Message:


Mr Ghaffari, I revised the ERP power measurement downward as you suggested for both the 410 to 490
kHz operation (1.5 micro-watts) and for the 10 to 24 kHz band (180 micro-watts). Both calculations were
based on OET 63 and the Part 15.209 rules for maximum field strength at 300 meters.