RF Exposure INfo

FCC ID: 2AFGB100100

RF Exposure Info

Download: PDF

                                                                                    Test Report Number: 4237167EMC06 Rev: 0
                                                                                Prima—Temp c/o Boulder BioMed / PT180207—04
                                                                                                                                    Page: 1 of 3



                               RF Exposure Evaluation
                     Project Number:               4261199
                      Report Number:               4261199EMCO6                       Revision Number:                 0
                                     Client:       Younes Medical



          Equipment Under Test: Wireless Sleep Diagnostics
                          Model Name:               Prodigy 2 System
                       Model Number:                Prodigy 2 (CMU)




           Applicable Standards:                   47 CFR §§ 2.1091 and 2.1093;
                                                    FCC KDB 447498 D01 General RF Exposure Guidance v06




Remarks: This report details the results of the testing carried out on one sample, the results contained in this test report do notrelate to other
samples of the same product. The manufacturer should ensure that all products in series production are in conformity with the product sample
detailed in this report.
This documentis issued by the Company under its General Conditions of Service accessible at hitp//www.sqs.com/en/Terms—and—
Conditions.aspx. Aftention is drawn to the limitation of liability, indemnification and jurisdiction issues defined therein.
Any holder of this documentis advised that information contained hereon reflects the Company‘s findings at thetime of its intervention only and
within the limits of Client‘s instructions, if any. The Company‘s sole responsibility is to its Client and this document does not exonerate parties to
a fransaction from exercising all their rights and obligations underthe transaction documents. Any unauthorized alteration, forgery orfalsification
of the content or appearance of this documentis unlawful and offenders may be prosecuted to the fullest extentof the law.


                                                                                 Test Report Number: 4237167EMC06 Rev: 0
                                                                              Prima-Temp c/o Boulder BioMed / PT180207-04
                                                                                                               Page: 2 of 3


1 General Information
1.1     Client Information
                         Name: Younes Medical Technologies
                       Address: Unit 5-55 Henlow Bay
      City, State, Zip, Country: Winnipeg Manitoba, R3Y 1G4, Canada

1.2     Test Laboratory
                         Name: SGS North America, Inc.
                       Address: 620 Old Peachtree Road NW, Suite 100
      City, State, Zip, Country: Suwanee, GA 30024, USA

             Accrediting Body: A2LA
                    Type of lab: Testing Laboratory
            Certificate Number: 3212.01

1.3     General Information of EUT
                           Type of Product:         Wireless Sleep Monitor
                              Model Name:           Prodigy 2 System
                            Model Number:           Prodigy 2
                            Serial Number:          CMU 7

                          Frequency Range: 2412-2472MHz
                               Data Modes: 802.11b – 1 Mbps (DSSS)
                                  Antenna: Integral

                              Rated Voltage: 3.7 Vdc Battery
                               Test Voltage: 3.7 Vdc Battery

                 Sample Received Date: 3/21/2018




 SGS North America Inc.   Consumer and Retail   620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024   t (770) 570-1800         www.sgs.com
                                                                                                                       Member of the SGS Group (SGS SA)


                                                                                                 Test Report Number: 4237167EMC06 Rev: 0
                                                                                              Prima-Temp c/o Boulder BioMed / PT180207-04
                                                                                                                               Page: 3 of 3


2 SAR Exclusion Calculations
                                                    447498 D01 General RF Exposure Guidance v06
                                                                    SAR test exclusion calculations
                            Section 4.3: General SAR test exclusion guidance / Section 4.3.1: Standalone SAR test exclusion considerations
                                               Select
                                    Input                Unit correction
                                               Units
                   Max Power:        3.4        dBm       2.187761624
     Min separation distance:         5          mm                   5
                 Frequency, f:       2442        MHz              2442

              Value reference         Values used
                                                         Reference number definition
                      Number         for Calculation
                           v1         2         mW       [max. power of channel, including tune-up tolerance, mW] 'Rounded to nearest mW
                             v2       5          mm      [min. test separation distance, mm] 'Rounded to nearest mm
                             v3     1.563                [√f(GHz)]

a) For 100 MHz to 6 GHz and test separation distances ≤ 50 mm, the 1-g and 10-g SAR test exclusion thresholds are determined by the following:
   [(max. power of channel, including tune-up tolerance, mW) / (min. test separation distance, mm)] · [√f(GHz)] ≤ 3.0 for 1-g SAR, and ≤ 7.5 for 10-g extremity SAR,
       1g Exclusion Threshold:      9.6         mW       <== 3 * v2 / v3
       10g Exlusion Threshold:       24.0        mW      <== 7.5 * v2 / v3

                                  The EUT max power is BELOW the threshold. SAR Testing is NOT required for Body applications
             Conclusions:         The EUT max power is BELOW the threshold. SAR Testing is NOT required for Extremity applications




 SGS North America Inc.       Consumer and Retail         620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024                 t (770) 570-1800           www.sgs.com
                                                                                                                                                 Member of the SGS Group (SGS SA)



Document Created: 2019-01-24 16:15:36
Document Modified: 2019-01-24 16:15:36

© 2019 FCC.report