EMC Test report

FCC ID: 2AARU-FM008

Test Report

Download: PDF
FCCID_4053650

EMC TEST
REPORT
Report No.: 18070636-FCC-E
Supersede Report No: N/A
Applicant          Medtrum Technologies Inc.
Product Name       Personal Diabetes Manager
Model No.          MD-FM-008
Serial No.         N/A
Test Standard      FCC Part 15 Subpart B Class B, ANSI C63.4: 2014
Test Date          August 03 to September 05, 2018
Issue Date         September 07, 2018
Test Result             Pass      Fail
Equipment complied with the specification
Equipment did not comply with the specification




            Evans He                     David Huang
       Test Engineer                     Checked By

                       This test report may be reproduced in full only
     Test result presented in this test report is applicable to the tested sample only


                                          Issued by:
                       SIEMIC (SHENZHEN-CHINA) LABORATORIES
                   Zone A, Floor 1, Building 2 Wan Ye Long Technology Park
       South Side of Zhoushi Road, Bao’ an District, Shenzhen, Guangdong China 518108
                  Phone: +86 0755 2601 4629801 Email: China@siemic.com.cn


                                          Test Report       18070636-FCC-E
                                          Page
                                          No.               2 of 24



                             Laboratories Introduction

SIEMIC, headquartered in the heart of Silicon Valley, with superior facilities in US and Asia, is one
of the leading independent testing and certification facilities providing customers with one-stop
shop services for Compliance Testing and Global Certifications.




In addition to testing and certification, SIEMIC provides initial design reviews and compliance
management throughout a project. Our extensive experience with China, Asia Pacific, North
America, European, and International compliance requirements, assures the fastest, most cost
effective way to attain regulatory compliance for the global markets.


                       Accreditations for Conformity Assessment
            Country/Region                                            Scope
                   USA                            EMC, RF/Wireless, SAR, Telecom
                 Canada                           EMC, RF/Wireless, SAR, Telecom
                  Taiwan                           EMC, RF, Telecom, SAR, Safety
               Hong Kong                                RF/Wireless, SAR, Telecom
                 Australia                         EMC, RF, Telecom, SAR, Safety
                  Korea                          EMI, EMS, RF, SAR, Telecom, Safety
                  Japan                            EMI, RF/Wireless, SAR, Telecom
                Singapore                                EMC, RF, SAR, Telecom
                  Europe                           EMC, RF, SAR, Telecom, Safety


                                      Test Report   18070636-FCC-E
                                      Page
                                      No.           3 of 24




This page has been left blank intentionally.


                                                                         Test Report                18070636-FCC-E
                                                                         Page
                                                                         No.                        4 of 24




CONTENTS

1.      REPORT REVISION HISTORY.............................................................................................................................5

2.      CUSTOMER INFORMATION ................................................................................................................................5

3.      TEST SITE INFORMATION ...................................................................................................................................5

4.      EQUIPMENT UNDER TEST (EUT) INFORMATION .........................................................................................6

5.      TEST SUMMARY ....................................................................................................................................................8

6.      MEASUREMENTS, EXAMINATION AND DERIVED RESULTS .....................................................................9

6.1 AC POWER LINE CONDUCTED EMISSIONS .....................................................................................................9

6.2 RADIATED EMISSIONS ......................................................................................................................................... 15

ANNEX A. TEST INSTRUMENT ................................................................................................................................. 20

ANNEX B. TEST SETUP AND SUPPORTING EQUIPMENT ................................................................................ 21

ANNEX C. USER MANUAL / BLOCK DIAGRAM / SCHEMATICS / PARTLIST/ DECLARATION OF
SIMILARITY..................................................................................................................................................................... 24


                                            Test Report   18070636-FCC-E
                                            Page
                                            No.           5 of 24




1. Report Revision History

         Report No.            Report Version             Description                 Issue Date
      18070636-FCC-E                 NONE                  Original              September 07, 2018




2. Customer information


Applicant Name          Medtrum Technologies Inc.
Applicant Add           7F, Building 8, No.200 Niudun Road, Shanghai 201203, China
Manufacturer            Medtrum Technologies Inc.
Manufacturer Add        7F, Building 8, No.200 Niudun Road, Shanghai 201203, China



3. Test site information
 Lab performing tests       SIEMIC (Shenzhen-China) LABORATORIES
                            Zone A, Floor 1, Building 2 Wan Ye Long Technology Park
 Lab Address                South Side of Zhoushi Road, Bao’ an District, Shenzhen, Guangdong China
                            518108
 FCC Test Site No.          535293
 IC Test Site No.           4842E-1
 Test Software of
                            Radiated Emission Program-To Shenzhen v2.0
 Radiated Emission
 Test Software of
                            EZ-EMC(ver.Icp-03A1)
 Conducted Emission


                                           Test Report   18070636-FCC-E
                                           Page
                                           No.           6 of 24




 4. Equipment under Test (EUT) Information
Description of EUT:             Personal Diabetes Manager



Main Model:                     MD-FM-008


Serial Model:                   N/A


Antenna Gain:                   BLE: 1.6dBi


Antenna Type:                   Chip antenna


                                Adapter:
                                Model: UES06WNCPU-050100SPA
                                Input: AC 100-240V~50/60Hz, 0.2A
                                Output: DC 5.0V, 1.0A
Input Power:                    Battery:
                                Model: RC2
                                Voltage: 3.8V
                                Capacity: 490mAh(1.9Wh)
                                Maximum Charge Voltage: 4.35V


Equipment Category:             JAB

Type of Modulation:             BLE: GFSK



RF Operating Frequency (ies):   BLE: 2402-2480 MHz


Number of Channels:             BLE: 40CH


Port:                           Pls refer to the user’ s manual


Trade Name :                    Medtrum



FCC ID:                         2AARU-FM008


Date EUT received:              August 02, 2018


                       Test Report    18070636-FCC-E
                       Page
                       No.            7 of 24

Test Date(s):   August 03 to September 05, 2018


                                          Test Report     18070636-FCC-E
                                          Page
                                          No.             8 of 24




 5. Test Summary

The product was tested in accordance with the following specifications.
All testing has been performed according to below product classification:


         FCC Rules                               Description of Test                        Result
 §15.107; ANSI C63.4: 2014     AC Power Line Conducted Emissions                          Compliance
 §15.109; ANSI C63.4: 2014     Radiated Emissions                                         Compliance


Measurement Uncertainty

                     Parameter                                              Uncertainty
        AC Power Line Conducted Emissions
                                                                             ±3.11dB
                  (150kHz~30MHz)
         Radiated Emission(30MHz~1GHz)                                       ±5.12dB
          Radiated Emission(1GHz~6GHz)                                       ±5.34dB


                                             Test Report           18070636-FCC-E
                                             Page
                                             No.                   9 of 24




 6. Measurements, Examination And Derived Results

6.1 AC Power Line Conducted Emissions

Temperature                                                25 oC
Relative Humidity                                          57%
Atmospheric Pressure                                       1019mbar
Test date :                                                August 30, 2018
Tested By :                                                Evans He


Requirement(s):

Spec              Item    Requirement                                                            Applicable
                          For Low-power radio-frequency devices that is designed to be
                          connected to the public utility (AC) power line, the radio frequency
                          voltage that is conducted back onto the AC power line on any
                          frequency or frequencies, within the band 150 kHz to 30 MHz, shall
                          not exceed the limits in the following table, as measured using a 50
47CFR§15.                 [mu] H/50 ohms line impedance stabilization network (LISN). The
                    a)
     107                  lower limit applies at the boundary between the frequencies ranges.
                            Frequency ranges                          Limit (dBµV)
                                  (MHz)                      QP                      Average
                                0.15 ~ 0.5              66 – 56                      56 – 46
                                  0.5 ~ 5                    56                        46
                                  5 ~ 30                     60                        50




Test Setup




                  1. The EUT and supporting equipment were set up in accordance with the requirements of
                     the standard on top of a 1.5m x 1m x 0.8m high, non-metallic table.
 Procedure
                  2. The power supply for the EUT was fed through a 50Ω /50mH EUT LISN, connected to
                     filtered mains.


                                          Test Report     18070636-FCC-E
                                          Page
                                          No.             10 of 24


            3. The RF OUT of the EUT LISN was connected to the EMI test receiver via a low-loss
                  coaxial cable.
            4. All other supporting equipment were powered separately from another main supply.
            5. The EUT was switched on and allowed to warm up to its normal operating condition.
            6. A scan was made on the NEUTRAL line (for AC mains) or Earth line (for DC power)
                  over the required frequency range using an EMI test receiver.
            7. High peaks, relative to the limit line, The EMI test receiver was then tuned to the
                  selected frequencies and the necessary measurements made with a receiver bandwidth
                  setting of 10 kHz.
            8. Step 7 was then repeated for the LIVE line (for AC mains) or DC line (for DC power).

  Remark

   Result      Pass                    Fail



Test Data   Yes                          N/A

Test Plot   Yes (See below)              N/A


                                          Test Report   18070636-FCC-E
                                          Page
                                          No.           11 of 24




Test Mode :        Charging Mode




Test Data


                                Phase Line Plot at 120Vac, 60Hz
     No.    P/L     Frequency   Reading      Detector   Corrected   Result   Limit    Margin

                      (MHz)      (dBuV)                   (dB}      (dBuV)   (dBuV)    (dB)

      1       L1     0.2358      23.16           QP      10.03      33.19    62.24    -29.05
      2       L1     0.2358       -1.06          AVG     10.03       8.97    52.24    -43.27
      3       L1     0.3411      20.97           QP      10.03      31.00    59.18    -28.18
      4       L1     0.3411       -2.26          AVG     10.03       7.77    49.18    -41.41
      5       L1     0.6102      16.51           QP      10.03      26.54    56.00    -29.46
      6       L1     0.6102       -3.86          AVG     10.03       6.17    46.00    -39.83
      7       L1     1.1094      13.62           QP      10.03      23.65    56.00    -32.35
      8       L1     1.1094       -4.93          AVG     10.03       5.10    46.00    -40.90
      9       L1     2.4666      12.66           QP      10.05      22.71    56.00    -33.29
      10      L1     2.4666       -4.91          AVG     10.05       5.14    46.00    -40.86
      11      L1     6.5256      12.27           QP      10.10      22.37    60.00    -37.63
      12      L1     6.5256       -4.33          AVG     10.10       5.77    50.00    -44.23


                                           Test Report   18070636-FCC-E
                                           Page
                                           No.           12 of 24




Test Mode :       Charging Mode




Test Data


                               Phase Neutral Plot at 120Vac, 60Hz

     No.    P/L    Frequency     Reading      Detector   Corrected   Result   Limit    Margin

                     (MHz)       (dBuV)                    (dB}      (dBuV)   (dBuV)    (dB)

      1       N      0.1851       21.92           QP       10.02     31.94    64.25    -32.31
      2       N      0.1851       -1.30           AVG      10.02      8.72    54.25    -45.53
      3       N      0.3333       22.80           QP       10.02     32.82    59.37    -26.55
      4       N      0.3333       -2.14           AVG      10.02      7.88    49.37    -41.49
      5       N      0.6024       19.37           QP       10.02     29.39    56.00    -26.61
      6       N      0.6024       -2.91           AVG      10.02      7.11    46.00    -38.89
      7       N      1.0626       13.41           QP       10.03     23.44    56.00    -32.56
      8       N      1.0626       -4.56           AVG      10.03      5.47    46.00    -40.53
      9       N      2.3379       16.10           QP       10.04     26.14    56.00    -29.86
      10      N      2.3379       -4.34           AVG      10.04      5.70    46.00    -40.30
      11      N      3.8502       14.94           QP       10.06     25.00    56.00    -31.00
      12      N      3.8502       -4.46           AVG      10.06      5.60    46.00    -40.40


                                           Test Report   18070636-FCC-E
                                           Page
                                           No.           13 of 24




Test Mode :        Charging Mode




Test Data


                                Phase Line Plot at 240Vac, 60Hz
     No.    P/L     Frequency    Reading      Detector   Corrected   Result   Limit    Margin

                      (MHz)      (dBuV)                     (dB}     (dBuV)   (dBuV)    (dB)

      1       L1      0.1812       17.06          QP       10.03      27.09   64.43    -37.34
      2       L1      0.1812       -1.71          AVG      10.03      8.32    54.43    -46.11
      3       L1      0.3333       17.35          QP       10.03      27.38   59.37    -31.99
      4       L1      0.3333       -0.94          AVG      10.03      9.09    49.37    -40.28
      5       L1      0.5517       16.59          QP       10.03      26.62   56.00    -29.38
      6       L1      0.5517       -1.27          AVG      10.03      8.76    46.00    -37.24
      7       L1      1.1523       14.29          QP       10.03      24.32   56.00    -31.68
      8       L1      1.1523       -2.84          AVG      10.03      7.19    46.00    -38.81
      9       L1      1.8036       13.77          QP       10.04      23.81   56.00    -32.19
      10      L1      1.8036       -4.15          AVG      10.04      5.89    46.00    -40.11
      11      L1      2.8293       12.54          QP       10.05      22.59   56.00    -33.41
      12      L1      2.8293       -4.50          AVG      10.05      5.55    46.00    -40.45


                                           Test Report    18070636-FCC-E
                                           Page
                                           No.            14 of 24




Test Mode :       Charging Mode




Test Data


                               Phase Neutral Plot at 240Vac, 60Hz

     No.    P/L    Frequency     Reading       Detector   Corrected    Result     Limit    Margin

                     (MHz)        (dBuV)                     (dB}     (dBuV)      (dBuV)    (dB)

      1       N      0.2319       23.29           QP        10.02      33.31      62.38    -29.07
      2       N      0.2319        4.65           AVG       10.02      14.67      52.38    -37.71
      3       N      0.3138       24.09           QP        10.02      34.11      59.87    -25.76
      4       N      0.3138        4.83           AVG       10.02      14.85      49.87    -35.02
      5       N      0.4464       22.32           QP        10.02      32.34      56.94    -24.60
      6       N      0.4464        2.68           AVG       10.02      12.70      46.94    -34.24
      7       N      1.1562       20.86           QP        10.03      30.89      56.00    -25.11
      8       N      1.1562        -0.33          AVG       10.03          9.70   46.00    -36.30
      9       N      1.9284       19.63           QP        10.04      29.67      56.00    -26.33
      10      N      1.9284        -1.17          AVG       10.04          8.87   46.00    -37.13
      11      N      3.8619       14.95           QP        10.06      25.01      56.00    -30.99
      12      N      3.8619        -2.95          AVG       10.06          7.11   46.00    -38.89


                                                  Test Report           18070636-FCC-E
                                                  Page
                                                  No.                   15 of 24




6.2 Radiated Emissions

Temperature                                                     25 oC
Relative Humidity                                               57%
Atmospheric Pressure                                            1019mbar
Test date :                                                     August 30, 2018
Tested By :                                                     Evans He


Requirement(s):

Spec              Item       Requirement                                                               Applicable
                             Except higher limit as specified elsewhere in other section, the
                             emissions from the low-power radio-frequency devices shall not
                             exceed the field strength levels specified in the following table and
                             the level of any unwanted emissions shall not exceed the level of
                             the fundamental emission. The tighter limit applies at the band
47CFR§15.
                       a)    edges
   109(d)                          Frequency range (MHz)                     Field Strength (µV/m)
                                            30 – 88                                  100
                                            88 – 216                                 150
                                            216 - 960                                200
                                           Above 960                                 500




Test Setup




                  1.        The EUT was switched on and allowed to warm up to its normal operating condition.
                  2.        The test was carried out at the selected frequency points obtained from the EUT
                            characterization. Maximization of the emissions, was carried out by rotating the EUT,
Procedure
                            changing the antenna polarization, and adjusting the antenna height in the following
                            manner:
                            a.        Vertical or horizontal polarization (whichever gave the higher emission level


                                       Test Report      18070636-FCC-E
                                       Page
                                       No.              16 of 24


                           over a full rotation of the EUT) was chosen.
                    b.     The EUT was then rotated to the direction that gave the maximum
                           emission.
                   c.      Finally, the antenna height was adjusted to the height that gave the maximum
                           emission.
            3.     The resolution bandwidth and video bandwidth of test receiver/spectrum analyzer is
                   120 kHz for Quasiy Peak detection at frequency below 1GHz.
            4.     The resolution bandwidth of test receiver/spectrum analyzer is 1MHz and video
                   bandwidth is 3MHz with Peak detection for Peak measurement at frequency above
                   1GHz.
                    The resolution bandwidth of test receiver/spectrum analyzer is 1MHz and the video
                    bandwidth with Peak detection for Average Measurement as below at frequency
                    above 1GHz.
                    ■ 1 kHz (Duty cycle < 98%) □ 10 Hz (Duty cycle > 98%)
            5.     Steps 2 and 3 were repeated for the next frequency point, until all selected frequency
                   points were measured.

  Remark

   Result        Pass                Fail



Test Data   Yes                        N/A

Test Plot   Yes (See below)            N/A


                                            Test Report     18070636-FCC-E
                                            Page
                                            No.             17 of 24




Test Mode :       Charging Mode

Below 1GHz




Test Data
                                    Horizontal Polarity Plot @3m
No.   P/L   Frequency   Reading    Ant_F    PA_G    Cab_L     Result   Limit    Margin   Height   Degree

                                                                       (dBuV/
              (MHz)     (dBuV/m)   (dB/m)   (dB)     (dB)   (dBuV/m)             (dB)    (cm)       (º)
                                                                         m)

1     H     104.5361     29.87     11.19    22.33    1.14    19.87     43.50    -23.63   100       126

2     H     43.8119      29.46     11.38    22.29    0.76    19.31     40.00    -20.69   100       174

3     H     133.6188     28.48     13.01    22.39    1.23    20.33     43.50    -23.17   100       353

4     H     202.8104     31.71     12.06    22.37    1.55    22.95     43.50    -20.55   100       346

5     H     790.6188     28.88     21.29    21.17    2.94    31.94     46.00    -14.06   100       213

6     H     378.5843     25.89     15.25    22.07    2.02    21.09     46.00    -24.91   100       74


                                            Test Report     18070636-FCC-E
                                            Page
                                            No.             18 of 24


Below 1GHz




Test Data
                                      Vertical Polarity Plot @3m
No.   P/L   Frequency   Reading    Ant_F    PA_G    Cab_L    Result    Limit    Margin   Height   Degree

                                                                       (dBuV/
              (MHz)     (dBuV/m)   (dB/m)   (dB)     (dB)   (dBuV/m)             (dB)    (cm)       (º)
                                                                         m)

1     V     52.9453      40.20     8.08     22.39   0.79     26.68     40.00    -13.32   100       78

2     V     61.1316      40.29     7.36     22.41   0.78     26.02     40.00    -13.98   100       292

3     V     118.1862     35.54     13.58    22.36   1.16     27.92     43.50    -15.58   100       320

4     V     80.0806      34.21     7.60     22.42   1.05     20.44     40.00    -19.56   100       68

5     V     153.2004     42.31     12.60    22.32   1.36     33.95     43.50    -9.55    100       346

6     V     261.0583     34.09     11.89    22.29   1.72     25.41     46.00    -20.59   100       204


                                                Test Report     18070636-FCC-E
                                                Page
                                                No.             19 of 24



Above 1GHz


 Frequency   Read_level                Height      Polarity    Level         Factors     Limit    Margin    Detector
                          Azimuth
   (MHz)      (dBµV/m)                  (cm)        (H/V)     (dBµV/m)        (dB)     (dBµV/m)      (dB)   (PK/AV)


 1376.95       67.22        175         100            V      -18.73         48.01       74       -25.99      PK


 3374.34       69.97         3          100            V      -12.98         47.22       74       -26.78      PK


 1261.62       65.54        294         100            V      -20.15         47.03       74       -26.97      PK


 1868.58       64.49        317         100            H      -15.55         43.21       74       -30.79      PK


 1908.64       63.13        192         100            H      -14.87         47.73       74       -26.27      PK


 3504.61       61.44        105         100            H      -13.29         46.01       74       -27.99      PK




 Note1: The highest frequency of the EUT is 2480 MHz, so the testing has been conformed to 5*2480MHz

 =12,400MHz.

Note2: The frequency that above 3GHz is mainly from the environment noise.
Note3: The AV measurement performed, more than 20dB below limit so AV test data was not presented.


                                    Test Report     18070636-FCC-E
                                    Page
                                    No.             20 of 24




Annex A. TEST INSTRUMENT


       Instrument         Model              Serial #           Cal Date    Cal Due    In use

AC Line Conducted Emissions
   EMI test receiver     ESCS30            8471241027          09/15/2017 09/14/2018
   Line Impedance
                         LI-125A             191106            09/23/2017 09/22/2018
 Stabilization Network
   Line Impedance
                         LI-125A             191107            09/23/2017 09/22/2018
 Stabilization Network
         LISN            ISN T800            34373             09/23/2017 09/22/2018
   Transient Limiter     LIT-153             531118            08/29/2018 08/28/2019
Radiated Emissions
   EMI test receiver      ESL6               100262            09/15/2017 09/14/2018
OPT 010 AMPLIFIER
                          8447E            2727A02430          08/29/2018 08/28/2019
    (0.1-1300MHz)
Microwave Preamplifier
                          8449B            3008A02402          03/22/2018 03/21/2019
    (1~26.5GHz)
    Bilog Antenna
                           JB6              A110712            09/19/2017 09/18/2018
   (30MHz~6GHz)
 Double Ridge Horn
                         AH-118              71259             09/22/2017 09/21/2018
       Antenna


                                     Test Report     18070636-FCC-E
                                     Page
                                     No.             21 of 24




Annex B. TEST SETUP AND SUPPORTING EQUIPMENT


Annex B.ii.   TEST SET UP BLOCK


Block Configuration Diagram for Conducted Emissions



    LISN 1
    V=120V/240V
    F=60Hz




                           Adapter
                                               EUT




                                                                       Test Table


                                                                      80cm above
                                                                      ground plane


                                     Test Report     18070636-FCC-E
                                     Page
                                     No.             22 of 24




Block Configuration Diagram for Radiated Emissions




                                               EUT
                           Adapter                                     Test Table


                                                                      80cm above
                                                                      ground plane




                                                        d=3meter




                                       Receiving Antenna


                                            Test Report      18070636-FCC-E
                                            Page
                                            No.              23 of 24




Annex B. ii. SUPPORTING EQUIPMENT DESCRIPTION

The following is a description of supporting equipment and details of cables used with the EUT.


Supporting Equipment:



                                    Equipment
       Manufacturer                                            Model                     Serial No
                                    Description

   Medtrum Technologies                                   UES06WNCPU-
                                      Adaptor                                               N/A
            Inc.                                            050100SPA


Supporting Cable:

    Cable type           Shield Type         Ferrite Core         Length                 Serial No


    USB Cable             Un-shielding             No              0.8m                      N/A


                             Test Report   18070636-FCC-E
                             Page
                             No.           24 of 24




Annex C. User Manual / Block Diagram / Schematics / Partlist/
DECLARATION OF SIMILARITY



Please see the attachment



Document Created: 2019-03-27 17:19:01
Document Modified: 2019-03-27 17:19:01

© 2024 FCC.report