Darpa NESD BISC FCC Exhibit

0049-EX-CN-2018 Text Documents

Blackrock Microsystems, LLC

2018-01-19ELS_203615

       Exhibit of the research project, necessity of the wireless
          communication, and inadequateness of existing
                        communication facility

                                                             Table of Contents
 1.   Exhibit of the research project ............................................................................................................................... 1
 2.   Exhibit of the necessity of the wireless communication .............................................................................. 6
 3.   Exhibit of the inadequateness of existing communication facility ........................................................... 7




 1. EXHIBIT OF THE RESEARCH PROJECT
The authorization is to be used for providing wireless communication for an advanced wireless
neurosensing device that will be used for monitoring and stimulation of cortical neural signals (for
the use in a future visual prosthesis) in a clinical setting and also in the home environment for
research participants engaged in research sessions to investigate the DARPA NESD BISC I neural
interface. The wireless neurosensing device is dedicated for recording neural signals from the
patients’ cortical areas to facilitate recording and stimulation of neural signals for a visual
prosthesis and to investigate the principle feasibility of the device. The device is a fully implantable
device consisting of a 65,536-channel active CMOS flexible electrode array allowing simultaneous
recording and stimulation from up to 1,024 channels simultaneously. The array of 65,536 channels
for electrode inputs are connected to column and row selectors followed by filters to ultralow noise
ultralow power neural amplifiers (LNA), successive approximate analog to digital converters for
recording and a stimulator control circuit and a wireless data link embedded in one flexible CMOS
platform.
The device will transmit and receive data over an RF link and receive power over an inductive link.
The BISC I is intended to be implanted in a human patient for up to 4 months. An external, head
mounted relay station will power the device, receive data and wirelessly relay it to a computer based
base station for data acquisition and stimulation control.
External to the patient, the implantable wireless neurosensing device will interface with the
implanted device via a near field link, record the neural signals, and wirelessly transmit those to
a receiver unit for further signal processing, display, and storage. It will also receive wireless
commands for stimulation. The experiments will be carried out in a controlled and secured
environment with an area of less than 10 m diameter. The wireless neural interface device will
be used for indoor short distance (a few meters) application. A detailed description of the wireless
device is shown in Figure 1.


                                                      4GHz    4GHz




                                                                               TX                   Freq Synth
                     CDR            RX                                                                 ×20

                        DATA

                    Digital                           4GHz   4GHz                         `

                                                                                      Digital
                                    TX                                       RX/CDR                    ÷2
                    Control                                                           Control
          VCO                                                                                                     Master Clock
    fVCO = 200MHz                                 12.5 MHz   12.5 MHz                                            fREF = 200MHz

                           PLL
                                         LA                                   PA                       ÷16
                           ×16
                                              Power Coil        Power Coil




                              (a)                                                             (b)
Figure 1. Architecture of implantable neural interface: (a); the Wireless Neural Interface: (b)

The BISC I device incorporates two sub-units, the implant unit and the Wireless Neural Interface
(WNI) unit (external head mounted) as shown in Figure 1. The implant will be mounted between the
dura and cortex, the WNI will be mounted on top of the skin. The BISC I device uses three additional
customized wireless links for data and power transfer:
   1. The data downlink from the implant to the WNI uses a simple On-Off Keying (OOK)
      modulation scheme at a nominal carrier center frequency of 4GHz for the transmission of
      data from the implant to the WNI. The signal that modulates the carrier is a single 100
      Mbps digital serial bit stream that is the encoded version of the digital neural data. The
      data will be encoded in a 1ns pulse which has 1GHz bandwidth from the BISC chip to the
      wireless neural interface. And 2 ns pulse which 500MHz has bandwidth from the wireless
      neural interface to the BISC chip. In order to reduce the RF emission to the user (patient),
      the time-averaged power of the implantable wireless devices is ultralow and limited to be
      <0.35 mW or -4.5dBm. The simulation results are shown below (Figure 2-3), which clearly
      indicates the maximum average power it transmits is much less than the 1.0mW threshold
      per FCC 447498 D01 General RF Exposure Guidance v05r02 “Mobile and Portable Devices
      RF Exposure Procedures and Equipment Authorization Policies”, “4.2.4 Transmitters
      implanted in the body of a user”, “When the aggregate of the maximum power available at the
      antenna port and radiating structures of an implanted transmitter, under all operating
      circumstances, is ≤ 1.0 mW, SAR test exclusion may be applied.” Therefore, the data downlink
      from the implant to the WNI falls into SAR test exclusion.


 V (mV)




 I (mA)




P (mW)




Figure 2. Simulated instantaneous voltage (V), current (I), and power (P) of a single pulse on
implantable antenna input. P = V * I




Figure 3. Simulated integrated instantaneous power ∫Pdt in pico-Joule. The pulse has 10% duty
factor. Assuming maximum pulse repetition rate 100MHz, time-averaged power of a pulse in one
period T=10ns is ∫Pdt /T = 0.336mW.

2. The data uplink from the WNI to the implant also uses a 4GHz OOK modulation ultralow
   power transmitter and is limited to be 0.135 mW or -8.3 dBm at the antenna input port.
   The simulation results (Figure 4) below show the schematic of the WNI module, which
   includes both the data uplink and the wireless power transfer module (which will


be address later in #3). For the uplink from WNI to the implant, as shown in Figure 4., the
output of ADF4351BCPZ is set to +5dBm. Using the circuit configuration in “Linear
Technology AN-98-8 ‘Nanosecond Pulse Width Generator’”, 2ns pulse width 18dBm pulses
can be generated from the pulse generator with maximum pulse repetition rate of 50 MHz.
The pulses then attenuated 10 dB by Mini-Circuits LAT-10+. The conversion loss of the mixer
is 6dB, and insertion loss the ADI HMC1118 switch is 0.7dB. The peak power at antenna port
18 – 10 – 6 – 0.7 = 1.3 dBm or 1.35mW. The time-averaged power at antenna will be 1.35mW
* 2ns/20ns = 0.135mW or -8.7 dBm. Considering possible minimum distance between the
external transmitter of the wireless device and the user under normal condition is > 10mm,
, for Mobile and Portable Devices (KDB 447498), according to FCC 447498 D01 General RF
Exposure Guidance v05r02 “Mobile and Portable Devices RF Exposure Procedures and
Equipment Authorization Policies”, “4.3. General SAR test reduction and exclusion guidance”,
the 1-g and 10-g SAR test exclusion thresholds for 100 MHz to 6 GHz at test separation
distances ≤ 50 mm are determined by: [(max. power of channel, including tune-up tolerance,
mW)/(min. test separation distance, mm)] · [√f(GHz)] ≤ 3.0 for 1-g SAR and ≤ 7.5 for 10-g
extremity SAR. For the proposed wireless device, max. power of channel=1mW, min. test
separation distance=10mm, f(GHz)= 4.0GHz. It will give us: 1/10*√4= 0.1, which is much less
than both 3.0 for 1-g SAR and 7.5 for 10-g extremity SAR. In additional, according to “Mobile
and Portable Devices RF Exposure Procedures and Equipment Authorization Policies”,
“Appendix A SAR Test Exclusion Thresholds for 100 MHz – 6 GHz and ≤ 50 mm”, for 5.2GHz
at 10mm, the SAR Test Exclusion Threshold is 13 mW, which is much larger than the
proposed device, therefore the proposed device data uplink falls into “Standalone SAR test
exclusion”.


                                            Wireless Power


                                                                      12.5MHz Reference to Power Amplifier


                            VGA & PA
        Power Coil         ADI AD8331



                                                                              4GHz Carrier to Mixer                 12.5MHz Reference to FPGA

                                                                               RF Mixer               VCO & PLL
                                                                        MiniCircuits MAC1-85+     ADI ADF4351BCPZ




   UWB Antenna
                      RF Switch               Wireless Data
                     ADI HMC1118                                                                                    50 MBps Data to Mixer



                                                                                     Attenuator
                                                                                MiniCircuits LAT-10+
                                                                                                         Pulse Generator
                                                                                                       LT1721 and 74AHC08                   FPGA
                                                                                                                    100 MBps Data to FPGA



                                            BPF                        LNA              RF Power Detector     Comparator
                                   TDK DEA453960BT-3007B1   MiniCircuits ZX60-83LN+       ADI ADL5511         LT LTC6754




Figure 4. Schematic of Wireless Interface Board.


3. The power link from the WNI to the implant use a high efficiency inductive link running at 12
   MHz. The schematic of the power transmitter is shown in Figure 4. The VCO maximum output
   power is 5dBm and the variable gain amplifier AD8331 is set to be no more than 22dB.
   Therefore, the maximum output power at the input of the WNI primary coil will be 5+22 =
   27dBm or 508mW. Figure 5-6 shows the primary and secondary coils design and the
   simulated results of the power link indicating a maximum transmitting power of 508mW and
   an efficiency of 11.2%. According to FCC 447498 D01 General RF Exposure Guidance v05r02
   “Mobile and Portable Devices RF Exposure Procedures and Equipment Authorization
   Policies”, “4.3. General SAR test reduction and exclusion guidance”, “4.3.1 c) 2)”, the
   maximum allowed power is calculated as:
        a. 1-g: 3*50/√0.0125*[1+log(100/12.5)]*1/2=1276mW
        b. 10-g: 7.5*50/√0.0125*[1+log(100/12.5)]*1/2=3192mW
   Since the maximum power transmitted by the primary coil on the WNI will be only 508mW,
   it is much less than both the 1-g and 10-g threshold for SAR exclusion under the operating
   frequency and distance, then the power link of the BSIC device also falls in the SAR exclusion
   category.


Figure 5. Configuration of the power coil




Figure 6. Simulation results of the power coil. The total required power input to the primary side of
the power coil on is 509 mW. Received power from the secondary coil on the implantable chip is
57.4 mW.


2. EXHIBIT OF THE NECESSITY OF THE WIRELESS COMMUNICATION
The wireless communication proposed for this application is one of the key technical advances
for this project. Without the wireless communication, a traditional wired neurosensing system
has to be used requiring a tethered connection between the subject and the system, which not
only greatly limits the patient mobility and patient care cost, but requires a percutaneous
connector. In addition, the ultralow power feature provides minimum RF emission and low heat
generation of the device. The 100 Mbps data is produced by the 65,535-channel high resolution
recording of the device and is a requirement of the research project. The use of OOK modulation
scheme is an essential tradeoff between wireless transition power, device simplicity, wireless
link fidelity, and patient safety. Due to its simplicity, OOK can reduce the electronics resources
needed to be implemented on to the wireless device, which will be attached to the subject’s head
(external wireless headstage and implantable device). This means, the size and power dissipation
of the device will be greatly reduced from the wired device, hence the weight of the device will be
greatly reduced as well.


3. EXHIBIT OF THE INADEQUATENESS OF EXISTING COMMUNICATION FACILITY
Considering the high data rate and low power requirements of the device used in the research
project, feasible commercial wireless communication electronics are inadequate at this point.
For instance, to satisfy the 100 Mbps data rate, 802.11 wireless LANs such as a, g, n, or ac, are
needed. However, these commercial wireless communication options usually deliver very high
power up to 1W (usually a few hundred mW) to the transmitting antenna. This is about 1000
times higher than the proposed wireless scheme. Such high RF output power could pose RF
emission risks to the subject and often requires even higher power dissipation (a few watts) at
the transmitter causing increased heat generation and larger form factor of the head-mounted
wireless device.