PROBE CALIBRATION CERTIFICATES


Bay Area Compliance Laboratories Corp. (Kunshan)


Bay Area Compliance Laboratories Corp. (Kunshan)




        EX3DV4 — SN:7441                                                                 November 15, 2016




                              Probe EX3DV4

                                                  SN:7441

                                          Manufactured:             May 31, 2016
                                          Calibrated:               November 15, 2016



                                         Galibrated for DASY/EASY Systems
                                            (Note: non—compaltible with DASY2 system!)




        Certificate No: EX4—7441_Now16                     Page 3 of 11


Bay Area Compliance Laboratories Corp. (Kunshan)




        EX3DV4— SN:7441                                                                                                             November 15, 2016



        DASY/EASY — Parameters of Probe: EX3DV4 — SN:7441

        Basic Calibration Parameters
                                                        Sensor X                          Sensor Y                    SensorZ               Unc (k=2)
         Norm {pwVi(Wim}®)*                                 0.40                           045                           0.36                £10.1 %
         DCP ({mV)"                                        100.2                           101.2                         104.2


        Modulation Calibration Parameters
                                                                                    A           B             C
                                                                                    ds        dBy
                                                                                    0.0                       1
                                                                                    6.           8.0          1.0
                                                                                    00           0.0          1.0


         The reported uncertainty of measurement is stated as the standard uncertainty of measurement
         multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
         probability of approximately 95% .

        * The uncertainties of Norm X.¥2 do not affect the E"—feld uncertainty inside TSL {see Pages 5 and 6.
        * Nurnerical linearization parameter: uncertainty mot required
        © Uncertainty is determined using the max. deviation from lingar response aprplying rectangular distributlion and is expressed for the square of the
        fald valug.




        Cerificate No: EX3—7441_Mov1i6                                      Page 4 of 11


Bay Area Compliance Laboratories Corp. (Kunshan)




        EX3DV4— SN:7441                                                                                                              Nowembsar 15, 2016




        DASY/EASY — Parameters of Probe: EX3DV4 — SN:7441

        Calibration Parameter Determined in Head Tissue Simulating Media

             f{(MHz) ©       Pa:Rr:ll:IIII:i'tyF   cm[gfi pity          ComE X         ComvFY         ConvFZ         flp_l!gi      D{.r:r:          (22;]

                150                52.3                0.76              13.25          13.25          13.25          0.00         1.00         13.3 %

                450                43.5                0.47              10.98          10.98          10.98          0.02         1.20         13.3 %

                750                41.9                0.89              10.73          10.73          10.73          0.53         0.85         £12.0%

                900                41.5                0.97              10.22          10.22          10.22          0.41         0.96         £12.0 %

               1750                40.1                 1.37             8.92           8.92           8.92           0.35         0.80         £12.0 %

               1900                40.0                 140              848            8.48           8.48           0.33         0.80         £12.0 %

               2450                39.2                 1.80              7.85          7.85           7.85           0.35         0.84         + 12.0 %

               2600                39.0                 1.96              7.53          7.53           7.53           0.36         0.80         & 12.0 %

               5250                35.9                4.71               5.67           5.67          5.67           0.35         1.80         £13.1 %

                5600               35.5                 5.07              5.00           5.00          5.00           0.40         1.80         @13.1 %

                5800               35.3                 5.27              5.11           5.11          5.11           040          1.80         #13.1 %


        © Frequency valicity above 300 MHz of + 400 MHz only applies for DASY v4.4 and higher (see Pagee 2), else it is restricted to + 50 MHz. The
        uncertainty is the RSS of the ConvF uncertainty at callbration: frequenty and the uncertainty for the indicaited frequency band. Frequancy validity
        below 300 MHz is + 10. 25, 40, 50 and 70 MMz for Convr assessments at 30, 64, 128, 150 and 220 MHz respectively. Above 5 GHz freq uency
        validity can be extanded to £ 110 MHz.
        " Aj Hroque ncles below 3 GHz, the validity of ltissue paremaeters (c and «) can be relaxe«! to + 10% if liquid compansation formule is applied to
        measured SAR values. At frequencies sbove 3 GHz, the validity of tissue parameters (e and a) is restricted to £ 5%. The uncertainty is the RSS of
        Ehac‘-onwF uncertainly for indicated Nerget fissue paremetars.                                                                             .
          Alpha/Dapth are determined during callbration.. SPEAG warrants that the remaining dovation due to the boundary affect afer compensation is
        always less than + 1% tor frequencies below 3. GHz and below + 2% for frequencies batween 3—6 GHz al any distance larger than hall the probe tip
        diameter from the bourrdary.




        Certificabe No: EX3—7441_Novi6                                       Page S of 11


Bay Area Compliance Laboratories Corp. (Kunshan)




        EX3DVA4— SN:7441                                                                                                         November 15, 2016




        DASY/EASY — Parameters of Probe: EX3DV4 — SN:7441

        Calibration Parameter Determined in Body Tissue Simulating Media
             f(MHz)®         Pa;nl:'lligt;f      cm;:mi‘fl"           ComFX         ConvEY         ConvEZ        Aipha®       "(:.;'L".u      l'i:::}
                150               61.9                0.80              12.58        12.58         12.58          0.00         1.00        #133%
                450               56.7                01.94             12.08        12.08         12.08          0.06         1.20        *13.3 %
                750               55.5                0.96              10.12        10.12         10.12          0.53         0.80        *12.0%
                800               55.0                1.05              9.85         9.85           5.85          0.44         0.80        #12.0 %
               17580              53.4                1.49              8.25         8.25           £.25          0.37         0.80        *120%
               1300               53.3                1.52              7.95         7.95           T.95          0.31         1.00        £12.0 %
               2450               52.7                1.95              7.67         7.67           7.67          0.36         0.80        #12.0 %

               2600               52.5                216               7.39         7.39           7.39          0.29         0.80        4 12.0 %

               5250               48.9                5.36              5.24         5.24           5.24          0.40         1.90        £#131%
               5600               48.5                5.77              4.33         4.33           4.33          0.50         1.80        #131%
               5800               48.2                6.00              4.48         448            4.48          0.50         1.90        131 %


        © Frequency validity above 300 MHz of + 100 MHz only applies for DASY v4.4 and higher (see Page 2). else it is restricted to + 50 MHz. The:
        uncertainty is the RSS of the ConvF uncertainty at calibration frequency and the uncertainty for the indicated frequency band . Frequency valadity
        below 300 MHz is + 10. 25, 40, 50 and 70 MHz for ConyP assessments at 30, 64, 128. 150 and 220 MHz respeciively. Above 5 GHz frequancy
        walidity can be extended to £ 110 MHz.
        " At frequencies below 3 GHz, the valldity of tissue parameters (e and 5) can be relamed to £ 10% if liquid compensation form ule is applied to
        measured SAR valuss. At frequencias above 3 GHz, the validity of tissue paremeters (c and a} is restricted to * 5%. The uncertainty is the RSS of
        Lns ConwF uncertainty forindicated targat tissue pararmaters.
          Alpha/Depth are detenmined during calibration. SPEAG warrants thait the remaining «leviation due to the bowndary affect atter compensation is
        always less than 4 1% for frequencies below 3 /CMz and below & 2% for frequancies between 3—6 GHz at any distance larger than haill the probe fip
        clameter from the boundary.




        Certificate No: EX3—7441_Movi6                                     Page 6 of 11


Bay Area Compliance Laboratories Corp. (Kunshan)


Bay Area Compliance Laboratories Corp. (Kunshan)




        EX3DV4= SN:7441                                                                                                             November 15, 2016




                                           Receiving Pattern (¢), 8 = 0°
                             f2800 MHz,.TEM                                                                 f=1800 MHz,R22




                    ®              a              e             ©                        #                      @         &                     @
                    Tot            xX             ¥             2                       Tat                     X         ¥                     Z




                        F                                                        |                      |

              s

                        !    150            100            :'.l i               ':;.‘           j           D   T   100         j       'I"'J

                                   a
                                                            e
                                                           [®
                                                                              Rok ["]               e                               ®
                                   MHz                 1'.[?117'.'!1.-                        1300 MHz                        Efiéfi'fiilw

                                         Uncertainty of Axial Isotropy Assessment: £0.5% (k=2)




        Certificate No: EX3—7441_Nov16                                   Page 8 of 11


Bay Area Compliance Laboratories Corp. (Kunshan)


Bay Area Compliance Laboratories Corp. (Kunshan)


Bay Area Compliance Laboratories Corp. (Kunshan)


Bay Area Compliance Laboratories Corp. (Kunshan)



DIPOLE CALIBRATION CERTIFICATES


Bay Area Compliance Laboratories Corp. (Kunshan)


Bay Area Compliance Laboratories Corp. (Kunshan)


Bay Area Compliance Laboratories Corp. (Kunshan)


Bay Area Compliance Laboratories Corp. (Kunshan)


Bay Area Compliance Laboratories Corp. (Kunshan)


Bay Area Compliance Laboratories Corp. (Kunshan)


Bay Area Compliance Laboratories Corp. (Kunshan)




      Impedance Measurement Plot for Body TSL




                                                                                             ¢ How 2016             12132146
               (CHH   sif      1       U   F$            11 54.168 a      —%4844a      37.004 pF                   450,000 000 MHz


               #




               Co                                                                 _C ~%,                   I|,
                                                                                 ’;'         °V    i 6


                                                                                                            /
               fi‘g
               16"                                                                     se

               H1d



               CH2    §$11     LOB                  2 dB/REF   —~20 c8                  11—20.092 d8               450.0AN AnA MH=

                      mm _off e                           +                              4                                 .
                                                .       —_4__ 6e                                       }ovne zl
                                                                   line                           se~*"1
                                                                             .               >         4




               iys
              PV
                                   4                      M          }       M

               HLd                 t            t         f


                             StaFT 259,000 000 MHz                                                     sToP      55¢,    100 MHz




     Certificale No: M4SOV3—096_Nov16                                      Page 8 of 8