Calibration Report EP80

FCC ID: 2AABK-TM800W630L

RF Exposure Info

Download: PDF
FCCID_3366854

   COMOSAR E-Field Probe Calibration Report


                                             Ref : ACR.189.1.16.SATU.A




 SHENZHEN MORLAB COMMUNICATIONS
       TECHNOLOGY CO., LTD
  FL3, BUILDING A, FEIYANG SCIENCE PARK, NO.8
               LONGCHANG ROAD,
BLOCK 67, BAOAN DISTRICT, SHENZHEN, GUANGDONG
              PROVINCE, P.R. CHINA
 MVG COMOSAR DOSIMETRIC E-FIELD PROBE
                    SERIAL NO.: SN 37/08 EP80
                           Calibrated at MVG US
                 2105 Barrett Park Dr. - Kennesaw, GA 30144




                        Calibration Date: 07/05/2016

    Summary:
    This document presents the method and results from an accredited COMOSAR Dosimetric E-Field
    Probe calibration performed in MVG USA using the CALISAR / CALIBAIR test bench, for use
    with a COMOSAR system only. All calibration results are traceable to national metrology
    institutions.


                         COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.189.1.16.SATU.A




                                  Name                                Function                         Date                   Signature
Prepared by :                Jérôme LUC                         Product Manager                     7/7/2016

Checked by :                 Jérôme LUC                         Product Manager                     7/7/2016

Approved by :          Kim RUTKOWSKI                             Quality Manager                    7/7/2016




                                                                         Customer Name
                                                                        Shenzhen Morlab
                                                                        Communications
                                       Distribution :
                                                                        Technology Co.,
                                                                              Ltd




   Issue                   Date                                                           Modifications
     A                   7/7/2016                  Initial release




                                                                Page: 2/9

                This document shall not be reproduced, except in full or in part, without the written approval of MVG.
                The information contained herein is to be used only for the purpose for which it is submitted and is not to
                be released in whole or part without written approval of MVG.


                             COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.189.1.16.SATU.A




                                                      TABLE OF CONTENTS



1    Device Under Test ..................................................................................................... 4
2    Product Description ................................................................................................... 4
    2.1      General Information _______________________________________________________ 4
3    Measurement Method ................................................................................................ 4
    3.1      Linearity ________________________________________________________________ 4
    3.2      Sensitivity _______________________________________________________________ 5
    3.3      Lower Detection Limit _____________________________________________________ 5
    3.4      Isotropy _________________________________________________________________ 5
    3.5      Boundary Effect __________________________________________________________ 5
4    Measurement Uncertainty .......................................................................................... 5
5    Calibration Measurement Results .............................................................................. 6
    5.1      Sensitivity in air __________________________________________________________ 6
    5.2      Linearity ________________________________________________________________ 7
    5.3      Sensitivity in liquid ________________________________________________________ 7
    5.4      Isotropy _________________________________________________________________ 8
6    List of Equipment ...................................................................................................... 9




                                                                    Page: 3/9

                    This document shall not be reproduced, except in full or in part, without the written approval of MVG.
                    The information contained herein is to be used only for the purpose for which it is submitted and is not to
                    be released in whole or part without written approval of MVG.


                          COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.189.1.16.SATU.A




1              DEVICE UNDER TEST

                                                      Device Under Test
      Device Type                                                     COMOSAR DOSIMETRIC E FIELD PROBE
      Manufacturer                                                    MVG
      Model                                                           SSE5
      Serial Number                                                   SN 37/08 EP80
      Product Condition (new / used)                                  Used
      Frequency Range of Probe                                        0.7 GHz-3GHz
      Resistance of Three Dipoles at Connector                        Dipole 1: R1=1.445 M
                                                                      Dipole 2: R2=1.467 M
                                                                      Dipole 3: R3=1.477 M

A yearly calibration interval is recommended.

2              PRODUCT DESCRIPTION

2.1          GENERAL INFORMATION
MVG’s COMOSAR E field Probes are built in accordance to the IEEE 1528, OET 65 Bulletin C and
CEI/IEC 62209 standards.




                             Figure 1 – MVG COMOSAR Dosimetric E field Dipole

      Probe Length                                                    330 mm
      Length of Individual Dipoles                                    4.5 mm
      Maximum external diameter                                       8 mm
      Probe Tip External Diameter                                     5 mm
      Distance between dipoles / probe extremity                      2.7 mm


         3     MEASUREMENT METHOD
The IEEE 1528, OET 65 Bulletin C, CENELEC EN50361 and CEI/IEC 62209 standards provide
recommended practices for the probe calibrations, including the performance characteristics of
interest and methods by which to assess their affect. All calibrations / measurements performed meet
the fore mentioned standards.

         3.1   LINEARITY
The evaluation of the linearity was done in free space using the waveguide, performing a power
sweep to cover the SAR range 0.01W/kg to 100W/kg.


                                                                 Page: 4/9

                 This document shall not be reproduced, except in full or in part, without the written approval of MVG.
                 The information contained herein is to be used only for the purpose for which it is submitted and is not to
                 be released in whole or part without written approval of MVG.


                            COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.189.1.16.SATU.A




        3.2    SENSITIVITY
The sensitivity factors of the three dipoles were determined using a two step calibration method (air
and tissue simulating liquid) using waveguides as outlined in the standards.

        3.3    LOWER DETECTION LIMIT
The lower detection limit was assessed using the same measurement set up as used for the linearity
measurement. The required lower detection limit is 10 mW/kg.

        3.4    ISOTROPY
The axial isotropy was evaluated by exposing the probe to a reference wave from a standard dipole
with the dipole mounted under the flat phantom in the test configuration suggested for system
validations and checks. The probe was rotated along its main axis from 0 - 360 degrees in 15 degree
steps. The hemispherical isotropy is determined by inserting the probe in a thin plastic box filled
with tissue-equivalent liquid, with the plastic box illuminated with the fields from a half wave dipole.
The dipole is rotated about its axis (0°–180°) in 15° increments. At each step the probe is rotated
about its axis (0°–360°).

        3.5    BOUNDARY EFFECT
The boundary effect is defined as the deviation between the SAR measured data and the expected
exponential decay in the liquid when the probe is oriented normal to the interface. To evaluate this
effect, the liquid filled flat phantom is exposed to fields from either a reference dipole or waveguide.
With the probe normal to the phantom surface, the peak spatial average SAR is measured and
compared to the analytical value at the surface.

4             MEASUREMENT UNCERTAINTY

The guidelines outlined in the IEEE 1528, OET 65 Bulletin C, CENELEC EN50361 and CEI/IEC
62209 standards were followed to generate the measurement uncertainty associated with an E-field
probe calibration using the waveguide technique. All uncertainties listed below represent an
expanded uncertainty expressed at approximately the 95% confidence level using a coverage factor
of k=2, traceable to the Internationally Accepted Guides to Measurement Uncertainty.
 Uncertainty analysis of the probe calibration in waveguide
                                                  Uncertainty            Probability                                                Standard
        ERROR SOURCES                                                                             Divisor               ci
                                                   value (%)             Distribution                                            Uncertainty (%)
 Incident or forward power                            3.00%               Rectangular                   3               1             1.732%

 Reflected power                                      3.00%               Rectangular                   3               1             1.732%
                                                                                                   1.732051
 Liquid conductivity                                  5.00%               Rectangular                   3               1             2.887%
                                                                                                   1.732051
 Liquid permittivity                                  4.00%               Rectangular                   3               1             2.309%
                                                                                                   1.732051
 Field homogeneity                                    3.00%               Rectangular                   3               1             1.732%
                                                                                                   1.732051
 Field probe positioning                              5.00%               Rectangular                   3               1             2.887%
                                                                                                   1.732051

                                                                   Page: 5/9

                   This document shall not be reproduced, except in full or in part, without the written approval of MVG.
                   The information contained herein is to be used only for the purpose for which it is submitted and is not to
                   be released in whole or part without written approval of MVG.


                           COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.189.1.16.SATU.A




 Field probe linearity                               3.00%               Rectangular                   3               1             1.732%
                                                                                                  1.732051
  Combined standard uncertainty                                                                                                      5.831%
       Expanded uncertainty
                                                                                                                                     12.0%
     95 % confidence level k = 2



        5     CALIBRATION MEASUREMENT RESULTS


                                                   Calibration Parameters
   Liquid Temperature                                         21 °C
   Lab Temperature                                            21 °C
   Lab Humidity                                               45 %


        5.1     SENSITIVITY IN AIR


                                     Normx dipole Normy dipole Normz dipole
                                     1 (μV/(V/m)2) 2 (μV/(V/m)2) 3 (μV/(V/m)2)
                                          5.13          5.62          5.15


                                     DCP dipole 1 DCP dipole 2 DCP dipole 3
                                       (mV)         (mV)         (mV)
                                        129          109          123


Calibration curves ei=f(V) (i=1,2,3) allow to obtain H-field value using the formula:
                                                       E  E1  E2  E3
                                                                    2         2        2




                                                                  Page: 6/9

                  This document shall not be reproduced, except in full or in part, without the written approval of MVG.
                  The information contained herein is to be used only for the purpose for which it is submitted and is not to
                  be released in whole or part without written approval of MVG.


                      COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.189.1.16.SATU.A




      5.2   LINEARITY




      5.3   SENSITIVITY IN LIQUID

  Liquid        Frequency             Permittivity         Epsilon (S/m)              ConvF
                 (MHz +/-
                100MHz)
  HL450             450                   42.17                  0.86                   7.55
  BL450             450                   57.65                  0.95                   7.77
  HL750             750                   40.03                  0.93                   6.44
  BL750             750                   56.83                  1.00                   6.68
  HL900             900                   42.08                  1.01                   6.13
  BL900             900                   55.25                  1.08                   6.37
  HL1800           1800                   41.68                  1.46                   5.21
  BL1800           1800                   53.86                  1.46                   5.38
  HL1900           1900                   38.45                  1.45                   5.61
  BL1900           1900                   53.32                  1.56                   5.71
  HL2450           2450                   37.50                  1.80                   4.82
  BL2450           2450                   53.22                  1.89                   4.96
  HL2600           2600                   39.80                  1.99                   4.74
  BL2600           2600                   52.52                  2.23                   4.93



LOWER DETECTION LIMIT: 8mW/kg




                                                             Page: 7/9

             This document shall not be reproduced, except in full or in part, without the written approval of MVG.
             The information contained herein is to be used only for the purpose for which it is submitted and is not to
             be released in whole or part without written approval of MVG.


                       COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.189.1.16.SATU.A




       5.4   ISOTROPY
HL900 MHz
- Axial isotropy:                              0.04 dB
- Hemispherical isotropy:                      0.05 dB




HL1800 MHz
- Axial isotropy:                              0.04 dB
- Hemispherical isotropy:                      0.07 dB




                                                              Page: 8/9

              This document shall not be reproduced, except in full or in part, without the written approval of MVG.
              The information contained herein is to be used only for the purpose for which it is submitted and is not to
              be released in whole or part without written approval of MVG.


                            COMOSAR E-FIELD PROBE CALIBRATION REPORT                                                     Ref: ACR.189.1.16.SATU.A




        6    LIST OF EQUIPMENT


                                                  Equipment Summary Sheet

    Equipment                  Manufacturer /                        Current                                               Next Calibration
                                              Identification No.
    Description                   Model                          Calibration Date                                               Date
                                                                                        Validated. No cal               Validated.       No cal
    Flat Phantom                       MVG                   SN-20/09-SAM71
                                                                                        required.                       required.
                                                                                        Validated. No cal               Validated.       No cal
COMOSAR Test Bench                  Version 3                          NA
                                                                                        required.                       required.
                              Rhode & Schwarz
  Network Analyzer                                                SN100132                         02/2016                       02/2019
                                   ZVA

   Reference Probe                     MVG                    EP 94 SN 37/08                       10/2015                       10/2016

      Multimeter                 Keithley 2000                     1188656                         12/2013                       12/2016

   Signal Generator             Agilent E4438C                  MY49070581                         12/2013                       12/2016

                                                                                         Characterized prior to Characterized prior to
       Amplifier                  Aethercomm                        SN 046
                                                                                         test. No cal required. test. No cal required.

     Power Meter                   HP E4418A                    US38261498                         12/2013                       12/2016

    Power Sensor                 HP ECP-E26A                    US37181460                         12/2013                       12/2016

                                                                                         Characterized prior to Characterized prior to
  Directional Coupler           Narda 4216-20                        01386
                                                                                         test. No cal required. test. No cal required.
                                                                                        Validated. No cal               Validated. No cal
     Waveguide                  Mega Industries             069Y7-158-13-712
                                                                                        required.                       required.
                                                                                        Validated. No cal               Validated. No cal
Waveguide Transition            Mega Industries             069Y7-158-13-701
                                                                                        required.                       required.
                                                                                        Validated. No cal               Validated. No cal
Waveguide Termination           Mega Industries             069Y7-158-13-701
                                                                                        required.                       required.
Temperature / Humidity
                       Control Company                           150798832                         10/2015                       10/2017
      Sensor




                                                                   Page: 9/9

                   This document shall not be reproduced, except in full or in part, without the written approval of MVG.
                   The information contained herein is to be used only for the purpose for which it is submitted and is not to
                   be released in whole or part without written approval of MVG.