  Model:                                                          SpiroSphere - Main Unit
  Test Site:                                                      Eurofins Product Service GmbH
  Operator:                                                       W. Treffke
  Test Conditions:                                                Tnom: 20°C, Vnom: 120 VAC
  Antenna:                                                        Rohde & Schwarz HFH 2-Z2
  Measurement distance:                                           3 m converted to 30 m
  Mode:                                                           TX; Charging
  Test Date:                                                      2017-05-02
  Note:
                                                                                                                             Index 9


                                     FCC 15.209 QP         FCC 15.209 QP2     RBW: 9 kHz, Vertical Max Peak



                            60



                            50



                            40
Electrical Field (dBµV/m)




                            30



                            20



                            10



                             0



                            -10



                            -20



                             490 k                   1M                 2M      3M                 5M         10 M   20 M    30 M

                                                                                  Frequency (Hz)




                                                          Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
                                                                         Eurofins Product Service GmbH
                                                               Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                                                                                     Page 24 of 26


  ANNEX B                                   Receiver Radiated Spurious Emissions
  Spurious emissions according to FCC 15.209

  Project number: G0M-1702-6295

  Applicant:                                                       eResearchTechnology GmbH
  EUT Name:                                                        Spirometer
  Model:                                                           SpiroSphere - Main Unit
  Test Site:                                                       Eurofins Product Service GmbH
  Operator:                                                        W. Treffke
  Test Conditions:                                                 Tnom: 20°C, Vnom: 120 VAC
  Antenna:                                                         Rohde & Schwarz HFH 2-Z2
  Measurement distance:                                            3 m converted to 300 m
  Mode:                                                            TX; no charging
  Test Date:                                                       2017-05-02
  Note:                                                            Standby
                                                                                                                                                                    Index 10


                                   FCC 15.209 AV1            FCC 15.209 AV2         FCC 15.209 QP1              RBW: 9 kHz, Vertical Max Average
                                   RBW: 200 Hz, Vertical Max Average          RBW: 200 Hz, Vertical Max Peak



                            60



                            40



                            20
Electrical Field (dBµV/m)




                             0                                                                                                       1



                            -20



                            -40



                            -60



                            -80



                              9k                     20 k          30 k              50 k                      100 k                     200 k       300 k          490 k

                                                                                         Frequency (Hz)




  Frequency                                        Average                        Average Limit                   Average Difference               Average Status
  178.531 kHz                                      -4.9 dBµV/m                    22.6 dBµV/m                     -27.48 dB                        Pass




                                                        Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
                                                                          Eurofins Product Service GmbH
                                                                Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                                                                                                                        Page 25 of 26


  Spurious emissions according to FCC 15.209

  Project number: G0M-1702-6295

  Applicant:                                                 eResearchTechnology GmbH
  EUT Name:                                                  Spirometer
  Model:                                                     SpiroSphere - Main Unit
  Test Site:                                                 Eurofins Product Service GmbH
  Operator:                                                  W. Treffke
  Test Conditions:                                           Tnom: 20°C, Vnom: 120 VAC
  Antenna:                                                   Rohde & Schwarz HFH 2-Z2
  Measurement distance:                                      3 m converted to 30 m
  Mode:                                                      TX; Charging
  Test Date:                                                 2017-05-02
  Note:                                                      Standby
                                                                                                                                    Index 12


                                     FCC 15.209 QP    FCC 15.209 QP2     RBW: 9 kHz, Vertical Max Peak



                            60



                            50



                            40
Electrical Field (dBµV/m)




                            30



                            20



                            10



                             0



                            -10



                            -20



                             490 k           600 k               800 k            1M               1.2 M   1.4 M   1.6 M   1.8 M    2.1 M

                                                                             Frequency (Hz)




                                                     Test Report No.: G0M-1702-6295-TFC209LP-MU-V01
                                                                    Eurofins Product Service GmbH
                                                          Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                                                                                            Page 26 of 26



Document Created: 2017-05-18 16:14:35
Document Modified: 2017-05-18 16:14:35

© 2024 FCC.report